Juha Koivisto1, Maureen van Eijnatten2, John Ludlow3, Timo Kiljunen4, Xie-Qi Shi5, Jan Wolff6,7.
Abstract
The aim of the study was to estimate and to compare effective doses in the elbow region resulting from four different x-ray imaging modalities. Absorbed organ doses were measured using 11 metal oxide field effect transistor (MOSFET) dosimeters that were placed in a custom-made anthropomorphic elbow RANDO phantom. Examinations were performed using Shimadzu FH-21 HR radiography device, Siemens Sensation Open 24-slice MSCT-device, NewTom 5G CBCT device, and Planmed Verity CBCT device, and the effective doses were calculated according to ICRP 103 recommendations. The effective dose for the conventional radiographic device was 1.5 µSv. The effective dose for the NewTom 5G CBCT ranged between 2.0 and 6.7 µSv, for the Planmed Verity CBCT device 2.6 µSv and for the Siemens Sensation MSCT device 37.4 µSv. Compared with conventional 2D radiography, this study demonstrated a 1.4-4.6 fold increase in effective dose for CBCT and 25-fold dose for standard MSCT protocols. When compared with 3D CBCT protocols, the study showed a 6-19 fold increase in effective dose using a standard MSCT protocol. CBCT devices offer a feasible low-dose alternative for elbow 3D imaging when compared to MSCT.Entities:
Keywords: CBCT; effective dose; elbow; x-ray radiation
Year: 2021 PMID: 33811787 PMCID: PMC8130242 DOI: 10.1002/acm2.13245
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Exposure parameters of conventional radiography device, CBCT and MSCT scanners.
| Shimadzu | Siemens | NewTom 5G | Planmed | ||||
|---|---|---|---|---|---|---|---|
| FH‐21 HR | Sensation | CBCT | Verity | ||||
| AP+LAT | Open | "Hi Res" | "Standard Scan" | CBCT | |||
| Radiography | MSCT | 12 × 8 | 12 × 8 | 15 × 12 | 18 × 16 | 13 × 16 | |
| Potential (kV) | 59 | 120 | 110 | 110 | 110 | 110 | 92 |
| Tube current (mA) | — | 115 | 5.13 | 0.97 | 0.72 | 0.66 | 5 |
| Exposure time (s) | — | 1 | 5.4 | 3.6 | 3.6 | 3.6 | 4.5 |
| Q (mA*s) | 12.6 | 115 | 27.7 | 3.5 | 2.6 | 2.4 | 22.5 |
| CTDIvol (mGy) | — | 11.0 | 4.40 | 1.16 | 1.15 | 1.14 | — |
| DLP (mGy *cm) | — | 179 | 39.6 | 10.5 | 14.7 | 18.2 | — |
| Filtration (mm Al eq.) | 1.5 | 6.8 | 1.4 | 1.4 | 1.4 | 1.4 | 8.1 |
| Focal size (mm) | 1.2 | 0.7 | 0.3 | 0.6 | 0.6 | 0.6 | 0.5 |
| Anode angle (º) | 12º | 9º | 15º | 15º | 15º | 15º | 5º |
| Slice thickness (mm) | — | 2 | — | — | — | — | — |
| Pitch (mm) | — | 0.5 | — | — | — | — | — |
| Voxel H (mm) | — | — | 0.15 | 0.3 | 0.3 | 0.3 | 0.4 |
| Voxel L (mm) | — | — | 0.15 | 0.3 | 0.3 | 0.3 | 0.4 |
| Voxel W (mm) | — | — | 0.15 | 0.3 | 0.3 | 0.3 | 0.4 |
| Scan angle (º) | — | 360º | 360º | 360º | 360º | 360º | 210º |
| Frame number | — | — | 360 | 360 | 360 | 360 | 300 |
| Scan height (mm) | 130 | 130 | 80 | 80 | 120 | 160 | 130 |
| Scan diam. (mm) | 160 | 160 | 120 | 120 | 150 | 180 | 160 |
For NewTom 5G CBCT: mA is automatically adjusted by "SafeBeam™" acquisition technique.
