| Literature DB >> 28025782 |
Hiroshi Watanabe1,2, Kimiya Noto3, Tomokazu Shohji4, Yasuyoshi Ogawa5, Toshioh Fujibuchi6, Ichiro Yamaguchi7, Hitoshi Hiraki8, Tetsuo Kida9, Kazutoshi Sasanuma10, Yasushi Katsunuma11, Takurou Nakano12, Genki Horitsugi13, Makoto Hosono14.
Abstract
The goal of this study is to develop a more appropriate shielding calculation method for computed tomography (CT) in comparison with the Japanese conventional (JC) method and the National Council on Radiation Protection and Measurements (NCRP)-dose length product (DLP) method. Scattered dose distributions were measured in a CT room with 18 scanners (16 scanners in the case of the JC method) for one week during routine clinical use. The radiation doses were calculated for the same period using the JC and NCRP-DLP methods. The mean (NCRP-DLP-calculated dose)/(measured dose) ratios in each direction ranged from 1.7 ± 0.6 to 55 ± 24 (mean ± standard deviation). The NCRP-DLP method underestimated the dose at 3.4% in fewer shielding directions without the gantry and a subject, and the minimum (NCRP-DLP-calculated dose)/(measured dose) ratio was 0.6. The reduction factors were 0.036 ± 0.014 and 0.24 ± 0.061 for the gantry and couch directions, respectively. The (JC-calculated dose)/(measured dose) ratios ranged from 11 ± 8.7 to 404 ± 340. The air kerma scatter factor κ is expected to be twice as high as that calculated with the NCRP-DLP method and the reduction factors are expected to be 0.1 and 0.4 for the gantry and couch directions, respectively. We, therefore, propose a more appropriate method, the Japanese-DLP method, which resolves the issues of possible underestimation of the scattered radiation and overestimation of the reduction factors in the gantry and couch directions.Entities:
Keywords: Air kerma scatter factor; Computed tomography; Dose length product; Shielding calculation method; Workload
Mesh:
Year: 2016 PMID: 28025782 PMCID: PMC5487862 DOI: 10.1007/s12194-016-0387-9
Source DB: PubMed Journal: Radiol Phys Technol ISSN: 1865-0333
Fig. 1Schematic diagram of the dosimeter arrangement
Fig. 2Schematic diagram of the JC method for CT shielding calculation
Parameters and description of the JC methoda
| Parameter | Description (unit) |
|---|---|
|
| Effective dose due to the primary beam (mSv) |
|
| Effective dose due to secondary radiations (mSv) |
|
| Effective dose due to leakage from the tube housing (mSv) |
|
| Air Kerma of rate per 1 mA standardized at 1 m from the focus of X-ray tube (mGym2/mAs) |
|
| Air kerma transmission factor on barrier of thickness |
|
| Workload (mAs, in case of |
|
| Converting factor to effective dose from air kerma (Sv/Gy) |
|
| Use factor |
|
| Occupancy factor |
|
| Distance from the focus of X-ray tube to the point of interest for evaluation of |
|
| Distance from the patient to the point of interest for evaluation of |
|
| Distance from the focus of X-ray tube to the patient (m) |
|
| Distance from the focus of X-ray tube to the point of interest for evaluation of |
|
| Scaled scatter fraction (scattered radiation ratio to |
|
| Exposure field (cm2) |
|
| Leakage radiation dose rate (air kerma) from a tube housing standardized at 1 m from the housing (equal to 1 mGy/h according to the Ordinance for Enforcement of the Medical Care Act (mGym2/h) |
|
| Number of hours of beam on-time (h) |
aJC method: Japanese conventional method
Characteristics of examined CT scanners and comparison of DLPs and workloads (mAs)
