Raag D Airan1,2,3, Randall A Meyer2,4, Nicholas P K Ellens1, Kelly R Rhodes2, Keyvan Farahani1,5, Martin G Pomper1,4,6, Shilpa D Kadam7,8, Jordan J Green2,4,6,9. 1. Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine , Baltimore, Maryland 21231, United States. 2. Department of Biomedical Engineering and the Translational Tissue Engineering Center, Johns Hopkins University School of Medicine , Baltimore, Maryland 21231, United States. 3. Department of Radiology, Stanford University , Stanford, California 94305, United States. 4. Institute for NanoBioTechnology, Johns Hopkins University , Baltimore, Maryland 21231, United States. 5. National Cancer Institute/National Institutes of Health , Bethesda, Maryland 20892, United States. 6. Department of Oncology, Johns Hopkins University School of Medicine , Baltimore, Maryland 21231, United States. 7. Neuroscience Laboratory, Hugo Moser Research Institute, Kennedy Krieger Institute , Baltimore, Maryland 21287, United States. 8. Department of Neurology, Johns Hopkins Medical Institutions , Baltimore, Maryland 21287, United States. 9. Departments of Neurosurgery, Ophthalmology, and Materials Science and Engineering, Johns Hopkins University School of Medicine , Baltimore, Maryland 21231, United States.
Abstract
Targeted, noninvasive neuromodulation of the brain of an otherwise awake subject could revolutionize both basic and clinical neuroscience. Toward this goal, we have developed nanoparticles that allow noninvasive uncaging of a neuromodulatory drug, in this case the small molecule anesthetic propofol, upon the application of focused ultrasound. These nanoparticles are composed of biodegradable and biocompatible constituents and are activated using sonication parameters that are readily achievable by current clinical transcranial focused ultrasound systems. These particles are potent enough that their activation can silence seizures in an acute rat seizure model. Notably, there is no evidence of brain parenchymal damage or blood-brain barrier opening with their use. Further development of these particles promises noninvasive, focal, and image-guided clinical neuromodulation along a variety of pharmacological axes.
Targeted, noninvasive neuromodulation of the brain of an otherwise awake subject could revolutionize both basic and clinical neuroscience. Toward this goal, we have developed nanoparticles that allow noninvasive uncaging of a neuromodulatory drug, in this case the small molecule anesthetic propofol, upon the application of focused ultrasound. These nanoparticles are composed of biodegradable and biocompatible constituents and are activated using sonication parameters that are readily achievable by current clinical transcranial focused ultrasound systems. These particles are potent enough that their activation can silence seizures in an acute ratseizure model. Notably, there is no evidence of brain parenchymal damage or blood-brain barrier opening with their use. Further development of these particles promises noninvasive, focal, and image-guided clinical neuromodulation along a variety of pharmacological axes.
Entities:
Keywords:
Neuromodulation; focused ultrasound; gated drug release; nanoparticles
A long sought after goal of both clinical and basic neuroscience is
the ability to focally modulate the activity of a spatially delimited
region of the brain, noninvasively, and in a safe and reversible manner.[1] Recent advances in MR-guided focused ultrasound
(MRgFUS) suggest that this modality could meet this challenge and
enable clinically translatable neuromodulation.[2−7] However, the mechanism by which focused ultrasound (FUS) may directly
induce changes in neural activity is unknown and is a matter of debate.[6] Additionally, different studies describe divergent
effects of FUS on neural activity with some describing net stimulatory
effects[3] and others describing net inhibitory
effects.[2,5] Despite the excellent robustness and reliability
of focused ultrasound techniques it is unclear how FUS alone impacts
neural activity.We propose an alternate strategy for FUS-mediated
neuromodulation via FUS-gated drug delivery to the brain. This would
combine the predictability of the FUS-induced pressure field with
the robustness of pharmacology. Recent application of FUS for central
nervous system (CNS) drug delivery has enabled advances in the local
delivery of nanoparticle-based therapeutics for varied applications
including glioma treatment,[8] neurological
disorders,[9] and neuroregeneration.[10] Although promising, all of these prior nanoparticle-based
strategies depend on the transient physical opening of the blood-brain
barrier (BBB) via ultrasound-induced cavitation of microbubbles. Additionally,
