Literature DB >> 16145234

A novel radiation protection drape reduces radiation exposure during fluoroscopy guided electrophysiology procedures.

Joseph J Germano1, Gina Day, David Gregorious, Venkataraman Natarajan, Todd Cohen.   

Abstract

OBJECTIVE: The purpose of this study was to evaluate a novel disposable lead-free radiation protection drape for decreasing radiation scatter during electrophysiology procedures.
BACKGROUND: In recent years, there has been an exponential increase in the number of electrophysiology (EP) procedures exposing patients, operators and laboratory staff to higher radiation doses.
METHODS: The RADPAD was positioned slightly lateral to the incision site for pectoral device implants and superior to the femoral vein during electrophysiology studies. Each patient served as their own control and dosimetric measurements were obtained at the examiner's elbow and hand. Radiation badge readings for the operator were obtained three months prior to RADPAD use and three months after introduction.
RESULTS: Radiation dosimetry was obtained in twenty patients: 7 electrophysiology studies, 6 pacemakers, 5 catheter ablations, and 2 implantable cardioverter-defibrillators. Eleven women and nine men with a mean age of 63 +/- 4 years had an average fluoroscopy time of 2.5 +/- 0.42 minutes per case. Mean dosimetric measurements at the hand were reduced from 141.38 +/- 24.67 to 48.63 +/- 9.02 milliroentgen (mR) per hour using the protective drape (63% reduction; p < 0.0001). Measurements at the elbow were reduced from 78.78 +/- 7.95 mR per hour to 34.50 +/- 4.18 mR per hour using the drape (55% reduction; p < 0.0001). Badge readings for three months prior to drape introduction averaged 2.45 mR per procedure versus 1.54 mR per procedure for 3 months post-initiation (37% reduction).
CONCLUSION: The use of a novel radiation protection surgical drape can significantly reduce scatter radiation exposure to staff and operators during a variety of EP procedures.

Entities:  

Mesh:

Year:  2005        PMID: 16145234

Source DB:  PubMed          Journal:  J Invasive Cardiol        ISSN: 1042-3931            Impact factor:   2.022


  9 in total

1.  A double-blind, randomized, sham-controlled trial of the effect of a radiation-attenuating drape on radiation exposure to endoscopy staff during ERCP.

Authors:  Thiruvengadam Muniraj; Harry R Aslanian; Loren Laine; James Farrell; Maria M Ciarleglio; Yanhong Deng; Henry Ho; Priya A Jamidar
Journal:  Am J Gastroenterol       Date:  2015-03-31       Impact factor: 10.864

2.  The Efficacy of Shielding Systems for Reducing Operator Exposure during Neurointerventional Procedures: A Real-World Prospective Study.

Authors:  T R Miller; J Zhuo; G Jindal; R Shivashankar; N Beaty; D Gandhi
Journal:  AJNR Am J Neuroradiol       Date:  2016-12-22       Impact factor: 3.825

3.  Radiation Exposure During Transcatheter Valve Replacement: What Cardiac Surgeons Need to Know.

Authors:  Alejandro Aquino; Ali J Khiabani; Matthew C Henn; Alan Zajarias; Spencer J Melby; Marc Sintek; John Lasala; Puja Kachroo; Eric Novak; Hersh S Maniar
Journal:  Ann Thorac Surg       Date:  2019-07-06       Impact factor: 4.330

4.  Efficacy of RADPAD protective drape during coronary angiography.

Authors:  B Kherad; T Jerichow; F Blaschke; M Noutsias; B Pieske; C Tschöpe; F Krackhardt
Journal:  Herz       Date:  2017-04-07       Impact factor: 1.443

Review 5.  Heart Rythm Society expert consensus statement on electrophysiology laboratory standards: process, protocols, equipment, personnel, and safety.

Authors:  David E Haines; Salwa Beheiry; Joseph G Akar; Janice L Baker; Doug Beinborn; John F Beshai; Neil Brysiewicz; Christine Chiu-Man; Kathryn K Collins; Matthew Dare; Kenneth Fetterly; John D Fisher; Richard Hongo; Samuel Irefin; John Lopez; John M Miller; James C Perry; David J Slotwiner; Gery F Tomassoni; Esther Weiss
Journal:  Heart Rhythm       Date:  2014-05-07       Impact factor: 6.343

Review 6.  The Growing Culture Of A Minimally Fluoroscopic Approach In Electrophysiology Lab.

Authors:  Michela Casella; Eleonora Russo; Francesca Pizzamiglio; Sergio Conti; Ghaliah Al-Mohani; Daniele Colombo; Victor Casula; Yuri D Alessandra; Viviana Biagioli; Corrado Carbucicchio; Stefania Riva; Gaetano Fassini; Massimo Moltrasio; Fabrizio Tundo; Martina Zucchetti; Benedetta Majocchi; Vittoria Marino; Giovanni Forleo; Pasquale Santangeli; Luigi Di Biase; Antonio Dello Russo; Andrea Natale; Claudio Tondo
Journal:  J Atr Fibrillation       Date:  2014-08-31

Review 7.  [Update on radiation exposure in catheter ablation of atrial fibrillation].

Authors:  Tobias Schreiber; Nora Kähler; Verena Tscholl; Patrick Nagel; Ulf Landmesser; Martin Huemer; Philipp Attanasio
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2019-11-22

8.  A novel radiation protection device based on tungsten functional paper for application in interventional radiology.

Authors:  Hajime Monzen; Mikoto Tamura; Kohei Shimomura; Yuichi Onishi; Shinichi Nakayama; Takahiro Fujimoto; Kenji Matsumoto; Kohei Hanaoka; Takeshi Kamomae
Journal:  J Appl Clin Med Phys       Date:  2017-04-19       Impact factor: 2.102

9.  Results from a real-time dosimetry study during left atrial ablations performed with ultra-low dose radiation settings.

Authors:  T Schreiber; N Kähler; S Biewener; V Tscholl; P Nagel; P Attanasio; U Landmesser; M Huemer
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2021-05-11
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.