Literature DB >> 11811410

Electrochemically mediated electrodeposition/electropolymerization to yield a glucose microbiosensor with improved characteristics.

Xiaohong Chen1, Norio Matsumoto, Yibai Hu, George S Wilson.   

Abstract

A procedure is described that provides for electrochemically mediated deposition of enzyme and a polymer layer permselective for endogenous electroactive species. Electrodeposition was first employed for the direct immobilization of glucose oxidase to produce a uniform, thin, and compact film on a Pt electrode. Electropolymerization of phenol was then employed to form an anti-interference and protective polyphenol film within the enzyme layer. In addition, a stability-reinforcing membrane derived from (3-aminopropyl)trimethoxysilane was constructed by electrochemically assisted cross-linking. This hybrid film outside the enzyme layer contributed to the improved stability and permselectivity. The resulting glucose sensor was characterized by a short response time (<4 s), high sensitivity (1200 nA/mM x cm2), low interference from endogenous electroactive species, and working lifetime of more than 50 days.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11811410     DOI: 10.1021/ac015628m

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  20 in total

Review 1.  Current problems and potential techniques in in vivo glucose monitoring.

Authors:  Y Wickramasinghe; Y Yang; S A Spencer
Journal:  J Fluoresc       Date:  2004-09       Impact factor: 2.217

Review 2.  In Vivo Chemical Sensors: Role of Biocompatibility on Performance and Utility.

Authors:  Robert J Soto; Jackson R Hall; Micah D Brown; James B Taylor; Mark H Schoenfisch
Journal:  Anal Chem       Date:  2016-11-21       Impact factor: 6.986

3.  Molecular extraction in single live cells by sneaking in and out magnetic nanomaterials.

Authors:  Zhen Yang; Liangzi Deng; Yucheng Lan; Xiaoliu Zhang; Zhonghong Gao; Ching-Wu Chu; Dong Cai; Zhifeng Ren
Journal:  Proc Natl Acad Sci U S A       Date:  2014-07-16       Impact factor: 11.205

4.  Edge-plane microwire electrodes for highly sensitive H₂O₂ and glucose detection.

Authors:  Liangliang Qiang; Santhisagar Vaddiraju; Dipesh Patel; Fotios Papadimitrakopoulos
Journal:  Biosens Bioelectron       Date:  2011-02-18       Impact factor: 10.618

5.  Compatibility of Nitric Oxide Release with Implantable Enzymatic Glucose Sensors Based on Osmium (III/II) Mediated Electrochemistry.

Authors:  Kyoung Ha Cha; Mark E Meyerhoff
Journal:  ACS Sens       Date:  2017-08-23       Impact factor: 7.711

6.  Nanomaterials and biomaterials in electrochemical arrays for protein detection.

Authors:  James F Rusling; Gregory W Bishop; Nhi Doan; Fotios Papadimitrakopoulos
Journal:  J Mater Chem B       Date:  2014-01-07       Impact factor: 6.331

7.  Microsphere erosion in outer hydrogel membranes creating macroscopic porosity to counter biofouling-induced sensor degradation.

Authors:  S Vaddiraju; Y Wang; L Qiang; D J Burgess; F Papadimitrakopoulos
Journal:  Anal Chem       Date:  2012-10-05       Impact factor: 6.986

8.  Design Considerations for Silica-Particle-Doped Nitric-Oxide-Releasing Polyurethane Glucose Biosensor Membranes.

Authors:  Robert J Soto; Jonathon B Schofield; Shaylyn E Walter; Maggie J Malone-Povolny; Mark H Schoenfisch
Journal:  ACS Sens       Date:  2016-12-15       Impact factor: 7.711

Review 9.  Biocompatible materials for continuous glucose monitoring devices.

Authors:  Scott P Nichols; Ahyeon Koh; Wesley L Storm; Jae Ho Shin; Mark H Schoenfisch
Journal:  Chem Rev       Date:  2013-02-07       Impact factor: 60.622

10.  Electrochemical sensing via selective surface modification of iridium microelectrodes to create a platinum black interface.

Authors:  Paras R Patel; Matthew D Gibson; Kip A Ludwig; Nicholas B Langhals
Journal:  Int IEEE EMBS Conf Neural Eng       Date:  2013
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.