| Literature DB >> 28194362 |
Taha Özyürek1, Koray Yılmaz1, Gülşah Uslu1.
Abstract
OBJECTIVES: The aim of this study was to evaluate the cyclic fatigue resistance of the ProTaper Universal D1 file (Dentsply Maillefer) under continuous and adaptive motion.Entities:
Keywords: Cyclic fatigue; Dynamic model; Endodontics; Nickel titanium; ProTaper Universal Retreatment
Year: 2016 PMID: 28194362 PMCID: PMC5299753 DOI: 10.5395/rde.2017.42.1.34
Source DB: PubMed Journal: Restor Dent Endod ISSN: 2234-7658
Figure 1Dynamic cyclic fatigue testing device.
The mean and the standard deviation values of number of cycles to failure of two groups during dynamic cyclic fatigue test
| Group | NCF | |
|---|---|---|
| Rotary motion | 20 | 810.0 ± 62.0a |
| Adaptive motion | 20 | 1625.8 ± 59.7b |
*Different superscript letters mean statistically significant difference (p < 0.05).
NFC, number of cycles to failure.
Figure 2SEM appearances of the ProTaper Universal retreatment D1 files after cyclic fatigue testing. (a) and (c) General view of D1 instrument with fatigue zones, and an overload fast fracture zone (white arrows); (b) and (d) High-magnification view of D1 instrument showing fatigue striations typical of cyclic fatigue (white arrows). (a) and (b) Rotary motion group; (c) and (d) Adaptive motion group.