Optimizing non-Pb radiation shielding materials using bilayers.

J P McCaffrey1, E Mainegra-Hing, H Shen.   

Abstract

PURPOSE: The objective of this study was to demonstrate that the weight of non-Pb radiation shielding materials can be minimized by structuring the material as a bilayer composed of different metal-powder-embedded elastomer layers.
METHODS: Measurements and Monte Carlo (MC) calculations were performed to study the attenuation properties of several non-Pb metal bilayers over the x-ray energy range 30-150 keV. Metals for the layers were chosen on the basis of low cost, nontoxicity, and complementary photoelectric absorption characteristics. The EGSnrc user code cavity.cpp was used to calculate the resultant x-ray fluence spectra after attenuation by these metal layers. Air kerma attenuation was measured using commercially manufactured metal/elastomer test layers. These layers were irradiated using the primary standard calibration beams at the Institute for National Measurement Standards in Ottawa, Canada utilizing the six x-ray beam qualities recommended in the German Standard DIN 6857. Both the measurements and the calculations were designed to approximate surface irradiation as well as penetrating radiation at 10 mm depth in soft tissue. The MC modeling point and the position of the measurement detector for surface irradiation were both directly against the downstream face of the attenuating material, as recommended in DIN 6857.
RESULTS: The low-Z upstream/high-Z downstream ordering of the metal bilayers provided substantially more attenuation than the reverse order. Optimal percentages of each metal in each bilayer were determined for each x-ray radiation beam quality.
CONCLUSIONS: Depending on the x-ray quality, appropriate choices of two complementary metal-embedded elastomer layers can decrease the weight of radiation shielding garments by up to 25% compared to Pb-based elastomer garments while providing equivalent attenuation.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20095271     DOI: 10.1118/1.3260839

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  On the (f)utility of measuring the lead equivalence of protective garments.

Authors:  A Kyle Jones; Louis K Wagner
Journal:  Med Phys       Date:  2013-06       Impact factor: 4.071

2.  Lead Free Multilayered Polymer Composites for Radiation Shielding.

Authors:  Laurynas Gilys; Egidijus Griškonis; Paulius Griškevičius; Diana Adlienė
Journal:  Polymers (Basel)       Date:  2022-04-21       Impact factor: 4.967

3.  Assessment of the Radiation Attenuation Properties of Several Lead Free Composites by Monte Carlo Simulation.

Authors:  M Kazempour; M Saeedimoghadam; F Shekoohi Shooli; N Shokrpour
Journal:  J Biomed Phys Eng       Date:  2015-06-01

4.  A novel radiation protection device based on tungsten functional paper for application in interventional radiology.

Authors:  Hajime Monzen; Mikoto Tamura; Kohei Shimomura; Yuichi Onishi; Shinichi Nakayama; Takahiro Fujimoto; Kenji Matsumoto; Kohei Hanaoka; Takeshi Kamomae
Journal:  J Appl Clin Med Phys       Date:  2017-04-19       Impact factor: 2.102

5.  Estimation of the shielding ability of a tungsten functional paper for diagnostic x-rays and gamma rays.

Authors:  Hajime Monzen; Ikuo Kanno; Takahiro Fujimoto; Masahiro Hiraoka
Journal:  J Appl Clin Med Phys       Date:  2017-06-28       Impact factor: 2.102

6.  Early-Stage Growth Mechanism and Synthesis Conditions-Dependent Morphology of Nanocrystalline Bi Films Electrodeposited from Perchlorate Electrolyte.

Authors:  Daria Tishkevich; Sergey Grabchikov; Tatiana Zubar; Denis Vasin; Sergei Trukhanov; Alla Vorobjova; Dmitry Yakimchuk; Artem Kozlovskiy; Maxim Zdorovets; Sholpan Giniyatova; Dmitriy Shimanovich; Dmitry Lyakhov; Dominik Michels; Mengge Dong; Svetlana Gudkova; Alex Trukhanov
Journal:  Nanomaterials (Basel)       Date:  2020-06-27       Impact factor: 5.076

7.  Calculation of Attenuation Parameter for Ir-192 Gamma Source in Shielding Materials.

Authors:  Adila Hanim Aminordin Sabri; M Z Abdul Aziz; S F Olukotun; S M Tajudin
Journal:  J Med Phys       Date:  2022-03-31

8.  A simple quality control tool for assessing integrity of lead equivalent aprons.

Authors:  Roshan S Livingstone; Anna Varghese
Journal:  Indian J Radiol Imaging       Date:  2018 Apr-Jun
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.