Raphaela Avgousti1, Christina Armpilia1, Ioannis Floros2, Christos Antypas1.
Abstract
The current work describes the implementation of the American Association of Physicists in Medicine (AAPM)'s Task Group 119 report on a volumetric phantom (Delta4, Scandidos, Uppsala, Sweden) following the stated dose goals, to evaluate the step-and-shoot intensity modulated radiation therapy (IMRT) system. Delta4 consists of diode detectors, lying on two crossed planes, measuring the delivered dose, and providing two-dimensional dosimetric information. Seven plans of different goals and complexity were performed, with individual structure sets. TG199 structure sets and plans were transferred and implemented on the Delta4 phantom taking into account its cylindrical geometry. All plans were delivered with a 6 MV linear accelerator equipped with multileaf collimator of 1 cm thickness. Plan results for each test met the recommended dose goals. The evaluation was performed in terms of dose deviation, distance to agreement, and gamma index passing rate. In all test cases, the gamma index passing rate was measured >90%. Delta4 phantom has proven to be fast, applicable, and reliable for the step-and-shoot IMRT commissioning following TG119's recommended tests. Although AAPM's TG119 report is referring to the implementation of test plans that do not correspond to patient plans, it could be used as an evaluation tool of various IMRT systems, considering the local treatment planning system and the delivery system.Entities:
Keywords: Commissioning; intensity modulated radiation therapy; quality assurance; volumetric phantom
Year: 2017 PMID: 28405106 PMCID: PMC5370336 DOI: 10.4103/0971-6203.202419
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Treatment planning statistics for multitarget
Treatment planning statistics for C-shape (harder)
Treatment planning statistics for mock prostate
Treatment planning statistics for mock head/neck
Treatment planning statistics for mock C-shape (easier)
Results of dose deviation, distance to agreement, and gamma index for all the test cases
Figure 1Dose distribution, dose deviation, distance to agreement, and gamma index of test case I2: Prostate.
Figure 2Dose distribution, dose deviation, distance to agreement, and gamma index of test case I3: Head/neck.
Figure 3Dose distribution, dose deviation, distance to agreement, and gamma index of test case I2: C-shape (easier).