a recent set of studies has tried to enable robust pharmacological
neuromodulation via FUS-mediated BBB opening.[7] The BBB is a crucial component of the CNS as it maintains the optimal
microenvironment for neuronal activity and protects the neurons from
many endogenous and exogenous neurotoxins that are commonly found
in circulation.[11,12]We therefore focus on delivery
of agents that may readily cross the blood-brain barrier and propose
to use focused ultrasound-mediated drug uncaging from nanoparticle
carriers with the ultrasound focusing providing a limit on the spatial
extent of the drug-based neuromodulation. We then rely upon metabolism
and redistribution of the drug to limit the temporal extent of this
activity. While this limits us to small molecule lipophilic agents
that are known to cross the blood brain barrier passively without
the need for disruption,[13] many if not
most drugs of neurological and psychiatric interest fall under this
umbrella. In practice, after an intravenous infusion of the nanoparticles
inertly labels the blood pool of the subject, FUS application releases
the drug in the vascular bed of the tissue of interest in a region
that is spatially limited by the size of the ultrasound focus. The
drug would then cross the intact blood-brain barrier and act upon
the brain parenchyma during a first-pass of perfusion. Given the availability
of FDA-approved clinical MRgFUS systems that allow noninvasive transcranial
focal sonication of millimeter-sized regions of the brain,[14,15] this strategy could potentially allow focal, noninvasive, and safe
neuromodulation with an immediate path toward clinical translation.We have generated ultrasound-gated nanoparticle carriers of the
small molecule anesthetic propofol. These particles are modified forms
of prior described ultrasound-gated “phase-change” particles
that were originally designed for chemotherapeutic delivery.[16] These particles are made of a biodegradable,
biocompatible polyethylene glycol-b-polycaprolactone
block copolymer matrix encapsulating a liquid perfluorocarbon core
and the drug of interest. Under sonication, the perfluorocarbon core
undergoes a liquid to gas phase transition, thereby releasing the
drug cargo (Figure ). Perfluoropentane was chosen for the perfluorocarbon core given
its relatively high boiling point while encapsulated that would prevent
spontaneous phase change.[17] This amphiphilic
polyester block copolymer was chosen for the emulsifying agent as
polymer perfluorocarbon nanoemulsions have been demonstrated to be
more stable in general than analogous lipid nanoemulsions.[18] We have established the efficacy of drug release
from these particles in vitro (Figure ), as well as the in vivo biodistribution and clearance
kinetics of the nanoparticles (Figure ). As a proof-of-principle, we have further demonstrated
the potency of the nanoparticles to modulate neural activity in vivo
by using them to inducibly silence seizure activity in an acute ratseizure model (Figures ). We have then demonstrated the safety of this technique by observing
no appreciable injury nor BBB opening within the sonicated brain (Figure ). As the components
of these particles have been regarded as safe when utilized in other
clinical applications,[19] these particles
may be able to be readily combined with existent clinical transcranial
MRgFUS systems[14,15] to enable clinical translation.
Overall, this strategy provides a neuromodulation approach that has
an immediate pathway to clinical translation, has a well-defined mechanism
of action via the drug being delivered, does not rely upon invasive
neurosurgery, gene therapy, or a deleterious action upon the brain,
and is generalizable for neuromodulation via any drug that these particles
could encapsulate. Indeed, this approach provides a pathway for clinical
neuromodulation that is noninvasive, image-guided, and targeted to
spatially compact regions of the brain with the patient otherwise
able to participate in a neuropsychological assessment.
Figure 1
Schematic of
focused ultrasound-gated drug delivery nanoparticles’ preparation
and use. (A) To produce the propofol-loaded nanoemulsions, first the
block copolymer (yellow and blue lines) and drug (red circles) are
dissolved into THF, which is followed by a solvent extraction into
PBS to produce propofol-loaded polymeric micelles. These micelles
then emulsify liquid perfluoropentane (PFP; light blue) through sonication
at 20 kHz. (B) In use, the propofol-loaded nanoemulsions with a liquid
PFP core are sonicated at a higher frequency such as 1 MHz in these
experiments. That sonication induces a liquid to gas phase transition
of the PFP which thins the encoated drug-loaded polymer shell, inducing
drug release.