| Scanner | Manufacturer and model | Number of detector rows | Maximum beam width (cm) | Region | Total DLPa/week | DLPa/week/region | Total workload/week | Workload/week/region | Number of scans | Head examinations ratio as DLP basisb (%) |
|---|---|---|---|---|---|---|---|---|---|---|
| CT-1 | Toshiba | 64 | 3.2 | Head | 429,619 | 147,415 | – | – | 146 | 34.3 |
| Aquilion CX | Body | 282,204 | – | 290 | ||||||
| CT-2 | Toshiba | 64 | 3.2 | Head | 500,739 | 171,884 | – | – | 182 | 34.3 |
| Aquilion64 | Body | 328,855 | – | 344 | ||||||
| CT-3 | Toshiba | 64 | 3.2 | Head | 120,587 | 31,780 | 309,468 | 84,194 | 50 | 26.4 |
| Aquilion64 | Body | 88,807 | 225,274 | 153 | ||||||
| CT-4 | Toshiba | 80 | 4 | Head | 56,078 | 45,850 | 105,810 | 76,479 | 60 | 81.8 |
| Prime 80 | Body | 10,228 | 29,331 | 46 | ||||||
| CT-5 | Toshiba | 64 | 3.2 | Head | 240,577 | 64,314 | 678,783 | 202,591 | 79 | 26.7 |
| Aquilion 64 | Body | 176,263 | 476,192 | 249 | ||||||
| CT-6 | Hitachi | 64 | 4 | Head | 171,881 | 64,478 | 432,662 | 181,766 | 108 | 37.5 |
| Scenaria | Body | 107,404 | 250,897 | 99 | ||||||
| CT-7 | GE | 64 | 4 | Head | 349,317 | 27,276 | 797,230 | 148,788 | 32 | 7.8 |
| VCT | Body | 322,042 | 648,442 | 320 | ||||||
| CT-8 | Hitachi | 64 | 4 | Head | 263,440 | 86,603 | 857,834 | 354,165 | 95 | 32.9 |
| Scenaria | Body | 176,837 | 503,669 | 237 | ||||||
| CT-9 | Hitachi | 16 | 2 | Head | 15,967 | 5369 | 98,827 | 39,907 | 8 | 33.6 |
| Eclos | Body | 10,597 | 58,920 | 18 | ||||||
| CT-10 | GE | 64 | 4 | Head | 222,722 | 39,567 | 716,886 | 113,500 | 33 | 17.8 |
| VCT | Body | 183,156 | 603,387 | 223 | ||||||
| CT-11 | Siemens | 128 | 3.84 | Head | 368,432 | 27,624 | 1,592,397 | 38,484 | 15 | 7.5 |
| Definition Flash | Body | 340,808 | 1,553,913 | 464 | ||||||
| CT-12 | GE | 64 | 4 | Head | 442,330 | 14,316 | 1,472,663 | 49,664 | 17 | 3.2 |
| VCT | Body | 428,014 | 1,422,999 | 480 | ||||||
| CT-13 | Toshiba | 320 | 16 | Head | 262,459 | 51,570 | 741,219 | 87,495 | 46 | 19.6 |
| One320 | Body | 210,889 | 624,837 | 188 | ||||||
| CT-14 | Toshiba | 320 | 16 | Head | 133,360 | 42,147 | 287,042 | 49,348 | 65 | 31.6 |
| One320 | Body | 91,214 | 237,694 | 154 | ||||||
| CT-15 | Siemens | 64 | 3.84 | Head | 239,687 | 90,459 | 997,378 | 341,063 | 112 | 37.7 |
| Definition AS+ | Body | 149,228 | 656,315 | 371 | ||||||
| CT-16 | Siemens | 64 | 3.84 | Head | 225,482 | 37,670 | 894,669 | 79,236 | 63 | 16.7 |
| Somatom Definition Edge | Body | 187,811 | 815,433 | 365 | ||||||
| CT-17 | Siemens | 128 | 3.84 | Head | 162,523 | 26,020 | 692,390 | 65,053 | 41 | 16.0 |
| Somatom Definition Flash | Body | 136,504 | 627,337 | 315 | ||||||
| CT-18 | GE | 64 | 4 | Head | 166,343 | 52,435 | 386,864 | 120,144 | 48 | 31.5 |
| Lightspeed VCT | Body | 113,908 | 266,720 | 172 | ||||||
| Mean | 27.6 | |||||||||
| Standard deviation | 17.4 | |||||||||
aUnits of DLP is Gy cm
bHead examinations ratio as DLP basis was calculated as the ratio of head examination/head and body examinations
cIt could not obtain the actual workload
Converted doses and distances in the CT room
| Item | Point (°) | CT-1 | CT-2 | CT-3 | CT-4 | CT-5 | CT-6 | CT-7 | CT-8 | CT-9 | CT-10 | CT-11 | CT-12 | CT-13 | CT-14 | CT-15 | CT-16 | CT-17 | CT-18 |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Converted dose at 1 ma (mSv) | 0 | 46.74 | 53.80 | 16.08 | 8.31 | 24.09 | 27.09 | 35.06 | 37.23 | 2.23 | 32.83 | 48.50 | 47.93 | 29.94 | 12.73 | 33.59 | 32.02 | 24.13 | 26.79 |
