Accuracy of infrared thermometers in very low birth weight infants and impact on newborn behavioural states.

Melanie Jarvis1, Katelyn J Guy, Kai König.   

Abstract

AIM: To study the impact on newborn behavioural states and accuracy of three infrared thermometers compared with digital axillary thermometer measurements in very low birth weight infants.
METHODS: Single-centre prospective observational study. Preterm infants born <1500-g birth weight were eligible. Infants were observed for pre-measurement behaviour state using a five-point neonatal behaviour observation tool. One infrared temperature was taken from each of the devices, followed by an axillary measurement. Further behaviour-state observations were recorded following infrared and axillary measurements.
RESULTS: One hundred measurements were collected from each infrared device among a cohort of 42 very low birth weight infants. Only one infrared device showed satisfactory agreement with bias -0.071 (95% limits of agreement -0.68 to 0.54). The other two devices demonstrated poor agreement: bias -1.34; 95% limits of agreement -2.62 to -0.5 and bias -0.56; 95% limits of agreement -1.38 to 0.25. Neonatal behavioural scores showed only minimal changes when infrared measurements were performed but increased significantly following axillary measurements. The difference between the two modalities was statistically significant with a mean increase of 1.44 points following axillary measurements (95% confidence interval 1.21 to 1.67, P < 0.001).
CONCLUSIONS: Temperature measurements taken with infrared thermometers demonstrated less disruption to preterm infants' behavioural state, however accuracy of devices varied.
© 2013 The Authors. Journal of Paediatrics and Child Health © 2013 Paediatrics and Child Health Division (Royal Australasian College of Physicians).

Entities:  

Mesh:

Year:  2013        PMID: 23614674     DOI: 10.1111/jpc.12207

Source DB:  PubMed          Journal:  J Paediatr Child Health        ISSN: 1034-4810            Impact factor:   1.954


  3 in total

1.  A comparison of mid-forehead and axillary temperatures in newborn intensive care.

Authors:  J Robertson-Smith; F T McCaffrey; R Sayers; S Williams; B J Taylor
Journal:  J Perinatol       Date:  2014-08-14       Impact factor: 2.521

2.  Temporal Artery Temperature Measurement in the Neonate.

Authors:  Mashette E Syrkin-Nikolau; Karen J Johnson; Tarah T Colaizy; Ruthann Schrock; Edward F Bell
Journal:  Am J Perinatol       Date:  2017-04-10       Impact factor: 1.862

3.  Performance of a non-contact infrared thermometer in healthy newborns.

Authors:  Sara Sollai; Carlo Dani; Elettra Berti; Claudia Fancelli; Luisa Galli; Maurizio de Martino; Elena Chiappini
Journal:  BMJ Open       Date:  2016-03-16       Impact factor: 2.692

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.