Relationship between left atrium catheter contact force and pacing threshold.

Teresa Barrio-López1,2, Mercedes Ortiz3, Eduardo Castellanos3, Carla Lázaro3, Jefferson Salas3, Sergio Madero3, Jesús Almendral3.   

Abstract

PURPOSE: The purpose of this study is to analyze the relationship between contact force (CF) and pacing threshold in left atrium (LA).
METHODS: Six to ten LA sites were studied in 28 consecutive patients with atrial fibrillation undergoing pulmonary vein isolation. Median CF, bipolar and unipolar electrogram voltage, impedance, and bipolar and unipolar thresholds for consistent constant capture and for consistent intermittent capture were measured at each site.
RESULTS: Pacing threshold measurements were performed at 188 LA sites. Both unipolar and bipolar pacing thresholds correlated significantly with median CF; however, unipolar pacing threshold correlated better (unipolar: Pearson R -0.45; p < 0.001; Spearman Rho -0.62; p < 0.001, bipolar: Pearson R -0.39; p < 0.001; Spearman Rho -0.52; p < 0.001). Consistent constant capture threshold had better correlation with median CF than consistent intermittent capture threshold for both unipolar and bipolar pacing (Pearson R -0.45; p < 0.001 and Spearman Rho -0.62; p < 0.001 vs. Pearson R -0.35; p < 0.001; Spearman Rho -0.52; p < 0.001). The best pacing threshold cutoff point to detect a good CF (>10 g) was 3.25 mA for unipolar pacing with 69% specificity and 73% sensitivity. Both increased to 80% specificity and 74% sensitivity for sites with normal bipolar voltage and a pacing threshold cutoff value of 2.85 mA.
CONCLUSIONS: Pacing thresholds correlate with CF in human not previously ablated LA. Since the combination of a normal bipolar voltage and a unipolar pacing threshold <2.85 mA provide reasonable parameters of validity, pacing threshold could be of interest as a surrogate for CF in LA.

Entities:  

Keywords:  Ablation; Contact force; Left atrium; Pacing threshold

Mesh:

Year:  2017        PMID: 28451774     DOI: 10.1007/s10840-017-0253-7

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  22 in total

1.  Effects of electrode-myocardial separation on cardiac stimulation in conductive solution.

Authors:  S B Knisley; A E Pollard; V G Fast
Journal:  J Cardiovasc Electrophysiol       Date:  2000-10

2.  Ablation effect indicated by impedance fall is correlated with contact force level during ablation for atrial fibrillation.

Authors:  Alessandro De Bortoli; Li-Zhi Sun; Eivind Solheim; Per Ivar Hoff; Peter Schuster; Ole-Jørgen Ohm; Jian Chen
Journal:  J Cardiovasc Electrophysiol       Date:  2013-07-19

3.  Initial impedance decrease as an indicator of good catheter contact: insights from radiofrequency ablation with force sensing catheters.

Authors:  Tobias Reichlin; Sven Knecht; Christopher Lane; Michael Kühne; Eyal Nof; Nagesh Chopra; Thomas M Tadros; Vivek Y Reddy; Beat Schaer; Roy M John; Stefan Osswald; William G Stevenson; Christian Sticherling; Gregory F Michaud
Journal:  Heart Rhythm       Date:  2013-10-28       Impact factor: 6.343

4.  The effects of electrode-tissue contact on radiofrequency lesion generation.

Authors:  B Avitall; K Mughal; J Hare; R Helms; D Krum
Journal:  Pacing Clin Electrophysiol       Date:  1997-12       Impact factor: 1.976

5.  Relation between impedance and endocardial contact during radiofrequency catheter ablation.

Authors:  S A Strickberger; V R Vorperian; K C Man; B D Williamson; S J Kalbfleisch; C Hasse; F Morady; J J Langberg
Journal:  Am Heart J       Date:  1994-08       Impact factor: 4.749

6.  Loss of pace capture on the ablation line: a new marker for complete radiofrequency lesions to achieve pulmonary vein isolation.

Authors:  Daniel Steven; Vivek Y Reddy; Keiichi Inada; Kurt C Roberts-Thomson; Jens Seiler; William G Stevenson; Gregory F Michaud
Journal:  Heart Rhythm       Date:  2009-11-12       Impact factor: 6.343

7.  Relationship between catheter contact force and radiofrequency lesion size and incidence of steam pop in the beating canine heart: electrogram amplitude, impedance, and electrode temperature are poor predictors of electrode-tissue contact force and lesion size.

Authors:  Atsushi Ikeda; Hiroshi Nakagawa; Hendrik Lambert; Dipen C Shah; Edouard Fonck; Aude Yulzari; Tushar Sharma; Jan V Pitha; Ralph Lazzara; Warren M Jackman
Journal:  Circ Arrhythm Electrophysiol       Date:  2014-11-07

8.  Predictive value of impedance changes for real-time contact force measurements during catheter ablation of atrial arrhythmias in humans.

Authors:  Saurabh Kumar; H M Haqqani; Martin Chan; Justin Lee; Matias Yudi; Michael C G Wong; Joseph B Morton; Liang-Han Ling; Timothy Robinson; Patrick M Heck; Nicholas F Kelland; Karen Halloran; Steven J Spence; Peter M Kistler; Jonathan M Kalman
Journal:  Heart Rhythm       Date:  2013-03-22       Impact factor: 6.343

9.  Activation time determination by high-resolution unipolar and bipolar extracellular electrograms in the canine heart.

Authors:  G Ndrepepa; E B Caref; H Yin; N el-Sherif; M Restivo
Journal:  J Cardiovasc Electrophysiol       Date:  1995-03

10.  Circumferential pulmonary vein isolation and linear left atrial ablation as a single-catheter technique to achieve bidirectional conduction block: the pace-and-ablate approach.

Authors:  Charlotte Eitel; Gerhard Hindricks; Philipp Sommer; Thomas Gaspar; Simon Kircher; Ulrike Wetzel; Nicos Dagres; Masahiro Esato; Andreas Bollmann; Daniela Husser; Sebastian Hilbert; Ruzbeh Zaker-Shahrak; Arash Arya; Christopher Piorkowski
Journal:  Heart Rhythm       Date:  2009-10-12       Impact factor: 6.343

View more
  2 in total

1.  Atrial mapping during pulmonary vein pacing to detect conduction gaps in a second pulmonary vein isolation procedure.

Authors:  Maria Teresa Barrio-López; Eduardo Castellanos; Mercedes Ortiz; Martín Arceluz; Carla Lázaro; Jefferson Salas; Sergio Madero; Jesús Almendral
Journal:  J Interv Card Electrophysiol       Date:  2018-05-25       Impact factor: 1.900

2.  Electrogram voltage and pacing threshold before ablation, measured by mini-electrodes, predict parameters indicative of transmural lesions in the human atrium.

Authors:  Carla Lázaro; Teresa Barrio-López; Eduardo Castellanos; Mercedes Ortiz; Martín Arceluz; Jesús Almendral
Journal:  J Interv Card Electrophysiol       Date:  2019-05-02       Impact factor: 1.900

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.