A novel radiation protection device based on tungsten functional paper for application in interventional radiology.

Hajime Monzen1, Mikoto Tamura1,2, Kohei Shimomura2, Yuichi Onishi3, Shinichi Nakayama3, Takahiro Fujimoto4, Kenji Matsumoto1,2, Kohei Hanaoka1,2, Takeshi Kamomae5.   

Abstract

Tungsten functional paper (TFP), which contains 80% tungsten by weight, has radiation-shielding properties. We investigated the use of TFP for the protection of operators during interventional or therapeutic angiography. The air kerma rate of scattered radiation from a simulated patient was measured, with and without TFP, using a water-equivalent phantom and fixed C-arm fluoroscopy. Measurements were taken at the level of the operator's eye, chest, waist, and knee, with a variable number of TFP sheets used for shielding. A Monte Carlo simulation was also utilized to analyze the dose rate delivered with and without the TFP shielding. In cine mode, when the number of TFP sheets was varied through 1, 2, 3, 5, and 10, the respective reduction in the air kerma rate relative to no TFP shielding was as follows: at eye level, 24.9%, 29.9%, 41.6%, 50.4%, and 56.2%; at chest level, 25.3%, 33.1%, 34.9%, 46.1%, and 44.3%; at waist level, 45.1%, 57.0%, 64.4%, 70.7%, and 75.2%; and at knee level, 2.1%, 2.2%, 2.1%, 2.1%, and 2.1%. In fluoroscopy mode, the respective reduction in the air kerma rate relative to no TFP shielding was as follows: at eye level, 24.8%, 30.3%, 34.8%, 51.1%, and 58.5%; at chest level, 25.8%, 33.4%, 35.5%, 45.2%, and 44.4%; at waist level, 44.6%, 56.8%, 64.7%, 71.7%, and 77.2%; and at knee level, 2.2%, 0.0%, 2.2%, 2.8%, and 2.5%. The TFP paper exhibited good radiation-shielding properties against the scattered radiation encountered in clinical settings, and was shown to have potential application in decreasing the radiation exposure to the operator during interventional radiology.
© 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  fluoroscopy; interventional radiology; radiation protection; scattered radiation; tungsten functional paper

Mesh:

Substances:

Year:  2017        PMID: 28422397      PMCID: PMC5689848          DOI: 10.1002/acm2.12083

