Consequences of air around an ionization chamber: Are existing solid phantoms suitable for reference dosimetry on an MR-linac?

S L Hackett1, B van Asselen1, J W H Wolthaus1, J G M Kok1, S J Woodings1, J J W Lagendijk1, B W Raaymakers1.   

Abstract

PURPOSE: A protocol for reference dosimetry for the MR-linac is under development. The 1.5 T magnetic field changes the mean path length of electrons in an air-filled ionization chamber but has little effect on the electron trajectories in a surrounding phantom. It is therefore necessary to correct the response of an ionization chamber for the influence of the magnetic field. Solid phantoms are used for dosimetry measurements on the MR-linac, but air is present between the chamber wall and phantom insert. This study aimed to determine if this air influences the ion chamber measurements on the MR-linac. The absolute response of the chamber and reproducibility of dosimetry measurements were assessed on an MR-linac in solid and water phantoms. The sensitivity of the chamber response to the distribution of air around the chamber was also investigated.
METHODS: Measurements were performed on an MR-linac and replicated on a conventional linac for five chambers. The response of three waterproof chambers was measured with air and with water between the chamber and the insert to measure the influence of the air volume on absolute chamber response. The distribution of air around the chamber was varied indirectly by rotating each chamber about the longitudinal chamber axis in a solid phantom and a water phantom (waterproof chambers only) and measuring the angular dependence of the chamber response, and varied directly by displacing the chamber in the phantom insert using a paper shim positioned at different orientations between the chamber casing and the insert.
RESULTS: The responses of the three waterproof chambers measured on the MR-linac were 0.7%-1.2% higher with water than air in the chamber insert. The responses of the chambers on the conventional linac changed by less than 0.3% when air in the insert was replaced with water. The angular dependence of the chambers ranged from 0.6% to 1.9% in the solid phantom on the MR-linac but was less than 0.5% in water on the MR-linac and less than 0.3% in the solid phantom on the conventional linac. Inserting a shim around the chamber induced changes of the chamber response in a magnetic field of up to 2.2%, but the change in chamber response on the conventional linac was less than 0.3%.
CONCLUSIONS: The interaction between the magnetic field and secondary electrons in the air around the chamber reduces the charge collected from 0.7% to 1.2%. The large angular dependence of ion chambers measured in the plastic phantom in a magnetic field appears to arise from a change of air distribution as the chamber is moved within the insert, rather than an intrinsic isotropy of the chamber sensitivity to radiation. It is recommended that reference dosimetry measurements on the MR-linac can be performed only in water, rather than in existing plastic phantoms.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27370114     DOI: 10.1118/1.4952727

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Monte Carlo study of the chamber-phantom air gap effect in a magnetic field.

Authors:  Daniel J O'Brien; Gabriel O Sawakuchi
Journal:  Med Phys       Date:  2017-05-26       Impact factor: 4.071

Review 2.  The transformation of radiation oncology using real-time magnetic resonance guidance: A review.

Authors:  William A Hall; Eric S Paulson; Uulke A van der Heide; Clifton D Fuller; B W Raaymakers; Jan J W Lagendijk; X Allen Li; David A Jaffray; Laura A Dawson; Beth Erickson; Marcel Verheij; Kevin J Harrington; Arjun Sahgal; Percy Lee; Parag J Parikh; Michael F Bassetti; Clifford G Robinson; Bruce D Minsky; Ananya Choudhury; Robert J H A Tersteeg; Christopher J Schultz
Journal:  Eur J Cancer       Date:  2019-10-12       Impact factor: 9.162

3.  Effect of Magnetic Field Strength on Plastic Scintillation Detector Response.

Authors:  F Therriault-Proulx; Z Wen; G Ibbott; S Beddar
Journal:  Radiat Meas       Date:  2018-06-09       Impact factor: 1.898

4.  Technical Note: Consistency of PTW30013 and FC65-G ion chamber magnetic field correction factors.

Authors:  S J Woodings; B van Asselen; T L van Soest; L A de Prez; J J W Lagendijk; B W Raaymakers; J W H Wolthaus
Journal:  Med Phys       Date:  2019-06-17       Impact factor: 4.071

5.  Characteristics of the Exradin W1 scintillator in the magnetic field.

Authors:  Jeongmin Yoon; Jung-In Kim; Chang Heon Choi; Jong Min Park
Journal:  J Appl Clin Med Phys       Date:  2019-08-28       Impact factor: 2.102

6.  Dosimetric evaluation of irradiation geometry and potential air gaps in an acrylic miniphantom used for external audit of absolute dose calibration for a hybrid 1.5 T MR-linac system.

Authors:  Neelam Tyagi; Ergys Subashi; Dale Michael Lovelock; Stephen Kry; Paola Elisa Alvarez; Margie A Hunt; Seng Boh Lim
Journal:  J Appl Clin Med Phys       Date:  2021-12-16       Impact factor: 2.102

7.  Commissioning a secondary dose calculation software for a 0.35 T MR-linac.

Authors:  Alex T Price; Nels C Knutson; Taeho Kim; Olga L Green
Journal:  J Appl Clin Med Phys       Date:  2022-02-15       Impact factor: 2.102

8.  Commissioning measurements on an Elekta Unity MR-Linac.

Authors:  Marcus Powers; John Baines; Robert Crane; Chantelle Fisher; Stephen Gibson; Linda Marsh; Bronwyn Oar; Ariadne Shoobridge; Emily Simpson-Page; Marchant Van der Walt; Glenn de Vine
Journal:  Phys Eng Sci Med       Date:  2022-03-02
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.