Diagnosis of small posterior fossa stroke on brain CT: effect of iterative reconstruction designed for brain CT on detection performance.

Taihei Inoue1, Takeshi Nakaura2, Morikatsu Yoshida3, Koichi Yokoyama3, Kenichiro Hirata2, Masafumi Kidoh2, Seitaro Oda2, Daisuke Utsunomiya2, Kazunori Harada4, Yasuyuki Yamashita2.   

Abstract

OBJECTIVES: In this study, we aimed to determine whether iterative model reconstruction designed for brain CT (IMR-neuro) would improve the accuracy of posterior fossa stroke diagnosis on brain CT.
METHODS: We enrolled 37 patients with ischaemic stroke in the posterior fossa and 37 patients without stroke (controls). Using axial images reconstructed using filtered back-projection (FBP) and IMR-neuro, we compared the CT numbers in infarcted areas, image noise in the pons, and contrast-to-noise ratios (CNRs) of infarcted and non-infarcted areas on scans subjected to IMR-neuro and FBP. To analyse the performance of hypo-attenuation detection, we used receiver-operating characteristic (ROC) curve techniques.
RESULTS: The image noise was significantly lower (2.2 ± 0.5 vs. 5.1 ± 0.9 Hounsfield units, p < 0.01) and the difference in CNR between the infarcted and non-infarcted areas was significantly higher with IMR-neuro than with FBP (2.2 ± 1.7 vs. 4.0 ± 3.6, p < 0.01). Furthermore, the average area under the ROC curve was significantly higher with IMR-neuro (0.90 vs. 0.86 for FBP, p = 0.04).
CONCLUSION: IMR-neuro yielded better image quality and improved hypo-attenuation detection in patients with ischaemic stroke. KEY POINTS: • Iterative model reconstruction of brain CT data can facilitate the diagnosis of ischaemic stroke. • IMR improved the detectability of low-contrast lesions in the posterior fossa. • IMR-neuro yielded better image quality and improved observer performance.

Entities:  

Keywords:  Computed tomography; Image processing; Posterior fossa; Radiation dosage; Stroke

Mesh:

Year:  2017        PMID: 28275901     DOI: 10.1007/s00330-017-4773-4

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  22 in total

1.  Cranial CT artifacts and gantry angulation.

Authors:  C Rozeik; O Kotterer; J Preiss; M Schütz; W Dingler; H K Deininger
Journal:  J Comput Assist Tomogr       Date:  1991 May-Jun       Impact factor: 1.826

2.  Comparison of iterative model, hybrid iterative, and filtered back projection reconstruction techniques in low-dose brain CT: impact of thin-slice imaging.

Authors:  Takeshi Nakaura; Yuji Iyama; Masafumi Kidoh; Koichi Yokoyama; Seitaro Oda; Shinichi Tokuyasu; Kazunori Harada; Yasuyuki Yamashita
Journal:  Neuroradiology       Date:  2015-12-29       Impact factor: 2.804

3.  Low-contrast-dose protocol in cardiac CT: 20% contrast dose reduction using 100 kVp and high-tube-current-time setting in 256-slice CT.

Authors:  Masafumi Kidoh; Takeshi Nakaura; Shinichi Nakamura; Shota Nakamura; Naritsugu Sakaino; Kazunori Harada; Shouzaburou Uemura; Yasuyuki Yamashita
Journal:  Acta Radiol       Date:  2013-08-22       Impact factor: 1.990

4.  Reduction of dental metallic artefacts in CT: value of a newly developed algorithm for metal artefact reduction (O-MAR).

