| Literature DB >> 28344603 |
Tanwiwat Jaikuna1, Phatchareewan Khadsiri1, Nisa Chawapun2, Suwit Saekho3, Ekkasit Tharavichitkul2.
Abstract
PURPOSE: To develop an in-house software program that is able to calculate and generate the biological dose distribution and biological dose volume histogram by physical dose conversion using the linear-quadratic-linear (LQL) model.Entities:
Keywords: EQD2 software; LQL model; biological dose
Year: 2017 PMID: 28344603 PMCID: PMC5346611 DOI: 10.5114/jcb.2017.66082
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Fig. 1The linear quadratic (LQ) and the linear-quadraticlinear (LQL) survival curves
Fig. 2The equivalent dose in 2 Gy fraction (EQD2) function in the Isobio software
Fig. 3The equivalent dose in 2 Gy fraction (EQD2) calculation and display function page in the Isobio software
Fig. 4The differentiation of the dose volume histogram in each organ between treatment planning and Computational Environment for Radiotherapy Research (CERR): (A) Pinnacle treatment planning system (TPS) version 9.8 vs. CERR, and (B) Oncentra TPS version 4.3 vs. CERR
Percentage differentiation of dose-volume histogram (DVH) between Computational Environment for Radiotherapy Research (CERR) and Pinnacle version 9.8 for external beam radiation therapy in D95% and D2%, and Oncentra version 4.3 for brachytherapy in D90% and D2cc from cervical cancer treatment plan
| Region of interest | Percentage dose difference | |
|---|---|---|
|
| ||
| D95% | CTV | 0.19 |
| D2% | Bladder | 0.22 |
|
| ||
| D90% | HR-CTV | 3.33 |
| D2cc | Bladder | 0.56 |
EBRT – external beam radiation therapy, BT – brachytherapy, CTV – clinical target volume, PTV – planning target volume
Percentage differentiation of biological effective dose (BED) and equivalent dose in 2 Gy fraction (EQD2) verification from external beam radiation therapy (EBRT) and brachytherapy (BT) between software calculation and manual calculation in cervix, bladder, and rectum for cervical cancer treatment plan
| Organ of interest |
| Cervical (HR-CTV and IR-CTV) | Bladder | Rectum | ||
|---|---|---|---|---|---|---|
| EBRT | BED | Manual | 100 | 64.16 ± 0.08 | 77.21 ± 0.31 | 72.94 ± 1.32 |
| Software | 100 | 64.16 ± 0.08 | 77.21 ± 0.31 | 72.94 ± 1.32 | ||
| % dif. | 0.00 | 0.00 | 0.00 | |||
|
| 0.717 | 0.158 | 0.171 | |||
| EQD2 | Manual | 100 | 53.47 ± 0.06 | 53.45 ± 0.21 | 48.21 ± 0.87 | |
| Software | 100 | 53.47 ± 0.06 | 53.45 ± 0.21 | 48.21 ± 0.87 | ||
| % dif. | 0.00 | 0.00 | 0.00 | |||
|
| 0.820 | 0.095 | 0.593 | |||
| BT | BED | Manual | 100 | 132.55 ± 11.68 | 13.64 ± 1.22 | 14.57 ± 1.25 |
| Software | 100 | 132.55 ± 11.68 | 13.64 ± 1.22 | 14.57 ± 1.25 | ||
| % dif. | 0.00 | 0.00 | 0.00 | |||
|
| 0.185 | 0.319 | 0.066 | |||
| EQD2 | Manual | 100 | 110.46 ± 9.73 | 9.44 ± 0.84 | 9.63 ± 0.83 | |
| Software | 100 | 110.46 ± 9.73 | 9.44 ± 0.84 | 9.63 ± 0.83 | ||
| % dif. | 0.00 | 0.00 | 0.00 | |||
|
| 0.240 | 0.320 | 0.849 | |||
xMB – the average value of manual BED calculation, xSB – the average value of software BED calculation, xMQ – the average value of manual EQD2 calculation, xSQ – the average value of software EQD2 calculation, SE – the standard error
Fig. 5The 2D physical dose distribution of (A) external beam radiation therapy (EBRT) (B). Brachytherapy (BT) and biological dose distribution from physical dose conversion using the linear-quadratic-linear (LQL) model of (C) EBRT and (D) BT
Fig. 6The 3D biological dose distribution from cervical cancer treatment in each organ of (A) external beam radiation therapy and (B) brachytherapy
Fig. 7Dose-volume histogram (DVH) and equivalent dose in 2 Gy fraction volume histogram (EQD2VH) in each organ of cervical cancer treatment planning from (A) external beam radiation therapy and (B) brachytherapy
Equivalent dose in 2 Gy fraction (EQD2) of region of interest in external beam radiation therapy and brachytherapy
| Region of interest | EQD2 (Gy) | |
|---|---|---|
|
| ||
| D95% | PTV | 51.76 |
| D2% | Bladder | 57.38 |
|
| ||
| D90% | HR-CTV | 9.72 |
| D2cc | Bladder | 7.53 |
EQD2 – equivalent dose at 2 Gy, EBRT – external beam radiation therapy, BT – brachytherapy, CTV – clinical target volume, PTV – planning target volume