Absorbed organ doses (mGy) of elbow with radiography, MSCT and CBCT devices.
| Dosim. | Organ | Shimadzu | Siemens | NewTom 5G | Planmed | |||
|---|---|---|---|---|---|---|---|---|
| FH‐21 HR | Sensation | CBCT | Verity | |||||
| AP+LAT | Open | "Hi Res" | "Standard Scan" | CBCT | ||||
| Radiography | MSCT | 12 × 8 | 12 × 8 | 15 × 12 | 18 × 16 | 13 × 16 | ||
| 1 | Humerus | 0.3 | 11.5 | 0.3 | 0.1 | 0.5 | 0.5 | 0.3 |
| 2 | Humerus | 0.5 | 17.1 | 2.9 | 0.9 | 0.8 | 1.0 | 1.3 |
| 3 | Triceps brachii | 0.6 | 18.8 | 1.5 | 0.5 | 1.1 | 1.1 | 1.4 |
| 4 | Bicep brachii | 1.1 | 21.6 | 4.9 | 1.3 | 1.1 | 1.3 | 1.4 |
| 5 | Humerus | 0.5 | 16.4 | 3.3 | 1.0 | 0.8 | 0.8 | 1.4 |
| 6 | Cubital nodes/ | 0.9 | 18.5 | 4.2 | 1.3 | 1.5 | 1.2 | 1.3 |
| 7 | Ulna | 0.3 | 15.1 | 3.0 | 1.0 | 0.9 | 0.9 | 1.3 |
| 8 | Ulna | 0.4 | 14.9 | 2.6 | 0.7 | 1.2 | 0.5 | 1.0 |
| 9 | Radius/Flexor c. | 1.1 | 19.3 | 0.5 | 0.2 | 0.6 | 1.0 | 1.1 |
| 10 | Ulna | 0.4 | 13.3 | 0.3 | 0.1 | 0.9 | 0.5 | 0.7 |
| 11 | Ulna | 0.4 | 11.1 | 0.6 | 0.2 | 1.0 | 0.3 | 0.4 |
| Average dose | 0.6 | 16.1 | 2.2 | 0.7 | 1.0 | 0.8 | 1.1 | |
| Absorbed organ dose ratio | 1.0 | 28.4 | 3.9 | 1.2 | 1.7 | 1.5 | 1.9 | |
Dose compared to the average absorbed organ dose of the radiographic device.
Fig. 1A schematic illustration of an anthropomorphic arm phantom with the exposed field of view.
Fig. 2Typical setup of an anthropomorphic elbow phantom placed into the bore of a CBCT device and the MOSFET readers.
MOSFET dosimeter locations in anthropomorphic RANDO elbow phantom.
| Dosimeter no | Layer | Location | Tissue |
|---|---|---|---|
| 1 | 17 | Humerus | Bone Marrow/Bone surface |
| 2 | 16 | Humerus | Bone Marrow/Bone surface |
| 3 | 16 | Triceps brachii | Bone surface/Muscle |
| 4 | 16 | Biceps brachii | Skin/Muscle |
| 5 | 15 | Humerus | Bone marrow/Bone surface |
| 6 | 14 | Cubital nodes/ | Bone marrow/Bone surface/Lymphatic nodes |
| 7 | 14 | Ulna | Bone marrow |
| 8 | 13 | Ulna | Bone marrow/Bone surface |
| 9 | 12 | Radius/Flexor carpi ulnaris | Bone surface/Muscle/Skin |
| 10 | 12 | Ulna | Bone marrow |
| 11 | 11 | Ulna | Bone marrow/Bone surface |
Fig. 3The placement of MOSFET dosimeters in the elbow phantom [AP (a), LAT (b)].
