Temporal Artery Temperature Measurement in the Neonate.

Mashette E Syrkin-Nikolau1, Karen J Johnson1, Tarah T Colaizy1, Ruthann Schrock1, Edward F Bell1.   

Abstract

Objective We compared an infrared temporal artery thermometer with our clinical standard axillary thermometer for temperature measurements in neonatal patients. Study Design We measured temporal artery (Tta), axillary (Tax, clinical standard), and rectal (Tr, gold standard) temperatures of 49 infants. The difference between Tr and Tta was compared with that between Tr and Tax, and the data were analyzed based on bed type and postmenstrual age. Results The mean Tta, Tax, and Tr were 37.16 (SD 0.36) °C, 36.61 (SD 0.30) °C, and 36.82 (SD 0.30) °C, respectively. The measurements by these methods were all significantly different. The mean Tr-Tax was 0.21 (SD 0.26) °C, and the mean Tr-Tta was -0.34 (SD 0.37) °C, indicating that Tax was closer to Tr than was Tta (p < 0.0001). Tta agreed more closely with Tr for infants in cribs than for those in incubators. Adjusting for bed type and body weight, with each week of postmenstrual age, the discrepancy between Tr-Tta and Tr-Tax decreased by 0.005°C (p = 0.034). Conclusion Compared with the gold standard, Tr, Tta is not more accurate than Tax. The temporal artery thermometer was less accurate for infants in incubators than for infants in cribs. The accuracy of temporal artery temperature increased with postmenstrual age. Thieme Medical Publishers, Inc. 333 Seventh Avenue, New York, NY 10001, USA.

Entities:  

Mesh:

Year:  2017        PMID: 28395367      PMCID: PMC5532079          DOI: 10.1055/s-0037-1601440

Source DB:  PubMed          Journal:  Am J Perinatol        ISSN: 0735-1631            Impact factor:   1.862


  18 in total

1.  ACCIDENTAL PERFORATION OF THE COLON AND RECTUM IN NEWBORN INFANTS.

Authors:  E W FONKALSRUD; H W CLATWORTHY
Journal:  N Engl J Med       Date:  1965-05-27       Impact factor: 91.245

2.  Infrared thermometry of newborn infants.

Authors:  K J Johnson; P Bhatia; E F Bell
Journal:  Pediatrics       Date:  1991-01       Impact factor: 7.124

3.  A practical comparison of temporal artery thermometry and axillary thermometry in neonates under different environments.

Authors:  Ming Ann Sim; Syen Yee Leow; Ying Hao; Cheo Lian Yeo
Journal:  J Paediatr Child Health       Date:  2016-04       Impact factor: 1.954

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Thermal features of the female axilla.

Authors:  R M Abrams; J P Royston; S J Humphrey; H S Wolff
Journal:  J Appl Physiol Respir Environ Exerc Physiol       Date:  1980-12

Review 6.  Infrared ear thermometry compared with rectal thermometry in children: a systematic review.

Authors:  Jean V Craig; Gillian A Lancaster; Stephen Taylor; Paula R Williamson; Rosalind L Smyth
Journal:  Lancet       Date:  2002-08-24       Impact factor: 79.321

7.  Accuracy of infrared thermometers in very low birth weight infants and impact on newborn behavioural states.

Authors:  Melanie Jarvis; Katelyn J Guy; Kai König
Journal:  J Paediatr Child Health       Date:  2013-04-25       Impact factor: 1.954

8.  Comparison of temporal artery, mid-forehead skin and axillary temperature recordings in preterm infants <1500 g of birthweight.

Authors:  Ridvan Duran; Ulfet Vatansever; Betül Acunaş; Necdet Süt
Journal:  J Paediatr Child Health       Date:  2009 Jul-Aug       Impact factor: 1.954

9.  Broken and retained rectal thermometers in infants and young children.

Authors:  J T Lau; G B Ong
Journal:  Aust Paediatr J       Date:  1981-06

Review 10.  Understanding Bland Altman analysis.

Authors:  Davide Giavarina
Journal:  Biochem Med (Zagreb)       Date:  2015-06-05       Impact factor: 2.313

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.