Automated identification and reduction of artifacts in cine four-dimensional computed tomography (4DCT) images using respiratory motion model.

Min Li1,2, Sarah Joy Castillo3,4, Richard Castillo3,4, Edward Castillo3,5,6, Thomas Guerrero3,5, Liang Xiao7, Xiaolin Zheng8.   

Abstract

PURPOSE: Four-dimensional computed tomography (4DCT) images are often marred by artifacts that substantially degrade image quality and confound image interpretation. Human observation remains the standard method of 4DCT artifact evaluation, which is time-consuming and subjective. We develop a method to automatically identify and reduce artifacts in cine 4DCT images.
METHODS: We proposed an algorithm that consisted of two main stages: deformable image registration and respiratory motion simulation. Specifically, each 4DCT phase image was registered to the breath-holding CT image using the block-matching method, with erroneous spatial matches removed by the least median of squares filter and the full displacement vector field generated by the moving least squares interpolation. The lung's respiratory motion trajectory was then recovered from the displacement vector field using the parameterized polynomial function, with fitting parameters estimated by combinatorial optimization. In this way, artifacts were located according to deviations between image points and their motion trajectories and further corrected based on position prediction.
RESULTS: The mean spatial error (standard deviation) was 1.00 (0.85) mm after registration as opposed to 6.96 (4.61) mm before registration. In addition, we took human observation conducted by medical experts as the gold standard. The average sensitivity, specificity, and accuracy of the proposed method in artifact identification were 0.97, 0.84, and 0.89, respectively.
CONCLUSIONS: The proposed method identified and reduced artifacts accurately and automatically, providing an alternative way to analyze 4DCT image quality and to correct problematic images for radiation therapy.

Entities:  

Keywords:  Artifacts; Deformable image registration (DIR); Four-dimensional computed tomography (4DCT)images; Respiratory motion

Mesh:

Year:  2017        PMID: 28197760     DOI: 10.1007/s11548-017-1538-0

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  35 in total

1.  Use of maximum intensity projections (MIP) for target volume generation in 4DCT scans for lung cancer.

Authors:  René W M Underberg; Frank J Lagerwaard; Ben J Slotman; Johan P Cuijpers; Suresh Senan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-09-01       Impact factor: 7.038

2.  Patient-specific biomechanical model as whole-body CT image registration tool.

Authors:  Mao Li; Karol Miller; Grand Roman Joldes; Barry Doyle; Revanth Reddy Garlapati; Ron Kikinis; Adam Wittek
Journal:  Med Image Anal       Date:  2015-01-30       Impact factor: 8.545

3.  Adaptation and applications of a realistic digital phantom based on patient lung tumor trajectories.

Authors:  Pankaj Mishra; Sara St James; W Paul Segars; Ross I Berbeco; John H Lewis
Journal:  Phys Med Biol       Date:  2012-06-07       Impact factor: 3.609

4.  Real-time respiration monitoring using the radiotherapy treatment beam and four-dimensional computed tomography (4DCT)--a conceptual study.

Authors:  Weiguo Lu; Kenneth J Ruchala; Ming-Li Chen; Quan Chen; Gustavo H Olivera
Journal:  Phys Med Biol       Date:  2006-08-22       Impact factor: 3.609

5.  Characterization and identification of spatial artifacts during 4D-CT imaging.

Authors:  Dongfeng Han; John Bayouth; Sudershan Bhatia; Milan Sonka; Xiaodong Wu
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

6.  Metal artifacts reduction using monochromatic images from spectral CT: evaluation of pedicle screws in patients with scoliosis.

Authors:  Yang Wang; Bangping Qian; Baoxin Li; Guochu Qin; Zhengyang Zhou; Yong Qiu; Xizhao Sun; Bin Zhu
Journal:  Eur J Radiol       Date:  2013-03-19       Impact factor: 3.528

7.  An evaluation of three commercially available metal artifact reduction methods for CT imaging.

Authors:  Jessie Y Huang; James R Kerns; Jessica L Nute; Xinming Liu; Peter A Balter; Francesco C Stingo; David S Followill; Dragan Mirkovic; Rebecca M Howell; Stephen F Kry
Journal:  Phys Med Biol       Date:  2015-01-14       Impact factor: 3.609

8.  On a PCA-based lung motion model.

Authors:  Ruijiang Li; John H Lewis; Xun Jia; Tianyu Zhao; Weifeng Liu; Sara Wuenschel; James Lamb; Deshan Yang; Daniel A Low; Steve B Jiang
Journal:  Phys Med Biol       Date:  2011-08-24       Impact factor: 3.609

9.  Patient-specific finite element modeling of respiratory lung motion using 4D CT image data.

Authors:  René Werner; Jan Ehrhardt; Rainer Schmidt; Heinz Handels
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

10.  Development of a geometry-based respiratory motion-simulating patient model for radiation treatment dosimetry.

Authors:  Juying Zhang; George X Xu; Chengyu Shi; Martin Fuss
Journal:  J Appl Clin Med Phys       Date:  2008-01-21       Impact factor: 2.102