Figure 2
Schematic and in vitro
characterization of nanoparticles enabling ultrasound-gated release
of propofol for targeted neuromodulation. (A) Schematic of in vitro
testing apparatus. A PCR tube containing the aqueous particle sample
(green) was held at the focal spot of the FUS transducer. A layer
of hexane was applied on top of the sample (yellow) to serve as a
chemical sink for the released propofol. (B) Sonication induces release
of propofol from particles into the medium with a dose response after
a threshold peak in situ pressure of 0.5 MPa (left) and after a threshold
burst length of 10 ms (right). The response to burst length saturates
at 50–100 ms. N = 3–4 samples/group.
(C) Histogram of particle sizes assessed by direct particle tracking
demonstrates a single nanoscale peak centered at 317.6 ± 148.2
nm (mean ±SD). (D) After 2 h of incubation, particles were tested
for release with 1.5 MPa peak in situ pressure and 50 ms burst lengths
(N = 4 samples/group). There was intact release ability
after incubation, although release efficacy is relatively reduced
at room (25 °C) and in vivo (37 °C) temperatures.
Figure 3
Biodistribution and clearance in vivo of the
propofol-loaded nanoparticles. (A) Time course of the amount of an
initial bolus of particles found in the intravascular space, as assessed
by fluorescence of timed whole-blood samples after administration
of propofol-loaded particles doped with an infrared fluorescent dye,
compared to assessment of the serum fluorescence to determine the
unbound dye kinetics. Presented are mean ± SD, normalized by
the initial whole-blood sample fluorescence (N =
4 rats). (B) Organ distribution of particle uptake at 24 h (mean ±
SD for 4 rats) show that particles are sequestered in expected organs
such as liver, spleen, and lung with minimal amounts seen in kidney
and heart that may represent blood pool activity. No significant uptake
is seen in the brain. Values are presented as their percentage of
the total fluorescence across the harvested organs. (C) Sample bright
field (left), fluorescence (middle), and bright field/fluorescence
merged (right) images for the spleen (S), kidney (K), liver (Li),
heart (H), lung (Lu), and brain (B) after harvest from a single rat.
Figure 4
Focused ultrasound-gated propofol release is
potent enough to silence seizure activity. (A) Schematic of rat positioning
for this demonstration of in vivo efficacy. After removal of the dorsal
scalp fur, rats were placed supine on the bed of a focused ultrasound
transducer, coupled to the transducer via degassed water (light blue),
a Kapton membrane filled with degassed water (orange-brown), and ultrasound
gel (not pictured). Rats underwent seizure induction using the chemoconvulsant
PTZ. A sonication focus (red ellipse) was developed at one target
within each hemisphere, 2.5 mm lateral to midline, and 15 mm caudal
to the eye center, which equals ∼5 mm caudal to bregma. Expected
location of the two sonication foci are overlaid onto ex vivo MRI
images with the red ellipse indicating the fwhm of the sonication
focus. (B) Schematic of experiment timing for seizure induction, particle
administration, and FUS application. (C) Sample traces of EEG voltage
from one rat receiving propofol-loaded particles before and after
seizure-induction and focused ultrasound application at the indicated
pressures. (D) Total EEG power normalized by baseline averaged across
rats receiving particles loaded with either propofol (blue) or no
drug (blank, red) across experiment time (N = 7 propofol,
5 blank). Gray bars indicate time of FUS application at the indicated
estimated in situ peak pressures in 50 ms bursts applied every 1 s
for 60 s. An electrical artifact precluded EEG analysis during FUS
applications. (E) Mean ± SD of normalized total (left) and theta
band (right) EEG power in the indicated time period across rats receiving
propofol-loaded particles or blank particles (N =
7 propofol, 5 blank). Two-way ANOVA across animals receiving both
FUS treatments demonstrates significant differences with FUS application
(p < 0.01) and with particle content (p < 0.05). Posthoc multiple comparison corrected tests
show significant (p < 0.01) differences of EEG
power between baseline and each of the post FUS application periods
for the propofol particle treated rats only. (F) Mean ± SD of
the HPLC-quantified serum propofol concentration of samples from N = 4 rats taken immediately after propofol-loaded particle
administration, immediately after sonication, and 10 min post sonication,
compared to a blank serum sample. There was no appreciable serum propofol
peak for the post sonication samples.