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


Introduction

Lead has been employed in radiation protection in a clinical setting for many years; its uses include shielding for operator and patient protection in X‐ray radiography and computed tomography, as well as during electron beam radiotherapy. In addition, radiation‐shielding materials made of lead or lead equivalents are not flexible.1, 2 To overcome these problems, some researchers have explored ways of providing effective X‐ray protection using alternative materials.3, 4, 5, 6 In an attempt to reduce the weight of protection materials, several vendors have developed composite lead‐equivalent materials using mixtures of different elements such as lead, tin, copper, bismuth‐antimony, and yttrium.7, 8 Tungsten functional paper (TFP: Toppan Printing Co., Ltd., Tokyo, Japan), which is the paper with thickness of 0.3 mm and containing 80% tungsten powder by weight, has been developed as a lead free radiation shielding alternative, that has the advantages of paper in that it is easy to cut, fold, and stick onto other materials. These characteristics make it particularly flexible for many radiation‐protection applications, such as in radiation‐shielding surgical drapes, which have been reported based on bismuth and barium.9, 10, 11, 12, 13, 14, 15 In a previous study, we have shown that TFP has radiation‐shielding properties for various peak X‐ray kilovoltages in the range of 60–120 kVp.16 We found that 10 TFP sheets (3.0 mm) could have 0.48 ± 0.02 mm lead equivalent for 100 kVp x‐ray.16 Furthermore, in electron‐based radiotherapy, Fujimoto et al. reported an evaluation of the shielding characteristics of TFP against electron beams with clinically relevant energies and demonstrated that TFP is a useful radiation‐shielding material with potential clinical applications.17 Interventional radiology (IR) is a medical specialty that utilizes fluoroscopic guidance for the diagnosis and treatment of disease, and can involve the administration of relatively large doses of radiation. In IR, radiation exposure to the operator results mainly from scattered radiation from the patient.18, 19 Because the operator must stand close to patient during procedures and thus cannot utilize distance for radiation protection, occupational shielding is especially important. It is difficult for the operator to maintain a large distance of separation from the radiation source, because they must operate the equipment close to the patient. Moreover, during complex operations, a large number of x‐ray fluoroscopy images may be required. For these reasons, in IR, the operator is limited in terms of distance from the radiation source and the exposure time; therefore, shielding is especially important. In IR, both the patient and operator face risks from radiation exposure. Patients are primarily at risk of potential skin effects from prolonged radiation exposure. Operators and staffs are at risk of radiation‐induced cataracts and potentially stochastic long‐term cancer risk.20 These risks are minimized in appropriately controlled conditions; however, radiation hazards for workers conducting IR in inappropriate controlled conditions have been reported.20 One principle of working with radiation is to ensure that the radiation exposure is “as low as reasonably achievable”.21 In order to achieve this, various practical protective measures have been developed to reduce the backscatter radiation from the patient, including ceiling mounted shields and table mounted leaded drapes.22 Chida et al. measured operator air kerma rates using a luminescence dosimeter (Luxel Badge; Landauer Inc., Glenwood, IL, USA) during cardiac catheterization procedures, and found that the use of both the table and the ceiling lead drape reduced the occupational radiation exposure by approximately 42%.23 The purpose of the present study was to investigate and measure the utility of a novel TFP as a new method to reduce the operator's radiation exposure from scattered X‐ray radiation arising from the patient during IR procedures. We measured the reduction in the air kerma rate for the operator using various thicknesses of TFP, and compared measurements and computed simulations using the particle and heavy‐ion transport code system (PHITS) 24 Monte Carlo (MC) simulation. The physical measurement of the air kerma rate from a phantom was difficult; this was because the air kerma rates for scattered radiation from the phantom were very low.25

Materials and methods

The TFP was constructed from tungsten powder on cellulose‐based paper, with a density of 0.06–0.21 g/cm2 and a thickness of 0.3 mm per sheet. This material has similar characteristics to paper, and can be cut, folded, and glued, so that it can take various forms. Figure 1 shows a schematic diagram of the measurement geometry. A C‐arm digital angiography system (AXIOM‐ArtisdFC: Medical Solutions, Erlangen, Germany) was used with the tube located beneath the table holding water equivalent slabs to simulate the patient. The X‐ray equipment has two modes: fluoroscopy mode and image acquisition (or cine) mode. The X‐ray tube has an added filtration of a 2.5 mm of Al and 0.6 mm of Cu and a 900‐mm separation between the source and the detector. The patient table was in the beam path, and was modeled as 0.86 mm Aluminum equivalent for the Monte Carlo simulations. The field of view was 180 × 180 mm2, and the detector entrance dose rates were 32 nGy/pulse in fluoroscopy mode (with 15 frames s−1 and a pulse width of 3.6 ms) and 0.2 μGy/frame in cine mode (with 15 frames s−1 and a pulse width of 6.2 ms), which were measured using an area detector of an ionization chamber (Diamentor K2S; PTW, Freiburg, Germany) during IR automatically. Table 1 lists the imaging parameters used in this study.
Figure 1

A schematic diagram of the measurement geometry for scattered radiation from the water equivalent slab phantom. The TFP was taped to the operator side of the phantom and the number of TFP sheets was varied through 1, 2, 3, 5, and 10.

