Lia Bally1, Hood Thabit2, Harald Kojzar3, Julia K Mader3, Jehona Qerimi-Hyseni3, Sara Hartnell4, Martin Tauschmann5, Janet M Allen5, Malgorzata E Wilinska5, Thomas R Pieber3, Mark L Evans2, Roman Hovorka6.
Abstract
BACKGROUND: Tight control of blood glucose concentration in people with type 1 diabetes predisposes to hypoglycaemia. We aimed to investigate whether day-and-night hybrid closed-loop insulin delivery can improve glucose control while alleviating the risk of hypoglycaemia in adults with HbA1c below 7·5% (58 mmol/mol).Entities:
Mesh:
Substances:
Year: 2017 PMID: 28094136 PMCID: PMC5379244 DOI: 10.1016/S2213-8587(17)30001-3
Source DB: PubMed Journal: Lancet Diabetes Endocrinol ISSN: 2213-8587 Impact factor: 44.867
Figure 1Trial profile
Baseline characteristics
| Sex | ||
| Female | 15 (52%) | |
| Male | 14 (48%) | |
| Age (years) | 41 (13) | |
| Bodyweight (kg) | 72·9 (13·0) | |
| BMI (kg/m2) | 25·1 (3·0) | |
| HbA1c (%) | 6·9 (0·5) | |
| HbA1c (mmol/mol) | 51·7 (4·8) | |
| Duration of diabetes (years) | 24 (12) | |
| Duration using pump (years) | 6 (4) | |
| Total daily insulin (U/kg/day) | 0·5 (0·1) | |
| Glucose sensor use | ||
| No previous glucose sensor use | 18 (62%) | |
| Real-time continuous glucose monitoring | 5 (17%) | |
| Flash glucose monitoring | 6 (21%) | |
Data are mean (SD) or n (%).
Overall day-and-night glucose control during closed-loop and control periods based on sensor glucose measurements
| Proportion of time with glucose concentration in range (%) | |||||
| 3·9–10·0 mmol/L | 76·2% (6·4) | 65·6% (8·1) | 10·5 (7·6 to 13·4) | <0·0001 | |
| >10·0 mmol/L | 20·4% (6·3) | 27·4% (9·6) | −6·9 (−10·2 to −3·5) | 0·0003 | |
| >13·9 mmol/L | 3·8% (2·6) | 6·9% (3·9) | −3·0 (−4·4 to −1·6) | 0·0002 | |
| >16·7 mmol/L | 0·9% (0·8) | 2·1% (1·8) | −1·2 (−1·9 to −0·6) | 0·0009 | |
| <3·9 mmol/L | 2·9% (2·3 to 4·0) | 5·3% (3·5 to 10·0) | 0·50 (0·41 to 0·63) | <0·0001 | |
| <3·5 mmol/L | 1·3% (0·8 to 2·3) | 3·4% (1·9 to 7·2) | 0·35 (0·26 to 0·47) | <0·0001 | |
| <3·3 mmol/L | 0·9% (0·5 to 1·7) | 2·6% (1·3 to 5·5) | 0·30 (0·22 to 0·43) | <0·0001 | |
| <2·8 mmol/L | 0·3% (0·1 to 0·5) | 1·0% (0·5 to 2·6) | 0·24 (0·14 to 0·41) | <0·0001 | |
| AUCday <3·5 mmol/L (min × mmol/L) | 9·1 (3·7 to 18·2) | 26·7 (13·1 to 65·5) | 0·27 (0·18 to 0·41) | <0·0001 | |
| Mean glucose concentration (mmol/L) | 7·9 (0·5) | 8·3 (0·9) | −0·4 (−0·7 to −0·1) | 0·0226 | |
| SD of glucose concentration (mmol/L) | 2·8 (0·4) | 3·3 (0·5) | −0·5 (−0·7 to −0·3) | <0·0001 | |
| Coefficient of variation of glucose concentration | |||||
| Within days (%) | 35·3 (3·0) | 40·3 (5·1) | −5·0 (−7·1 to −3·0) | <0·0001 | |
| Between days (%) | 12·8 (3·3) | 20·2 (4·6) | −7·5 (−9·7 to −5·3) | <0·0001 | |
| Sensor glucose concentration <3·5 mmol/L for at least 20 min | |||||
| Number of nights | 2·1 (1·0) | 5·6 (3·5) | −3·6 (−4·9 to −2·2) | <0·0001 | |
| Mean duration of each episode (min) | 45 (13) | 75 (25) | −29 (−38 to −20) | <0·0001 | |
Data are mean (SD) or median (IQR), unless otherwise stated. No significant period effect was observed. AUCday=sensor glucose area under the curve per day.
Unless specified otherwise, data are normally distributed and presented as mean difference of closed-loop period minus control period, with 95% CI for mean; a positive value indicates that the measurement was higher during the closed-loop period than in the control period.
Non-normally distributed data are presented as ratio of closed-loop data over control data, with 95% CI for ratio; a value greater than unity indicates that the measurement was higher in the closed-loop period than in the control period.
Primary outcome.