Figure 5
MRI and histological evaluation of brains following focused-ultrasound
gated propofol release. (A) Sample whole-brain ex vivo 17.6 T MRI
of rats treated with either propofol-loaded particles or blank particles
and that underwent the seizure model and FUS application of Figure . Red ellipses in
the left images indicate the expected location and fwhm of the sonication
foci, overlaid onto the “Blank” images. Black spots
at the periphery of the brain on the MRI images are microscopic air
bubbles that show a susceptibility related blooming artifact. Notably
no such findings are present near the expected sonication field to
indicate tissue damage due to either particle administration or sonication.
(B) The 11.7 T in vivo MRI images taken presonication (T2 and T1 weighted
images left and center) and postparticle administration, postsonication,
and postcontrast administration (right) show no evidence of parenchymal
damage or blood-brain barrier opening due to particle administration
and sonication. (C) Cresyl violet histology shows no evidence of parenchymal
damage on either wide-field views (top, 4×) or magnified views
(bottom, scale bar 40 μm) for either propofol-loaded or blank
particle-treated animals that received the full sonication protocol
of Figure . The more
medial dorsal dentate gyrus (DG) was within the sonication trajectory.
The more lateral ventral dentate gyrus was not within the sonication
trajectory and serves as a negative control for assessment of damage.
Schematic of
focused ultrasound-gated drug delivery nanoparticles’ preparation
and use. (A) To produce the propofol-loaded nanoemulsions, first the
block copolymer (yellow and blue lines) and drug (red circles) are
dissolved into THF, which is followed by a solvent extraction into
PBS to produce propofol-loaded polymeric micelles. These micelles
then emulsify liquid perfluoropentane (PFP; light blue) through sonication
at 20 kHz. (B) In use, the propofol-loaded nanoemulsions with a liquid
PFP core are sonicated at a higher frequency such as 1 MHz in these
experiments. That sonication induces a liquid to gas phase transition
of the PFP which thins the encoated drug-loaded polymer shell, inducing
drug release.Schematic and in vitro
characterization of nanoparticles enabling ultrasound-gated release
of propofol for targeted neuromodulation. (A) Schematic of in vitro
testing apparatus. A PCR tube containing the aqueous particle sample
(green) was held at the focal spot of the FUS transducer. A layer
of hexane was applied on top of the sample (yellow) to serve as a
chemical sink for the released propofol. (B) Sonication induces release
of propofol from particles into the medium with a dose response after
a threshold peak in situ pressure of 0.5 MPa (left) and after a threshold
burst length of 10 ms (right). The response to burst length saturates
at 50–100 ms. N = 3–4 samples/group.
(C) Histogram of particle sizes assessed by direct particle tracking
demonstrates a single nanoscale peak centered at 317.6 ± 148.2
nm (mean ±SD). (D) After 2 h of incubation, particles were tested
for release with 1.5 MPa peak in situ pressure and 50 ms burst lengths
(N = 4 samples/group). There was intact release ability
after incubation, although release efficacy is relatively reduced
at room (25 °C) and in vivo (37 °C) temperatures.Biodistribution and clearance in vivo of the
propofol-loaded nanoparticles. (A) Time course of the amount of an
initial bolus of particles found in the intravascular space, as assessed
by fluorescence of timed whole-blood samples after administration
of propofol-loaded particles doped with an infrared fluorescent dye,
compared to assessment of the serum fluorescence to determine the
unbound dye kinetics. Presented are mean ± SD, normalized by
the initial whole-blood sample fluorescence (N =
4 rats). (B) Organ distribution of particle uptake at 24 h (mean ±
SD for 4 rats) show that particles are sequestered in expected organs
such as liver, spleen, and lung with minimal amounts seen in kidney
and heart that may represent blood pool activity. No significant uptake