Table 1

Imaging parameters for the C‐arm digital angiography system (AXIOM‐ArtisdFC: Medical Solutions, Erlangen, Germany)

ModeFluoroscopyCine
Pulse/frame rate (s−1)1515
Pulse width (s)3.6 × 10‐3 6.2 × 10‐3
Tube voltage (kV)10066
Tube current (mA)144333
ED (mGy pulse−1 or mGy frame−1)32 × 10‐6 0.2 × 10‐3
DDO (%)5050

ED, detector entrance dose; DDO, dynamic density optimization.

This is a post‐processing parameter that functions in real‐time by harmonizing the distribution of gray steps in the image.

A schematic diagram of the measurement geometry for scattered radiation from the water equivalent slab phantom. The TFP was taped to the operator side of the phantom and the number of TFP sheets was varied through 1, 2, 3, 5, and 10. Imaging parameters for the C‐arm digital angiography system (AXIOM‐ArtisdFC: Medical Solutions, Erlangen, Germany) ED, detector entrance dose; DDO, dynamic density optimization. This is a post‐processing parameter that functions in real‐time by harmonizing the distribution of gray steps in the image. Four water equivalent slabs (TM phantom: Taisei Medical Inc. Kyoto, Japan) were placed on the table to simulate a typical patient. Each slab had dimensions of 300 mm (length) × 300 mm (width) × 50 mm (height), for a total phantom thickness of 200 mm. The survey meter of a solid state detector (Unfors Xi: Unfors Billdal, Sweden) shaped like a lollipop was used to measure the scattered radiation. This device is useful to measure the low level radiation such as scattered radiation from x‐ray tubes or examination rooms. We measured the air kerma rates and the locations of the measurements were at heights corresponding to the eyes (1600 mm from the floor), chest (1300 mm), waist (1000 mm) and knee (600 mm) of the operator; these values were based on Japanese male average height. The statistical errors were standard deviations estimated from five times measurements at each height. The TFP was cut into 300 × 200 mm2 strips and taped to the operator side of the TM phantom, and scattered radiation was measured with and without the TFP. The number of TFP sheets were varied through 1, 2, 3, 5, and 10, and the measurements were repeated twice. Data regression was performed using the Igor Pro (Version 6.33J: WaveMetrics, Lake Oswego, OR, USA) software package, and measurements were compared with the MC simulations because it was difficult to measure the air kerma rate for the operator during angiography, the air kerma rate for scattered radiation from the phantom were very low. We used the particle and heavy‐ion transport code system (PHITS; version 2.64: Japan Atomic Energy Agency, Japan), which has been used in many fields related to particle and heavy ion transport, including accelerator technology, radiotherapy, space radiation, and radiation protection.19 The geometry as shown in Fig. 1 was modeled with PHITS, with the patient table modeled as aluminum of a 0.86 mm thickness. The X‐ray spectrum was calculated using the Birch formula (28), with cut‐off kinetic energies of 1 keV for photons and 66 keV for electrons. The measured dose distribution at each height was simulated over a volume of 20.0 (length) × 10.0 (width) × 10.0 (height) mm3. The simulation domain was discretized into 180 points in the vertical direction (giving a grid size of 1.0 cm from the floor to a height of 180.0 cm). The number of layers of TFP was varied through 0, 1, 2, 3, 5, and 10. The statistical uncertainties in the PHITS calculations were within 3% at waist level.

Results

Table 2 lists the results of measurements involving 0, 1, 2, 3, 5, and 10 sheets of TFP. The measured air kerma rates were decreased when an increasing number of TFP sheets were used for the eye, chest, and waist protection. However, at the knee level the measured dose varied little as a function of the number of TFP sheets. The trends were similar using fluoroscopy mode and cine mode.
Table 2