Authors:  M Kidoh; T Nakaura; S Nakamura; S Tokuyasu; H Osakabe; K Harada; Y Yamashita
Journal:  Clin Radiol       Date:  2013-10-22       Impact factor: 2.350

5.  A method for simultaneous correction of spectrum hardening artifacts in CT images containing both bone and iodine.

Authors:  P M Joseph; C Ruth
Journal:  Med Phys       Date:  1997-10       Impact factor: 4.071

6.  Low radiation dose protocol in cardiac CT with 100 kVp: usefulness of display preset optimization.

Authors:  Takeshi Nakaura; Masafumi Kidoh; Naritsugu Sakaino; Shota Nakamura; Toshimitsu Nozaki; Akari Izumi; Kazunori Harada; Yasuyuki Yamashita
Journal:  Int J Cardiovasc Imaging       Date:  2013-04-09       Impact factor: 2.357

7.  Statistical limitations in x-ray reconstructive tomography.

Authors:  R A Brooks; G Di Chiro
Journal:  Med Phys       Date:  1976 Jul-Aug       Impact factor: 4.071

8.  Brain tissue water uptake after middle cerebral artery occlusion assessed with CT.

Authors:  Imanuel Dzialowski; Johannes Weber; Arnd Doerfler; Michael Forsting; Rüdiger von Kummer
Journal:  J Neuroimaging       Date:  2004-01       Impact factor: 2.486

9.  Computed tomography depiction of small pediatric vessels with model-based iterative reconstruction.

Authors:  Gonca Koc; Jesse L Courtier; Andrew Phelps; Peter A Marcovici; John D MacKenzie
Journal:  Pediatr Radiol       Date:  2014-02-16

10.  Assessment of a model-based, iterative reconstruction algorithm (MBIR) regarding image quality and dose reduction in liver computed tomography.

Authors:  Won Chang; Jeong Min Lee; Kyunghee Lee; Jeong Hee Yoon; Mi Hye Yu; Joon Koo Han; Byung Ihn Choi
Journal:  Invest Radiol       Date:  2013-08       Impact factor: 6.016

View more
  5 in total

1.  Detection of suspected brain infarctions on CT can be significantly improved with temporal subtraction images.

Authors:  Thai Akasaka; Masahiro Yakami; Mizuho Nishio; Koji Onoue; Gakuto Aoyama; Keita Nakagomi; Yoshio Iizuka; Takeshi Kubo; Yutaka Emoto; Kiyohide Satoh; Hiroyuki Yamamoto; Kaori Togashi
Journal:  Eur Radiol       Date:  2018-07-30       Impact factor: 5.315

2.  Model-Based Iterative Reconstruction (MBIR) for ASPECT Scoring in Acute Stroke Patients Selection: Comparison to rCBV and Follow-Up Imaging.

Authors:  Brieg Dissaux; Mourad Cheddad El Aouni; Julien Ognard; Jean-Christophe Gentric
Journal:  Tomography       Date:  2022-05-05

3.  Accuracy of thin-slice model-based iterative reconstruction designed for brain CT to diagnose acute ischemic stroke in the middle cerebral artery territory: a multicenter study.

Authors:  Hidenori Mitani; Fuminari Tatsugami; Toru Higaki; Yoko Kaichi; Yuko Nakamura; Ewoud Smit; Mathias Prokop; Chiaki Ono; Ken Ono; Yukunori Korogi; Kazuo Awai
Journal:  Neuroradiology       Date:  2021-06-30       Impact factor: 2.804

4.  Ultra-low-dose multiphase CT angiography derived from CT perfusion data in patients with middle cerebral artery stenosis.

Authors:  Xiaoling Wu; Yuelong Yang; Menghuang Wen; Lijuan Wang; Yunjun Yang; Yuhu Zhang; Zihua Mo; Kun Nie; Biao Huang
Journal:  Neuroradiology       Date:  2019-10-30       Impact factor: 2.804

5.  Impact of dose reduction and iterative model reconstruction on multi-detector CT imaging of the brain in patients with suspected ischemic stroke.

Authors:  Karolin J Paprottka; Karina Kupfer; Isabelle Riederer; Claus Zimmer; Meinrad Beer; Peter B Noël; Thomas Baum; Jan S Kirschke; Nico Sollmann
Journal:  Sci Rep       Date:  2021-11-15       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.