is seen in the brain. Values are presented as their percentage of
the total fluorescence across the harvested organs. (C) Sample bright
field (left), fluorescence (middle), and bright field/fluorescence
merged (right) images for the spleen (S), kidney (K), liver (Li),
heart (H), lung (Lu), and brain (B) after harvest from a single rat.Focused ultrasound-gated propofol release is
potent enough to silence seizure activity. (A) Schematic of rat positioning
for this demonstration of in vivo efficacy. After removal of the dorsal
scalp fur, rats were placed supine on the bed of a focused ultrasound
transducer, coupled to the transducer via degassed water (light blue),
a Kapton membrane filled with degassed water (orange-brown), and ultrasound
gel (not pictured). Rats underwent seizure induction using the chemoconvulsant
PTZ. A sonication focus (red ellipse) was developed at one target
within each hemisphere, 2.5 mm lateral to midline, and 15 mm caudal
to the eye center, which equals ∼5 mm caudal to bregma. Expected
location of the two sonication foci are overlaid onto ex vivo MRI
images with the red ellipse indicating the fwhm of the sonication
focus. (B) Schematic of experiment timing for seizure induction, particle
administration, and FUS application. (C) Sample traces of EEG voltage
from one rat receiving propofol-loaded particles before and after
seizure-induction and focused ultrasound application at the indicated
pressures. (D) Total EEG power normalized by baseline averaged across
rats receiving particles loaded with either propofol (blue) or no
drug (blank, red) across experiment time (N = 7 propofol,
5 blank). Gray bars indicate time of FUS application at the indicated
estimated in situ peak pressures in 50 ms bursts applied every 1 s
for 60 s. An electrical artifact precluded EEG analysis during FUS
applications. (E) Mean ± SD of normalized total (left) and theta
band (right) EEG power in the indicated time period across rats receiving
propofol-loaded particles or blank particles (N =
7 propofol, 5 blank). Two-way ANOVA across animals receiving both
FUS treatments demonstrates significant differences with FUS application
(p < 0.01) and with particle content (p < 0.05). Posthoc multiple comparison corrected tests
show significant (p < 0.01) differences of EEG
power between baseline and each of the post FUS application periods
for the propofol particle treated rats only. (F) Mean ± SD of
the HPLC-quantified serum propofol concentration of samples from N = 4 rats taken immediately after propofol-loaded particle
administration, immediately after sonication, and 10 min post sonication,
compared to a blank serum sample. There was no appreciable serum propofol
peak for the post sonication samples.MRI and histological evaluation of brains following focused-ultrasound
gated propofol release. (A) Sample whole-brain ex vivo 17.6 T MRI
of rats treated with either propofol-loaded particles or blank particles
and that underwent the seizure model and FUS application of Figure . Red ellipses in
the left images indicate the expected location and fwhm of the sonication
foci, overlaid onto the “Blank” images. Black spots
at the periphery of the brain on the MRI images are microscopic air
bubbles that show a susceptibility related blooming artifact. Notably
no such findings are present near the expected sonication field to
indicate tissue damage due to either particle administration or sonication.
(B) The 11.7 T in vivo MRI images taken presonication (T2 and T1 weighted
images left and center) and postparticle administration, postsonication,
and postcontrast administration (right) show no evidence of parenchymal
damage or blood-brain barrier opening due to particle administration
and sonication. (C) Cresyl violet histology shows no evidence of parenchymal
damage on either wide-field views (top, 4×) or magnified views
(bottom, scale bar 40 μm) for either propofol-loaded or blank
particle-treated animals that received the full sonication protocol
of Figure . The more
medial dorsal dentate gyrus (DG) was within the sonication trajectory.
