Monte Carlo study of the chamber-phantom air gap effect in a magnetic field.

Daniel J O'Brien1, Gabriel O Sawakuchi1,2.   

Abstract

PURPOSE: The aim of this study was to examine the effect of submillimeter air gaps that may exist between an ionization chamber and solid phantoms when measurements are performed in a magnetic field.
METHODS: Geant4 Monte Carlo simulations were performed using a model of a PTW 30013 Farmer chamber in a water phantom. Symmetrical and asymmetrical air gaps of various thicknesses were modeled surrounding the chamber, and the dose to the air cavity of the chamber was scored in each case. Magnetic fields were modeled parallel to the long axis of the chamber with strengths of 0, 0.35 T, 1.0 T, and 1.5 T. To examine the phenomenon in more detail, the gyroradii of the electrons responsible for the energy deposited in the chamber were scored as they entered the chamber and the total energy deposited was split into three components: energy originating from inside the chamber, in the immediate vacinity of the chamber, or outside the chamber.
RESULTS: Differences in the chamber dose of 1.6% were observed for asymmetric air gaps just 0.2 mm thick. No effect greater than 0.5% was observed for the symmetrical air gaps investigated in this work (1.4 mm thick or less) for this chamber/magnetic field configuration. The mean gyroradius of contributing electrons as they first enter the chamber was 4 mm. The presence of the air gap reduced the energy contributions from electrons released in the immediate vicinity of the chamber, and this loss was not completely compensated for when a magnetic field was present.
CONCLUSIONS: The gyroradius of most electrons was too large to be responsible for the air gap effect via the electron return effect; instead, the effect is attributed to the loss of energy contributions from electrons originating inside the air gap volume, which is not completely compensated for by more distant electrons owing to their reduced range in the magnetic field. When the chamber is parallel with the magnetic field, symmetric air gaps have a smaller effect (< 0.5%) compared to asymmetric air-gaps (up to 1.6%) on the chamber response.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  MRIgRT; air gaps; dosimetry; magnetic field; solid phantoms

Mesh:

Year:  2017        PMID: 28432792      PMCID: PMC5503156          DOI: 10.1002/mp.12290

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  History by history statistical estimators in the BEAM code system.

Authors:  B R B Walters; I Kawrakow; D W O Rogers
Journal:  Med Phys       Date:  2002-12       Impact factor: 4.071

2.  Magnetic-field-induced dose effects in MR-guided radiotherapy systems: dependence on the magnetic field strength.

Authors:  A J E Raaijmakers; B W Raaymakers; J J W Lagendijk
Journal:  Phys Med Biol       Date:  2008-01-18       Impact factor: 3.609

3.  Integrating a 1.5 T MRI scanner with a 6 MV accelerator: proof of concept.

Authors:  B W Raaymakers; J J W Lagendijk; J Overweg; J G M Kok; A J E Raaijmakers; E M Kerkhof; R W van der Put; I Meijsing; S P M Crijns; F Benedosso; M van Vulpen; C H W de Graaff; J Allen; K J Brown
Journal:  Phys Med Biol       Date:  2009-05-19       Impact factor: 3.609

4.  First MR images obtained during megavoltage photon irradiation from a prototype integrated linac-MR system.

Authors:  B G Fallone; B Murray; S Rathee; T Stanescu; S Steciw; S Vidakovic; E Blosser; D Tymofichuk
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

5.  Quantification of static magnetic field effects on radiotherapy ionization chambers.

Authors:  J Agnew; F O'Grady; R Young; S Duane; G J Budgell
Journal:  Phys Med Biol       Date:  2017-01-10       Impact factor: 3.609

6.  Electron contamination modeling and reduction in a 1 T open bore inline MRI-linac system.

Authors:  B M Oborn; S Kolling; P E Metcalfe; S Crozier; D W Litzenberg; P J Keall
Journal:  Med Phys       Date:  2014-05       Impact factor: 4.071

7.  Reference dosimetry in magnetic fields: formalism and ionization chamber correction factors.

Authors:  D J O'Brien; D A Roberts; G S Ibbott; G O Sawakuchi
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

8.  Charged particle transport in magnetic fields in EGSnrc.

Authors:  V N Malkov; D W O Rogers
Journal:  Med Phys       Date:  2016-07       Impact factor: 4.071

9.  Consequences of air around an ionization chamber: Are existing solid phantoms suitable for reference dosimetry on an MR-linac?

Authors:  S L Hackett; B van Asselen; J W H Wolthaus; J G M Kok; S J Woodings; J J W Lagendijk; B W Raaymakers
Journal:  Med Phys       Date:  2016-07       Impact factor: 4.071

  9 in total
  4 in total

1.  Feasibility of spinal stereotactic body radiotherapy in Elekta Unity® MR-Linac.

Authors:  Eun Young Han; Manik Aima; Neil Hughes; Tina M Briere; Debra N Yeboa; Pam Castillo; Jihong Wang; Jinzhong Yang; Sastry Vedam
Journal:  J Radiosurg SBRT       Date:  2020

2.  Relative dosimetry with an MR-linac: Response of ion chambers, diamond, and diode detectors for off-axis, depth dose, and output factor measurements.

Authors:  Daniel J O'Brien; James Dolan; Stefan Pencea; Nicholas Schupp; Gabriel O Sawakuchi
Journal:  Med Phys       Date:  2017-12-21       Impact factor: 4.071

3.  MOSFET dosimeter characterization in MR-guided radiation therapy (MRgRT) Linac.

Authors:  Poonam Yadav; Abdelbasset Hallil; Dinesh Tewatia; David A P Dunkerley; Bhudatt Paliwal
Journal:  J Appl Clin Med Phys       Date:  2019-12-18       Impact factor: 2.102

4.  Dosimetric evaluation of irradiation geometry and potential air gaps in an acrylic miniphantom used for external audit of absolute dose calibration for a hybrid 1.5 T MR-linac system.

Authors:  Neelam Tyagi; Ergys Subashi; Dale Michael Lovelock; Stephen Kry; Paola Elisa Alvarez; Margie A Hunt; Seng Boh Lim
Journal:  J Appl Clin Med Phys       Date:  2021-12-16       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.