An Amperometric Glucose Sensor Integrated into an Insulin Delivery Cannula: In Vitro and In Vivo Evaluation.

W Kenneth Ward1, Gabriel Heinrich1, Matthew Breen1, Sheila Benware1, Nicole Vollum1, Kristin Morris1, Chad Knutsen1, Joseph D Kowalski1, Scott Campbell1, Jerry Biehler1, Mark S Vreeke1, Scott M Vanderwerf1, Jessica R Castle2, Robert S Cargill1.   

Abstract

BACKGROUND: Labeling prohibits delivery of insulin at the site of subcutaneous continuous glucose monitoring (CGM). Integration of the sensing and insulin delivery functions into a single device would likely increase the usage of CGM in persons with type 1 diabetes.
METHODS: To understand the nature of such interference, we measured glucose at the site of bolus insulin delivery in swine using a flexible electrode strip that was laminated to the outer wall of an insulin delivery cannula. In terms of sensing design, we compared H2O2-measuring sensors biased at 600 mV with redox mediator-type sensors biased at 175 mV.
RESULTS: In H2O2-measuring sensors, but not in sensors with redox-mediated chemistry, a spurious rise in current was seen after insulin lis-pro boluses. This prolonged artifact was accompanied by electrode poisoning. In redox-mediated sensors, the patterns of sensor signals acquired during delivery of saline and without any liquid delivery were similar to those acquired during insulin delivery.
CONCLUSION: Considering in vitro and in vivo findings together, it became clear that the mechanism of interference is the oxidation, at high bias potentials, of phenolic preservatives present in insulin formulations. This effect can be avoided by the use of redox mediator chemistry using a low bias potential.

Entities:  

Keywords:  Continuous glucose monitoring; Continuous subcutaneous insulin infusion; Insulin delivery; Insulin delivery cannula.; Type 1 diabetes

Mesh:

Substances:

Year:  2017        PMID: 28221814      PMCID: PMC5399735          DOI: 10.1089/dia.2016.0407

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  22 in total

1.  Electrochemically mediated electrodeposition/electropolymerization to yield a glucose microbiosensor with improved characteristics.

Authors:  Xiaohong Chen; Norio Matsumoto; Yibai Hu; George S Wilson
Journal:  Anal Chem       Date:  2002-01-15       Impact factor: 6.986

2.  Current state of type 1 diabetes treatment in the U.S.: updated data from the T1D Exchange clinic registry.

Authors:  Kellee M Miller; Nicole C Foster; Roy W Beck; Richard M Bergenstal; Stephanie N DuBose; Linda A DiMeglio; David M Maahs; William V Tamborlane
Journal:  Diabetes Care       Date:  2015-06       Impact factor: 19.112

3.  Effects of phenol and meta-cresol depletion on insulin analog stability at physiological temperature.

Authors:  Brandon M Teska; Javier Alarcón; Ronald J Pettis; Theodore W Randolph; John F Carpenter
Journal:  J Pharm Sci       Date:  2014-06-06       Impact factor: 3.534

4.  The Accuracy and Efficacy of the Dexcom G4 Platinum Continuous Glucose Monitoring System.

Authors:  Katherine Nakamura; Andrew Balo
Journal:  J Diabetes Sci Technol       Date:  2015-03-23

5.  Mediated electron transfer in glucose oxidising enzyme electrodes for application to biofuel cells: recent progress and perspectives.

Authors:  Paul Kavanagh; Dónal Leech
Journal:  Phys Chem Chem Phys       Date:  2013-04-14       Impact factor: 3.676

6.  "Wired" enzyme electrodes for amperometric determination of glucose or lactate in the presence of interfering substances.

Authors:  T J Ohara; R Rajagopalan; A Heller
Journal:  Anal Chem       Date:  1994-08-01       Impact factor: 6.986

7.  No apparent local effect of insulin on microdialysis continuous glucose- monitoring measurements.

Authors:  Jeroen Hermanides; Iris M Wentholt; Augustinus A Hart; Joost B Hoekstra; J Hans DeVries
Journal:  Diabetes Care       Date:  2008-03-10       Impact factor: 19.112

Review 8.  Continuous Glucose Monitoring: Impact on Hypoglycemia.

Authors:  Cornelis A J van Beers; J Hans DeVries
Journal:  J Diabetes Sci Technol       Date:  2016-11-01

9.  Glucose microbiosensor based on alumina sol-gel matrix/electropolymerized composite membrane.

Authors:  Xiaohong Chen; Yibai Hu; George S Wilson
Journal:  Biosens Bioelectron       Date:  2002-12       Impact factor: 10.618

10.  Use of the site of subcutaneous insulin administration for the measurement of glucose in patients with type 1 diabetes.

Authors:  Stefan Lindpointner; Stefan Korsatko; Gerd Köhler; Hans Köhler; Roland Schaller; Ruthy Kaidar; Ofer Yodfat; Lukas Schaupp; Martin Ellmerer; Thomas R Pieber; Werner Regittnig
Journal:  Diabetes Care       Date:  2009-12-29       Impact factor: 17.152

View more
  6 in total

1.  Insulin Infusion Sets and Continuous Glucose Monitoring Sensors: Where the Artificial Pancreas Meets the Patient.

Authors:  Gregory P Forlenza
Journal:  Diabetes Technol Ther       Date:  2017-04       Impact factor: 6.118

2.  Artificial Pancreas in Young Children.

Authors:  Rebecca A Ohman-Hanson; Gregory P Forlenza
Journal:  Diabetes Technol Ther       Date:  2017-05       Impact factor: 6.118

3.  A Disposable Tear Glucose Biosensor-Part 5: Improvements in Reagents and Tear Sampling Component.

Authors:  Chi En Lin; Yuka Ito; Anna Deng; Jared Johns; Daniel Matloff; Curtiss B Cook; Koji Sode; Jeffrey T La Belle
Journal:  J Diabetes Sci Technol       Date:  2018-04-18

4.  Measuring glucose at the site of insulin delivery with a redox-mediated sensor.

Authors:  Peter G Jacobs; Nichole S Tyler; Scott M Vanderwerf; Clara Mosquera-Lopez; Thomas Seidl; Robert Cargill; Deborah Branigan; Katrina Ramsey; Kristin Morris; Sheila Benware; W Kenneth Ward; Jessica R Castle
Journal:  Biosens Bioelectron       Date:  2020-04-29       Impact factor: 12.545

Review 5.  Monitoring of Pediatric Type 1 Diabetes.

Authors:  Brynn E Marks; Joseph I Wolfsdorf
Journal:  Front Endocrinol (Lausanne)       Date:  2020-03-17       Impact factor: 5.555

6.  Systematic in vivo evaluation of the time-dependent inflammatory response to steel and Teflon insulin infusion catheters.

Authors:  Jasmin R Hauzenberger; Julia Münzker; Petra Kotzbeck; Martin Asslaber; Vladimir Bubalo; Jeffrey I Joseph; Thomas R Pieber
Journal:  Sci Rep       Date:  2018-01-18       Impact factor: 4.379

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.