| Literature DB >> 28393321 |
Hitoshi Imamura1,2, Hitoshi Tabuchi3, Shunsuke Nakakura3, Daisuke Nagasato3, Hiroaki Baba3, Yoshiaki Kiuchi4.
Abstract
PURPOSE: To investigate the usability and the reproducibility of the tear meniscus values via swept-source optical coherence tomography (SS-OCT) and the conventional slit lamp microscope method with a graticule.Entities:
Keywords: Reproducibility; SS-OCT; Slit lamp; Tear meniscus; Usability
Mesh:
Year: 2017 PMID: 28393321 PMCID: PMC5932105 DOI: 10.1007/s10792-017-0517-3
Source DB: PubMed Journal: Int Ophthalmol ISSN: 0165-5701 Impact factor: 2.031
Fig. 1A tear meniscus image obtained by SS-OCT (male, 79 years of age). The horizontal arrow indicates the tear meniscus height (0.288 mm). The area within the manually plotted lines indicates the tear meniscus area (0.029 mm2)
Fig. 2An anterior segment image obtained by the slit lamp method with a graticule (female, 30 years of age). One scale increment was equivalent to approximately 200 µm. In this case, the tear meniscus height was estimated to be 400 µm
Inter-rater ICC (2, 1) and intra-rater ICC (1, 1) reliability in SS-OCT and slit lamp method
| ICC (2, 1) | ICC (1, 1) | ||
|---|---|---|---|
| Rater 1 | Rater 2 | ||
| SS-OCT | |||
| TMH | 0.932 (0.865, 0.966) | 0.981 (0.965, 0.990) | 0.932 (0.881, 0.964) |
| TMA | 0.982 (0.956, 0.992) | 0.990 (0.982, 0.995) | 0.971 (0.948, 0.985) |
| Slit lamp | |||
| TMH | 0.761 (0.558, 0.878) | 0.974 (0.954, 0.987) | 0.853 (0.752, 0.921) |
The values in parentheses indicate 95% confidence intervals
Range and median values of TMH in 3 groups (µm)
| Age groups | Median | Maximum | Minimum |
|---|---|---|---|
| Group 1 | |||
| Slit lamp | 200 | 300 | 120 |
| SS-OCT | 320 | 481 | 187 |
| Group 2 | |||
| Slit lamp | 200 | 300 | 120 |
| SS-OCT | 322 | 601 | 202 |
| Group 3 | |||
| Slit lamp | 200 | 340 | 100 |
| SS-OCT | 339 | 514 | 173 |
Fig. 3Box plots showing the distribution of tear meniscus height (TMH) measured via SS-OCT and slit lamp methods. No significant differences were found among the three subject groups using either method The TMH values measured by the SS-OCT were significantly higher than those measured by slit lamp method in all age groups (P < 0.0001)
Distribution of TMA Values in 3 groups (×102 µm2)
| Age groups | Median | Maximum | Minimum |
|---|---|---|---|
| Group 1 | 238 | 678 | 104 |
| Group 2 | 246 | 551 | 89 |
| Group 3 | 279 | 707 | 97 |