Literature DB >> 21549439

A national dosimetric audit of IMRT.

Geoff Budgell1, Joe Berresford, Michael Trainer, Ellie Bradshaw, Peter Sharpe, Peter Williams.   

Abstract

BACKGROUND AND
PURPOSE: A dosimetric audit of IMRT has been carried out within the UK between June 2009 and March 2010 in order to provide an independent check of safe implementation and to identify problems in the modelling and delivery of IMRT. METHODS AND MATERIALS: A mail based audit involving film and alanine dosimeters was utilized. Measurements were made for each individual field in an IMRT plan isocentrically in a flat water-equivalent phantom at a depth of 5cm. The films and alanine dosimeters were processed and analysed centrally; additional ion chamber measurements were made by each participating centre.
RESULTS: 57 of 62 centres participated, with a total of 78 plans submitted. For the film measurements, all 176 fields from the less complex IMRT plans (including prostate and breast plans) achieved over 95% pixels passing a gamma criterion of 3%/3mm within the 20% isodose. For the more complex IMRT plans (mainly head and neck) 8/245 fields (3.3%) achieved less than 95% pixels passing a 4%/4mm gamma criterion. Of the alanine measurements, 4/78 (5.1%) of the measurements differed by >5% from the dose predicted by the treatment planning system. Three of these were large deviations of -77.1%, -29.1% and 14.1% respectively. Excluding the three measurements outside 10%, the mean difference was 0.05% with a standard deviation of 1.5%. The out of tolerance results have been subjected to further investigations.
CONCLUSIONS: A dosimetric audit has been successfully carried out of IMRT implementation by over 90% of UK radiotherapy departments. The audit shows that modelling and delivery of IMRT is accurate, suggesting that the implementation of IMRT has been carried out safely.
Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21549439     DOI: 10.1016/j.radonc.2011.03.016

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  11 in total

1.  A national dosimetry audit of intraoperative radiotherapy.

Authors:  D J Eaton; B Earner; P Faulkner; N Dancer
Journal:  Br J Radiol       Date:  2013-10-16       Impact factor: 3.039

2.  Adaptation and validation of a commercial head phantom for cranial radiosurgery dosimetry end-to-end audit.

Authors:  Alexis Dimitriadis; Antony L Palmer; Russell A S Thomas; Andrew Nisbet; Catharine H Clark
Journal:  Br J Radiol       Date:  2017-05-25       Impact factor: 3.039

3.  Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials.

Authors:  Catharine H Clark; Edwin G A Aird; Steve Bolton; Elizabeth A Miles; Andrew Nisbet; Julia A D Snaith; Russell A S Thomas; Karen Venables; David I Thwaites
Journal:  Br J Radiol       Date:  2015-09-02       Impact factor: 3.039

4.  Multi-institutional evaluation using the end-to-end test for implementation of dynamic techniques of radiation therapy in Thailand.

Authors:  Lalida Tuntipumiamorn; Puangpen Tangboonduangjit; Taweap Sanghangthum; Rattapol Rangseevijitprapa; Chirasak Khamfongkhruea; Thida Niyomthai; Boonrut Vuttiprasertpong; Supranee Supanant; Nongnuch Chatchaipaiboon; Porntip Iampongpaiboon; Pitchayut Nakkrasae; Tanwiwat Jaikuna
Journal:  Rep Pract Oncol Radiother       Date:  2018-11-28

5.  Inter-departmental dosimetry audits - development of methods and lessons learned.

Authors:  David J Eaton; Steve Bolton; Russell A S Thomas; Catharine H Clark
Journal:  J Med Phys       Date:  2015 Oct-Dec

6.  Confidence limits for patient-specific IMRT dose QA: a multi-institutional study in Korea.

Authors:  Jung-In Kim; Jin-Beom Chung; Ju-Young Song; Sung Kyu Kim; Yunseok Choi; Chang Heon Choi; Won Hoon Choi; Byungchul Cho; Jin Sung Kim; Sung Jin Kim; Sung-Joon Ye
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

7.  Validation of the RapidArc Delivery System Using a Volumetric Phantom as Per Task Group Report 119 of the American Association of Physicists in Medicine.

Authors:  Lalit Kumar; Girigesh Yadav; Vimal Kishore; Manindra Bhushan; Munish Gairola; Deepak Tripathi
Journal:  J Med Phys       Date:  2019 Apr-Jun

8.  Regional cancer centre demonstrates voluntary conformity with the national Radiation Oncology Practice Standards.

Authors:  Stephen Manley; Andrew Last; Kenneth Fu; Stuart Greenham; Andrew Kovendy; Thomas P Shakespeare
Journal:  J Med Radiat Sci       Date:  2015-04-08

9.  A multi-institutional study for tolerance and action levels of IMRT dose quality assurance measurements in Korea.

Authors:  Jung-in Kim; Jin-Beom Chung; Yang-Kyun Park; Ju-Young Song; Sung Kyu Kim; Sung Hwan Ahn; Chang Heon Choi; Won Hoon Choi; Byungchul Cho; Sang Gyu Ju; Sung Jin Kim; Sung-Joon Ye
Journal:  J Appl Clin Med Phys       Date:  2013-03-04       Impact factor: 2.102

10.  Different Dosimeters/Detectors Used in Small-Field Dosimetry: Pros and Cons.

Authors:  Wrya Parwaie; Soheila Refahi; Mahdieh Afkhami Ardekani; Bagher Farhood
Journal:  J Med Signals Sens       Date:  2018 Jul-Sep
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.