Measured air kerma rates using different numbers of TFP sheets

ModeFluoroscopy (μGy/min)Cine (μGy/min)
Number of TFP sheets01235100123510
Eye level4.53.43.12.92.21.930.122.621.117.514.913.2
Chest level14.710.99.89.58.08.298.573.665.964.153.254.9
Waist level33.418.514.411.89.57.6235.4137.7101.283.968.958.4
Knee level36.335.536.835.535.335.4240.4235.4235.1235.4235.4235.3
Measured air kerma rates using different numbers of TFP sheets Figure 2 shows the air kerma rate as a function of the number of TFP sheets at eye, chest, waist, and knee level. Figure 3 shows the fitted curves for the percent reduction in rate at each level calculated from Fig. 2 data using Igor Pro software. With five sheets of TFP the rate reduction in the air kerma rates relative to those measured with no TFP shielding using cine mode were 50.4%, 46.1%, 70.7%, and 2.1% at eye, chest, waist, and knee levels, respectively. Using fluoroscopy mode, the respective rate reduction in the air kerma rates as compared with no TFP shielding were 51.1%, 45.2%, 71.7%, and 2.8% at eye, chest, waist, and knee levels, respectively. Figure 4(a) shows a comparison between the calculations using PHITS and the measured data for the relative air kerma distributions at operator height without TFP shielding, and Fig. 4(b) shows the relative air kerma distributions using five sheets of TFP. The relative air kerma distributions at operator height were normalized to the value at waist level for each distribution. The MC simulations were in close agreement with our measurements, which the differences for each average were within 3% (Fig. 4).
Figure 2

The air kerma rates as a function of the number of TFP sheets at eye‐, chest‐, waist‐, and knee‐level measurements.

Figure 3

The measured percentage reduction in the radiation dose as a function of the number of layers of TFP, with curve fitting. (a) At eye level, (b) at chest level, (c) at waist level, and (d) at knee level.

Figure 4

The relative air kerma rate distributions normalized to the waist level values as a function of height, obtained from measurements and PHITS simulations. (a) Without TFP and (b) with five TFP sheets.

The air kerma rates as a function of the number of TFP sheets at eye‐, chest‐, waist‐, and knee‐level measurements. The measured percentage reduction in the radiation dose as a function of the number of layers of TFP, with curve fitting. (a) At eye level, (b) at chest level, (c) at waist level, and (d) at knee level. The relative air kerma rate distributions normalized to the waist level values as a function of height, obtained from measurements and PHITS simulations. (a) Without TFP and (b) with five TFP sheets.

Discussion

We evaluated the radiation‐shielding ability of TFP regarding scattered radiation from a TM phantom. Although the statistical uncertainties of our measurements were approximately 15%, their accuracy was verified by comparing them with MC simulations because the air kerma rate measurements are difficult for all locations and thicknesses of TFP as shown Fig. 4. TFP was effective in shielding the operator from scattered radiation during IR. Using only one TFP sheet it was possible to reduce scattered radiation from the TM phantom by > 40%. When the number of TFP sheets was increased from 5 (1.5 mm) to 10 (3.0 mm), a further reduction of only 10% was measured, indicating that five TFP sheets exhibited saturated shielding ability concerning the lateral scattered radiation from the TM phantom. The reduction in the dose rate at chest and eye level was smaller than that at waist level, because the direct scatter attenuated by the TFP was less due to scatter angle. In addition, at knee level, the air kerma rate could not be decreased by the TFP sheets placed at the side of the phantom body, because the scattered radiation from direct irradiation did not pass the TFP sheets. Further development of TFP for applications such as surgical drapes will require additional modifications, including single‐sided water or blood absorption layers, and a waterproofing layer on the rear. We have already attempted to create waterproof layers on TFP sheets. The simplest method is to laminate water‐repellent material to one or both sides of the TFP sheet. TFP has the potential to be a practical substitute for existing protective devices such as lead drapes for the treatment table and ceiling, surgical drapes, and to serve as a general‐purpose radiation‐shielding material.