Figure 2Median sensor glucose and insulin delivery for the 24 h duration over the study period
Median (IQR) sensor glucose concentration (A) and insulin delivery (B) during closed-loop period (solid red line and red shaded area) and control period (dashed black line and grey shaded area) for the 24 h duration. The horizontal dashed lines show the lower and upper limits of the glucose target range (3·9–10·0 mmol/L).
Figure 3Individual values of mean sensor glucose and proportion of time spent with glucose concentration below the target range for the whole study
The size of the circles denotes the proportion of time spent with low glucose (<3·5 mmol/L).
Insulin delivery over 24 h period
| Total daily insulin (U/day) | 37·5 (13·8) | 37·4 (12·6) | 0·8 (−1·0 to 2·6) | 0·36 |
| Total bolus insulin (U/day) | 18·6 (7·9) | 20·2 (8·4) | −1·1 (−2·4 to 0·2) | 0·11 |
| Total basal insulin (U/day) | 18·9 (7·8) | 17·2 (5·7) | 1·9 (0·7 to 3·1) | 0·0038 |
Data are mean (SD), unless otherwise stated.
Normally distributed data are presented as mean difference of closed-loop period minus control period, with 95% CI for mean; a positive value indicates that the measurement was higher in the closed-loop period than in the control period.
Night-time and daytime glucose control during closed-loop and control periods based on sensor glucose measurements
| Proportion of time with glucose concentration in range (%) | |||||
| 3·9–10·0 mmol/L | 82·0% (9·7) | 64·5% (11·5) | 17·2 (12·0 to 22·4) | <0·0001 | |
| >10·0 mmol/L | 14·9% (8·5) | 25·4% (11·8) | −10·2 (−15·4 to −5·1) | 0·0004 | |
| <3·9 mmol/L | 3·2% (1·6 to 4·0) | 9·0% (4·6 to 15·7) | 0·33 (0·24 to 0·45) | <0·0001 | |
| <3·5 mmol/L | 1·1% (0·4 to 2·2) | 5·4% (2·9 to 11·4) | 0·19 (0·12 to 0·28) | <0·0001 | |
| <2·8 mmol/L | 0·1% (0·0 to 0·6) | 1·7% (0·9 to 5·7) | 0·14 (0·08 to 0·26) | <0·0001 | |
| AUCday <3·5 mmol/L (mmol/L × min) | 5·1 (1·4 to 15·5) | 41·5 (21·5 to 122·8) | 0·11 (0·06 to 0·20) | <0·0001 | |
| Mean glucose concentration (mmol/L) | 7·5 (0·6) | 8·0 (1·0) | −0·4 (−0·8 to −0·1) | 0·0211 | |
| SD of glucose concentration (mmol/L) | 2·5 (0·6) | 3·2 (0·6) | −0·7 (−1·0 to −0·4) | <0·0001 | |
| Coefficient of variation of glucose concentration between nights (%) | 24·9 (7·1) | 34·4 (6·4) | −9·6 (−13·2 to −5·9) | <0·0001 | |
| Proportion of time with glucose concentration in range (%) | |||||
| 3·9–10·0 mmol/L | 74·3% (6·9) | 66·1% (8·8) | 8·1 (5·3 to 11·0) | <0·0001 | |
| >10·0 mmol/L | 22·2% (7·2) | 27·9% (10·4) | −5·6 (−8·9 to −2·3) | 0·0023 | |
| <3·9 mmol/L | 2·7% (1·9 to 4·5) | 4·4% (2·8 to 8·6) | 0·61 (0·49 to 0·76) | 0·0001 | |
| <3·5 mmol/L | 1·2% (0·7 to 2·2) | 2·3% (1·4 to 6·0) | 0·47 (0·35 to 0·64) | <0·0001 | |
| <2·8 mmol/L | 0·2% (0·1 to 0·5) | 0·5% (0·2 to 1·3) | 0·48 (0·29 to 0·80) | 0·0076 | |
| AUCday <3·5 mmol/L (mmol/L × min) | 8·3 (3·4 to 14·3) | 15·9 (7·5 to 45·7) | 0·39 (0·25 to 0·86) | 0·0001 | |
| Mean glucose concentration (mmol/L) | 8·0 (0·6) | 8·4 (1·0) | −0·3 (−0·6 to 0·0) | 0·0498 | |
| SD of glucose concentration (mmol/L) | 2·9 (0·4) | 3·3 (0·6) | −0·5 (−0·7 to −0·2) | 0·0002 | |
| Coefficient of variation of glucose concentration between days (%) | 13·8 (3·0) | 20·6 (5·1) | −6·8 (−9·1 to −4·6) | <0·0001 | |
Data are presented as mean (SD) or median (IQR), unless otherwise stated. No significant period effect was observed. AUCday=sensor glucose area under the curve per day.
Unless specified otherwise, data are normally distributed and presented as mean difference of closed-loop period minus control period, with 95% CI for mean; a positive value indicates that the measurement was higher in the closed-loop period than in the control period.
Non-normally distributed data are presented as ratio of closed-loop data over control data, with 95% CI for ratio; a value greater than unity indicates that the measurement was higher in the closed-loop period than in the control period.