ICRP 103 weighting factors (w), fractions irradiated (f), and and dosimeters used to calculate effective dose.
| Organ | wT | fi | Dosim. No. |
|---|---|---|---|
| Bone marrow | 0.12 | ||
| Ulna | 0.19% | 7, 8, 10, 11 | |
| Radius | 0.21% | 6, 8, 10, 11 | |
| Humerus | 0.60% | 1, 2, 5, 6 | |
| Bone surface | 0.01 | ||
| Ulna | 0.40% | 6, 9 | |
| Radius | 0.39% | 6, 9 | |
| Humerus | 0.93% | 1, 2, 5, 6 | |
| Skin | 0.01 | ||
| Flexor carpi ulnaris | 0.83% | 9 | |
| Extensor carpi radialis longus | 0.80% | 8 | |
| Bicep | 0.81% | 4 | |
| Tricep | 0.42% | 3 | |
| Remainder | 0.12 | ||
| Lymphatic nodes | |||
| Cubital nodes | 5.00% | 6 | |
| Muscle | |||
| Flexor carpi ulnaris | 0.04% | 9 | |
| Extensor carpi radialis longus | 0.05% | 6 | |
| Brachioradialis | 0.18% | 6 | |
| Bicep | 0.22% | 4 | |
| Tricep | 0.40% | 3 | |
Equivalent and effective doses (µSv) for the imaging of elbow with MSCT, CBCT, and radiography devices.
| Tissue | Shimadzu | Siemens | NewTom 5G | Planmed | |||
|---|---|---|---|---|---|---|---|
| FH‐21 HR | Sensation | CBCT | Verity | ||||
| AP+LAT | Open | "Hi Res" | "Standard Scan" | CBCT | |||
| Radiography | MSCT | 12 × 8 | 12 × 8 | 15 × 12 | "18 × 16" | 13 × 16 | |
| Bone marrow | 0.6 | 19.0 | 3.4 | 1.0 | 1.2 | 1.0 | 1.3 |
| Bone surface | 0.1 | 2.7 | 0.4 | 0.1 | 0.2 | 0.1 | 0.2 |
| Skin | 0.2 | 5.5 | 0.8 | 0.2 | 0.3 | 0.3 | 0.4 |
| Remainder | |||||||
| Lymphatic nodes | 0.4 | 8.5 | 2.0 | 0.6 | 0.7 | 0.5 | 0.6 |
| Muscle | 0.1 | 1.6 | 0.2 | 0.1 | 0.1 | 0.1 | 0.1 |
| Effective dose | 1.5 | 37.4 | 6.7 | 2.0 | 2.5 | 2.1 | 2.6 |
| Norm.E (µSv/mAs) | 0.12 | 0.32 | 0.24 | 0.58 | 0.95 | 0.86 | 0.11 |
| Effective dose ratio | 1.0 | 25.4 | 4.6 | 1.4 | 1.7 | 1.4 | 1.8 |
Exposure unit (mAs) normalized effective dose.
Dose compared to the radiographic device.
Fig. 4Anterior‐posterior elbow projections using the Shimadzu FH21‐HR radiography device (a), the Siemens Sensation Open MSCT device (b), the Planmed Verity CBCT device (c) and the NewTom 5G CBCT device HiRes 12 × 8 cm FOV (d), Standard Scan 12 × 8 cm (e), Standard Scan 15 × 12 cm FOV (f), Standard Scan 18 × 16 cm FOV (g).
Fig. 5Exposing the elbow in an upright position using a CT device.
Effective doses (µSv) of the Planmed Verity CBCT using the standard dose kVp and mA exposure values.
| 3.8 mA | 7.5 mA | 12 mA | |
|---|---|---|---|
| 80 kVp | 1.15 | 2.27 | 3.63 |
| 84 kVp | 1.34 | 2.65 | 4.24 |
| 88 kVp | 1.63 | 3.21 | 5.14 |
| 90 kVp | 1.74 | 3.44 | 5.50 |
| 92 kVp | 1.96 | 3.88 | 6.20 |
| 96 kVp | 2.30 | 4.54 | 7.26 |