Conclusion

We have described measurements and simulations of the radiation‐shielding ability of TFP for IR, especially concerning scattered radiation. We expect that TFP will play an important role in reducing the level of scattered radiation more flexibly than lead, because TFP is easy to cut, paste, and roll as a result of its paper‐like properties.

conflict of Interest

Hajime Monzen has a consultancy agreement with, and financial interest in, TOPPAN PRINTING CO., LTD, Tokyo.
  22 in total

1.  Operator shielding: how and why.

Authors:  Beth A Schueler
Journal:  Tech Vasc Interv Radiol       Date:  2010-09

2.  Radiation attenuation by lead and nonlead materials used in radiation shielding garments.

Authors:  J P McCaffrey; H Shen; B Downton; E Mainegra-Hing
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

3.  Optimizing non-Pb radiation shielding materials using bilayers.

Authors:  J P McCaffrey; E Mainegra-Hing; H Shen
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

4.  Dosimetric shield evaluation with tungsten sheet in 4, 6, and 9MeV electron beams.

Authors:  Takahiro Fujimoto; Hajime Monzen; Manabu Nakata; Takashi Okada; Shinsuke Yano; Toru Takakura; Junichi Kuwahara; Makoto Sasaki; Kyoji Higashimura; Masahiro Hiraoka
Journal:  Phys Med       Date:  2014-06-19       Impact factor: 2.685

5.  Computation of bremsstrahlung X-ray spectra and comparison with spectra measured with a Ge(Li) detector.

Authors:  R Birch; M Marshall
Journal:  Phys Med Biol       Date:  1979-05       Impact factor: 3.609

6.  A novel radiation protection drape reduces radiation exposure during fluoroscopy guided electrophysiology procedures.

Authors:  Joseph J Germano; Gina Day; David Gregorious; Venkataraman Natarajan; Todd Cohen
Journal:  J Invasive Cardiol       Date:  2005-09       Impact factor: 2.022

7.  Reduction of scatter radiation during transradial percutaneous coronary angiography: a randomized trial using a lead-free radiation shield.

Authors:  Luigi Politi; Giuseppe Biondi-Zoccai; Luca Nocetti; Tiziana Costi; Daniel Monopoli; Rosario Rossi; Fabio Sgura; Maria Grazia Modena; Giuseppe M Sangiorgi
Journal:  Catheter Cardiovasc Interv       Date:  2011-04-25       Impact factor: 2.692

8.  Reduction in operator radiation exposure during transradial catheterization and intervention using a simple lead drape.

Authors:  Ali F Iqtidar; Cathy Jeon; Richard Rothman; Randall Snead; Christopher T Pyne
Journal:  Am Heart J       Date:  2012-11-13       Impact factor: 4.749

Review 9.  Radiation injuries after fluoroscopic procedures.

Authors:  Fred A Mettler; Titus R Koenig; Louis K Wagner; Charles A Kelsey
Journal:  Semin Ultrasound CT MR       Date:  2002-10       Impact factor: 1.875

10.  Radiation Dose Reduction during Radial Cardiac Catheterization: Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape.

Authors:  Andrew Ertel; Jeffrey Nadelson; Adhir R Shroff; Ranya Sweis; Dean Ferrera; Mladen I Vidovich
Journal:  ISRN Cardiol       Date:  2012-09-04
View more
  2 in total

1.  Dosimetric evaluation of skin collimation with tungsten rubber for electron radiotherapy: A Monte Carlo study.

Authors:  Kazuki Wakabayashi; Hajime Monzen; Mikoto Tamura; Kenji Matsumoto; Yoshiki Takei; Yasumasa Nishimura
Journal:  J Appl Clin Med Phys       Date:  2021-02-26       Impact factor: 2.102

2.  Estimating radiation exposure of the brain of a physician with a protective flap in interventional radiology: A phantom study.

Authors:  Shota Hattori; Hajime Monzen; Mikoto Tamura; Hiroyuki Kosaka; Yasunori Nakamura; Yasumasa Nishimura
Journal:  J Appl Clin Med Phys       Date:  2022-01-19       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.