The more lateral ventral dentate gyrus was not within the sonication
trajectory and serves as a negative control for assessment of damage.Particles that encapsulated propofol
with a liquid perfluorocarbon core and a biodegradable, biocompatible
polymer coating were produced (Figure ) and sized via nanoparticle tracking analysis (Figure C). There was a single
nanoscale peak of 320 ± 150 nm (mean ± SD). Encapsulation
efficiency of the propofol was 11.8% ± 1.2% (mean ± SD)
yielding an encapsulated 177 μg ± 19 μg propofol
per mL of particles. In vitro particle release efficacy was assessed
by focused ultrasound applied at 1 MHz center frequency in short continuous
bursts with 0.5 Hz burst frequency for a total of 2 min with varying
peak in situ pressure and the individual burst length, that is, the
short amount of time that sonication is applied continuously. The
amount of released propofol was assessed by extraction into a hexane
sink (Figure A) and
quantified via UV fluorescence. There was a dose response evident
for propofol release with peak in situ pressures past a threshold
of 0.5 MPa. For burst length, a release threshold of 10 ms was present
with saturation of a dose response between 50 and 100 ms (Figure B). Particles kept
in storage and in vivo-like conditions for 2 h followed by sonication
at room temperature showed intact release ability, although release
efficacy was reduced after 2 h of incubation at room and in vivo temperatures
(Figure D) possibly
due to diffusion of the perfluoropentane (PFP) from the core of the
particles with higher temperature incubation.To evaluate the
in vivo biodistribution and intravascular residence time of the nanoparticles,
the particles were initially doped with a custom synthesized hydrophobic
dye. Following intravenous administration of these doped nanoparticles,
timed blood samples demonstrated that the whole-blood fluorescence
has a decay profile that is faithfully characterized with a double
exponential decay model (Figure A). The initial phase decay half-life was 8.8 min and
the second phase decay half-life was 270 min. Notably, the whole blood
samples were expected to contain both intact particles and free and
micelle-bound portions of the dye. In contrast, the serum of these
samples would contain free dye or potentially PEG−PCL micelle-bound
dye after high-speed centrifugation pellets the cellular and nanoparticle
constituents. The serum fractions showed a markedly lower fluorescence
that cleared more rapidly than the whole-blood fluorescence signal,
with no appreciable serum fluorescence by 2 h. The serum sample fluorescence
decayed with a monoexponential profile with a half-life calculated
as 8 min, notably similar to the short half-life component of the
whole blood samples. After 24 h from particle administration, there
was no remnant intravascular signal above background. End organ fluorescence
demonstrated no evidence of nonspecific particle binding to the brain
(Figure B,C). Instead,
the nanoparticles were principally taken up by the liver, spleen,
and to a lesser extent the lungs, with minimal amounts in the kidney
and heart.To demonstrate and assess the functional potency
of particle release in vivo, an acute pentylenetetrazol (PTZ)-induced
status epilepticus protocol[20] was developed
for adult male Fischer 344 rats (Figure A,B). We specifically chose this protocol
and preparation as prior groups have used this system to assess the
degree to which FUS may directly modulate neural activity.[21] Following seizure induction and particle administration,
there was no significant difference in baseline EEG power between
animals receiving propofol-loaded particles and particles generated
with no drug (“Blank”; see Supporting Information). Importantly, following FUS administration first
at 1.0 MPa estimated peak in situ pressure and then at 1.5 MPa, immediate
statistically significant declines of total and theta band EEG power
were seen in the animals receiving propofol-loaded particles, but
not in the animals receiving blank particles (Figure C–F).Ex vivo 17.6 T MRI, in
vivo 11.7 T MRI, and histology confirmed that no deleterious effect
of FUS and particle administration was visible (Figure ). In particular, given the high susceptibility
dependence of the MRI protocol used here (note the blooming artifact
from microscopic air bubbles along the brain periphery in Figure A), the lack of any
noted susceptibility artifact or brain parenchymal signal change within
the sonicated region confirms the lack of petechial hemorrhage or
other cavitation induced damage to the brain parenchyma. Notably,
the 17.6 T MRI evaluation covered the entire brain in both axial and
coronal planes, without interslice gaps, ensuring that a complete
evaluation of the parenchyma was completed for each brain. All MRI
images were reviewed by a board-certified neuroradiologist. In vivo
MRI also confirmed no damage to the brain parenchyma of particle administration
and sonication, and no evidence of blood-brain barrier opening with
this technique (Figure B). Whole-brain histological sections and more focused evaluation
of the sonicated dorsal dentate gyrus in comparison with the nonsonicated
ventral dentate gyrus showed no evidence of parenchymal damage and
certainly no damage that could be attributed to sonication (Figure C).We have
therefore described nanoparticles that allow focused ultrasound-induced
uncaging of the small molecule anesthetic agent propofol (Figure ) and demonstrated
the in vitro and in vivo efficacy of the nanoparticles as a proof-of-principle.
