Fatigue Life of Reciproc and Mtwo instruments subjected to static and dynamic tests.

Hélio P Lopes1, Carlos N Elias, Márcia V B Vieira, José F Siqueira, Marcelo Mangelli, Weber S P Lopes, Victor T L Vieira, Flávio R F Alves, Julio C M Oliveira, Thaiane G Soares.   

Abstract

INTRODUCTION: This study evaluated the influence of flexibility and reciprocating movement on the fatigue life of endodontic instruments subjected to static and dynamic tests.
METHODS: The rotary nickel-titanium instruments used in this study were Reciproc and Mtwo. The instruments were initially subjected to a cantilever-bending test and then to static and dynamic fatigue tests. Reciproc instruments were operated in reciprocating movement, whereas Mtwo instruments were worked in continuous rotation.
RESULTS: The means of bending resistance (maximum load in grams) of the instruments were 274.9 for Reciproc and 429 for Mtwo. The mean times (in seconds) to fracture of the instruments subjected to static and dynamic tests were 214.5 (static) and 286.3 (dynamic) for Reciproc and 38.9 (static) and 99 (dynamic) for Mtwo. The Student's t test revealed significant differences in all tests (P < .05).
CONCLUSIONS: The results of the present study showed longer fatigue life for instruments with higher flexibility, driven by reciprocating movement, and in the dynamic testing model. These findings reinforce the assumption that use of reciprocating movement is a means to prolong the fatigue life of rotary nickel-titanium endodontic instruments during instrumentation of curved canals.
Copyright © 2013 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23611393     DOI: 10.1016/j.joen.2012.11.048

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  21 in total

1.  Cyclic fatigue of instruments for endodontic glide path.

Authors:  Gianluca Gambarini; Gianluca Plotino; GianPaolo Sannino; Nicola Maria Grande; Alessio Giansiracusa; Lucila Piasecki; Ulisses Xavier da Silva Neto; Dina Al-Sudani; Luca Testarelli
Journal:  Odontology       Date:  2013-11-07       Impact factor: 2.634

2.  A comparative study of cyclic fatigue of 6 endodontic systems. An in vitro study.

Authors:  Jorge Rubio; José-Ignacio Zarzosa; Susana Aranda; Alberto Casino; Antonio Pallarés
Journal:  J Clin Exp Dent       Date:  2022-07-01

3.  Cyclic fatigue resistance of R-Pilot, WaveOne Gold Glider, and ProGlider glide path instruments.

Authors:  Cangül Keskin; Uğur İnan; Murat Demiral; Ali Keleş
Journal:  Clin Oral Investig       Date:  2018-02-17       Impact factor: 3.573

4.  Influence of motion pattern on apical transportation and centering ability of WaveOne single-file technique in curved root canals.

Authors:  Mandana Naseri; Payam Paymanpour; Ali Kangarloo; Sahar Haddadpur; Omid Dianat; Mohammad Ali Ketabi
Journal:  Dent Res J (Isfahan)       Date:  2016 Jan-Feb

5.  Assessment of the role of cross section on fatigue resistance of rotary files when used in reciprocation.

Authors:  Vadhana Sekar; Ranjith Kumar; Suresh Nandini; Suma Ballal; Natanasabapathy Velmurugan
Journal:  Eur J Dent       Date:  2016 Oct-Dec

6.  Effect of adaptive motion on cyclic fatigue resistance of a nickel titanium instrument designed for retreatment.

Authors:  Taha Özyürek; Koray Yılmaz; Gülşah Uslu
Journal:  Restor Dent Endod       Date:  2016-12-19

7.  Bending resistance and cyclic fatigue life of Reciproc Blue, WaveOne Gold, and Genius files in a double (S-shaped) curved canal.

Authors:  Taha Özyürek; Mustafa Gündoğar; Koray Yılmaz; Gülşah Uslu
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2017-12-13

8.  Comparison of Cyclic Fatigue Resistance of Five Nickel Titanium Rotary File Systems with Different Manufacturing Techniques.

Authors:  Mohsen Aminsobhani; Naghmeh Meraji; Ehsan Sadri
Journal:  J Dent (Tehran)       Date:  2015-09

9.  Resistance to torsional failure and cyclic fatigue resistance of ProTaper Next, WaveOne, and Mtwo files in continuous and reciprocating motion: An in vitro study.

Authors:  N O Varghese; Rajesh Pillai; U-Nu Sujathen; Shan Sainudeen; Abe Antony; Sinju Paul
Journal:  J Conserv Dent       Date:  2016 May-Jun

10.  Management of apical periodontitis using WaveOne gold reciprocating files, single-cone endodontic approach: A case series author.

Authors:  Ranjdar Mahmood Talabani
Journal:  Ann Med Surg (Lond)       Date:  2021-05-13
View more