| Literature DB >> 28042557 |
Shinho Hong1, Keun-Myoung Park1, Yong Sun Jeon2, Soon Gu Cho2, Kee Chun Hong1, Woo Young Shin1, Yun-Mee Choe1.
Abstract
PURPOSE: Implementing an inferior vena cava (IVC) filter is a relatively safe procedure but potential negative long-term effects. The complications for filter retrieval have been noted. We examined filter characteristics on pre-retrieval computed tomography (CT) that were associated with complicated retrieval (CR) of IVC filters.Entities:
Keywords: Device removal; Vena cava filters; Venous thrombosis
Year: 2016 PMID: 28042557 PMCID: PMC5198764 DOI: 10.5758/vsi.2016.32.4.175
Source DB: PubMed Journal: Vasc Specialist Int ISSN: 2288-7970
Fig. 1.(A) Axial computed tomography (CT) image shows grade 2 perforation. (B) Inferior vena cava (IVC) filter shape changes due to adhesion and external compression. (C) Sagittal CT image shows embedding to the IVC anterior wall. (D) Catheter-steered snare wire use.
A comparison of characteristics between the CR group and the non-CR group
| Variable | CR group (n=24) | Non-CR group (n=56) | P-value |
|---|---|---|---|
| Demographics | |||
| Age (y) | 53.0±23.2 (39–76) | 59.0±29.5 (36–82) | 0.652 |
| Sex, male | 16 (66.7) | 32 (57.1) | 0.749 |
| Indications for insertion | |||
| Pulmonary embolism | 2 (8.3) | 5 (8.9) | |
| Thrombolysis | 16 (66.7) | 40 (71.4) | 0.853 |
| Contraindications for anticoagulation | 6 (25.0) | 8 (14.3) | |
| Dwelling time (d) | 50.0±78.1 (15–198) | 38.0±42.5 (14–135) | <0.010 |
| Within 45 | 10 | 20 | |
| Over 45 | 14 | 36 | |
| Type | |||
| Optease® | 18 (75.0) | 47 (83.9) | 0.754 |
| Tulip® | 6 (25.0) | 9 (16.1) |
Values are presented as mean±standard deviation (range), number (%), number only.
CR, complicated retrieval.
Independent t-test,
Fisher’s exact test.
A comparison of computed tomogram findings between the CR group and the non-CR group
| Variable | CR group (n=24) | Non-CR group (n=56) | P-value |
|---|---|---|---|
| Tilt angle | |||
| Anteroposterior (mean, °) | 15 | 3 | <0.01 |
| Mediolateral (mean, °) | 19 | 4 | <0.01 |
| Tilt over 15° | 22 | 11 | <0.01 |
| Tip embedding | 7 | 1 | <0.01 |
| Perforation degree | |||
| Grade 0 | 2 | 29 | <0.01 |
| Grade 1 | 8 | 23 | |
| Grade 2 | 14 | 4 | |
| Distance from renal vein (mean, mm) | 11.2 | 12.1 | 0.87 |
CR, complicated retrieval.
Independent t-test,
Fisher’s exact test.
A multivariate analysis of factors associated with complicated retrieval according to dwelling time
| Dwelling time | Over 45 days | Within 45 days | ||
|---|---|---|---|---|
|
|
| |||
| OR (95% CI) | P-value | OR (95% CI) | P-value | |
| Tilt over 15° | 6.476 (2.194–12.147) | <0.01 | 4.476 (0.912–11.084) | 0.08 |
| Tip embedding | 7.213 (3.234–11.231) | <0.01 | 3.742 (0.845–10.231) | 0.15 |
| Perforation degree (grade 2) | 2.165 (3.193–9.134) | <0.01 | 1.165 (0.693–6.267) | 0.45 |
OR, Odd ratio; CI, confidence interval.
Logistic regression test.