Given that these particles have a hydrodynamic diameter of approximately
320 nm and that similar perfluorocarbon-based phase-change particles
have been shown to increase diameter up to 5–6 times during
sonication,[16,22] the maximal diameter of these
particles after activation would be <2 μm, suggesting no
substantial risk of embolization of capillaries with these nanoparticles
and their use. These particles release their drug cargo with dose
responses with both peak in situ pressure and with sonication burst
length (Figure ).
The threshold peak in situ pressure of 0.5 MPa and the maximal pressure
of 1.5 MPa that were used here are both achievable by current clinical
transcranial MRgFUS systems.[14,15] Additionally, the dynamic
range of the burst length dose response between 10 ms and 50–100
ms is also achievable with these clinical transcranial MRgFUS systems.
These burst lengths and duty cycles are unlikely to induce substantial
heating of the brain parenchyma, especially given heat dissipation
by cerebral perfusion.We were able to use particles doped with
an IR fluorescent dye as a surrogate marker of particle intravascular
residence and distribution (Figure ). The serum fluorescence, which would contain the
unbound free dye fraction, showed a much more rapid clearance from
the blood pool than the whole-blood samples that represent both the
particle-bound and unbound fractions. The whole-blood fluorescence
particle elimination profile showed two phases: an initial rapid (9
min half-life) phase that likely corresponds to the unbound dye fraction
of the sample, and a slower (270 min half-life) phase that more represents
the particle decay profile itself (Figure A). This half-life would allow enough time
for a clinically relevant intervention with these particles, but not
so long of a particle vascular residence time that it would preclude
repeat particle administration or would suggest a potential toxicity
of extended particle residence in the body. The lack of particle uptake
in the brain (Figure B,C) confirms that our results are unlikely to be due to particle
crossing of the blood brain barrier, particle binding to the brain,
or some other nonspecific action of the particles upon the brain.
The finding that the liver and spleen primarily take up these particles
is expected as the reticuloendothelial system generally sequesters
nanoscale material.[23]The ability
of focused ultrasound to activate the intravascular propofol-loaded
particles and yield silencing of seizure activity in vivo (Figure ) indicates that
these particles indeed can enable a potent neuromodulatory effect
upon focused ultrasound application. Crucially, given that there was
no significant effect of sonication in rats receiving the blank particles
and seeing as the blank particles were otherwise constructed exactly
the same as the propofol-loaded particles, the effects seen here are
specifically related to the release of propofol in this system and
not a nonspecific effect of ultrasound or particle interaction with
neural tissue or of the individual polymer or perfluorocarbon particle
constituents. Given that our total encapsulation efficiency of 177 μg/mL
translates to ∼1 mg/kg in these experiments and that a normal
loading dose for anesthetic effect in rats is an order of magnitude
higher at 10 mg/kg,[24] it is unlikely that
our results are due to a nonspecific leak of the propofol from the
particles. Indeed, serum propofol concentrations taken immediately
and 10 min after sonication showed no appreciable propofol above the
background (Figure F), indicating that this propofol release was likely limited to the
brain, without nonspecific systemic delivery. The detectable serum
propofol seen immediately following particle administration likely
reflects a small amount of free propofol in the particle batch given
our method of production, although this level of ∼0.6 μg/mL
is an order of magnitude less than the typical serum concentrations
of propofol thought to be necessary for an anesthetic effect. Taken
together, these results suggest that these particles indeed yield
a higher local drug concentration in the brain following FUS application
than might be suggested by the raw total amount of drug delivered
in the bolus intravenous dose. Additionally, given the fast 2–3
min distribution half-life of propofol from the blood-pool,[25] given that we waited 5−10 min from particle
administration to FUS application, and given that we did not see significant
differences in the baseline EEG power between propofol and blank treated
rats, it is unlikely that free or loosely bound propofol in the particle
solution could have substantially contributed to our results. Notably,
the EEG power was seen to decrease immediately following FUS application
suggesting that the kinetics of this neuromodulatory effect are rapid
(Figure D). Additionally,
we did not observe any deleterious consequence to the brains with
ex vivo MRI, in vivo MRI, or post hoc histology (Figure ), suggesting that these effects
are not due to a nonspecific damage of the brain parenchyma. Indeed,
the lack of blood-brain barrier opening with this technique (Figure B) confirms the safety
of this technique and the distinction of this technique between it
and other proposed methods of FUS-mediated neuromodulation.[7,21]In this study, we have not directly visualized the particle
activation in vivo to assess the effective spatial resolution of this
technique. Additionally, the volume conduction effect of EEG signals,
particularly for subdermal EEG, and the nature of the generalized
status epilepticus model used for this study limits our ability to
spatially resolve this signal in this in vivo preparation. Similarly,
given the acute nature of PTZ-induced seizures that may not recur
substantially once they are aborted, this study protocol is limited
in ability to determine over what time interval the action of the
gated propofol persists. Those limitations said, the in situ ultrasound
focus induced by the particular ultrasound transducer used in this
study is known to have a fwhm of ∼1.5 mm transaxially and ∼5
mm longitudinally at 1 MHz (personal communication with the vendor,
FUS Instruments, Toronto, CA), providing an effective initial spatial
extent for the action of the particles in this preparation. Additionally,
we saw no substantial systemic propofol load with sonication (Figure F), confirming that
the propofol release was likely limited to the brain. While groups
have shown that activated perfluorocarbon particles may induce further
activation of unsonicated particles in static solutions,[16] we would expect cerebral perfusion to rapidly
clear the activated particles from the sonication field, especially
given the lack of particle binding to the brain (Figure ), thereby limiting this potential
confound. Additionally, the temporal residence of propofol in the
brain and its time of action is known to be rapid on the order of
minutes or even tens of seconds[26] and similar
to the time-scales used in this experiment. This time of action would
be clinically practical for neuropsychological assessment, as evidenced
by the current protocol of the Wada test,[27] which is used for clinical mapping of the laterality of brain functions.
Given the size of the rat cerebrum (∼15 × 15 × 10
mm[28]) and the technical limitations of
signal volume conduction in subdermal EEG, further characterization
of the spatial and temporal resolution of this technique will necessitate
experiments that assess baseline nonseizure neural activity, likely
in larger animal models, and potentially with a different measure
of neural activity, such as fMRI or PET. Nonetheless, our results
provide a proof-of-principle that these nanoparticles yield potent
inducible neuromodulation using noninvasive focused ultrasound and
that this approach has the potential to enable precise spatial (mm)
and temporal (min) control of brain activity, with a pathway to clinical
translation.With regard to clinical translation, each component
of these particles has been previously approved for clinical use in
different contexts.[19] Additionally, the
sonication pressures and burst lengths used in this study are well
achievable by FDA-approved transcranial MRgFUS systems that are currently
in clinical use.[14] Taken together, this
provides a pathway toward clinical translation that is otherwise unavailable
to other targeted molecular neuromodulation strategies. Further, the
chemistry that enables these particles to encapsulate a given drug
relies mainly upon the lipophilicity of the drug in question, so that
it may bind the hydrophobic domains of the encapsulating block copolymer
and the hydrophobic polymer–perfluorocarbon interface. Given
that most molecules that passively cross the blood-brain barrier are
highly lipophilic, this suggests that the nanotechnology strategy
presented here could be adapted for focal and targeted delivery of
most any small molecule that naturally crosses the blood-brain barrier,
including imaging agents as well as compounds that act directly upon
the adrenergic, serotonergic, or dopaminergic systems, in addition
to the excitation/inhibition axis that propofol modulates. This opens
the door to a wide variety of potential nanotechnological tools for
targeted clinical modulation of brain activity.
Authors: Pejman Ghanouni; Kim Butts Pauly; W Jeff Elias; Jaimie Henderson; Jason Sheehan; Stephen Monteith; Max Wintermark Journal: AJR Am J Roentgenol Date: 2015-07 Impact factor: 3.959
Authors: Zhi-De Deng; Bruce Luber; Nicholas L Balderston; Melbaliz Velez Afanador; Michelle M Noh; Jeena Thomas; William C Altekruse; Shannon L Exley; Shriya Awasthi; Sarah H Lisanby Journal: Annu Rev Pharmacol Toxicol Date: 2020-01-06 Impact factor: 13.820