| 45 | 82.42 | 106.77 | 12.78 | –c | 18.88 | 42.24 | 14.71 | 34.81 | 0.88 | 27.64 | 65.12 | 43.98 | 67.25 | 10.14 | 29.30 | 32.27 | 22.06 | 16.94 | |
| 90 | 3.49 | 4.81 | 1.82 | 0.26 | 1.74 | 1.85 | 1.70 | 2.36 | 0.26 | 2.74 | 2.51 | 5.59 | 1.46 | 0.88 | 1.13 | 2.24 | 0.79 | 1.39 | |
| 135 | 97.47 | 166.77 | 48.55 | 11.09 | 48.67 | 37.58 | 71.41 | 52.96 | 5.54 | 66.44 | 85.30 | 170.67 | 55.12 | 26.83 | 18.87 | 44.39 | 50.44 | 45.19 | |
| 180 | 21.29 | 23.91 | 8.66 | 2.14 | 15.29 | 6.75 | 18.10 | 18.89 | 1.44 | 21.92 | 35.65 | 30.64 | 12.62 | 6.42 | 14.91 | 15.26 | 13.18 | 16.05 | |
| 225 | 95.22 | 91.05 | 26.49 | –c | 54.68 | 31.29 | 44.11 | 92.88 | 5.08 | 87.01 | 112.66 | 147.90 | 56.56 | 26.73 | 47.59 | 47.72 | 48.76 | 30.26 | |
| 270 | 3.56 | 5.90 | 1.53 | 0.44 | 2.02 | 1.25 | 1.22 | 4.54 | 0.24 | 1.72 | 6.06 | 4.64 | 2.89 | 0.82 | 3.92 | 3.48 | 2.51 | 1.46 | |
| 315 | 94.23 | 107.49 | 5.58 | 5.90 | 16.61 | 22.29 | 7.71 | 67.27 | 0.55 | 19.39 | 66.86 | 52.57 | 62.37 | 18.00 | 36.55 | 24.66 | 28.10 | 18.81 | |
| Distanceb (cm) | 0 | 227 | 217 | 190 | 235 | 209 | 242 | 210 | 249 | 256 | 238 | 241 | 229 | 230 | 225 | 230 | 210 | 240 | 230 |
| 45 | 225 | 238 | 300 | 235 | 200 | 200 | 258 | 197 | 215 | 280 | 220 | 296 | 270 | 175 | 194 | 300 | 336 | 185 | |
| 90 | 225 | 214 | 275 | 360 | 198 | 297 | 175 | 200 | 158 | 203 | 302 | 221 | 200 | 160 | 213 | 248 | 296 | 210 | |
| 135 | 383 | 412 | 383 | 355 | 296 | 360 | 270 | 180 | 233 | 266 | 383 | 276 | 315 | 225 | 193 | 360 | 424 | 255 | |
| 180 | 435 | 417 | 494 | 391 | 473 | 362 | 390 | 366 | 380 | 334 | 417 | 439 | 390 | 360 | 390 | 420 | 416 | 420 | |
| 225 | 277 | 239 | 233 | 329 | 251 | 293 | 290 | 340 | 338 | 186 | 371 | 357 | 400 | 470 | 403 | 458 | 456 | 360 | |
| 270 | 300 | 300 | 250 | 214 | 180 | 233 | 215 | 348 | 185 | 180 | 267 | 296 | 255 | 185 | 240 | 330 | 320 | 210 | |
| 315 | 198 | 210 | 300 | 209 | 222 | 198 | 180 | 255 | 280 | 190 | 260 | 292 | 270 | 275 | 223 | 300 | 344 | 190 |
aDose was converted to 1 m from the isocenter
bDistance to the measurement position from the isocenter
cNot detected
Ratio of the calculated dose to the measured dose
| Point (°) | NCRP-DLP methoda/measured dose | J C methodb/measured dose | J C methodb/NCRP-DLP methoda | ||
|---|---|---|---|---|---|
| Mean | SD | Mean | SD | ||
| 0 | 3.2 | 0.8 | 21 | 16 | 6.5 |
| 45 | 4.3 | 4.1 | 25 | 15 | 5.8 |
| 90 | 55 | 24 | 404 | 340 | 7.4 |
| 135 | 1.7 | 0.6 | 11 | 8.7 | 6.9 |
| 180 | 6.5 | 2.0 | 43 | 37 | 6.6 |
| 225 | 1.9 | 0.6 | 12 | 8.3 | 6.4 |
| 270 | 42 | 23 | 270 | 195 | 6.4 |
| 315 | 4.6 | 4.3 | 26 | 16 | 5.5 |
| All | 15 | 21 | 102 | 150 | 6.4 |
SD standard deviation
aNCRP-DLP: National Council on Radiation Protection and Measurements, USA method utilizing Dose Length Product
bJC: Japanese conventional method
Reduction factorsa in the gantry and couch directions
| Direction | Point (°) | CT-1 | CT-2 | CT-3 | CT-4 | CT-5 | CT-6 | CT-7 | CT-8 | CT-9 | CT-10 | CT-11 | CT-12 | CT-13 | CT-14 | CT-15 | CT-16 | CT-17 | CT-18 | Mean | SD |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Gantry | 90 | 0.036 | 0.029 | 0.037 | 0.023 | 0.032 | 0.044 | 0.024 | 0.025 | 0.047 | 0.031 | 0.022 | 0.033 | 0.022 | 0.033 | 0.024 | 0.047 | 0.016 | 0.031 | 0.031 | 0.009 |
| 270 | 0.036 | 0.035 | 0.032 | 0.039 | 0.037 | 0.030 | 0.017 | 0.049 | 0.043 | 0.020 | 0.054 | 0.027 | 0.043 | 0.031 | 0.082 | 0.073 | 0.050 | 0.032 | 0.041 | 0.017 | |
| Couch | 180 | 0.218 | 0.143 | 0.178 | 0.193 | 0.280 | 0.160 | 0.253 | 0.203 | 0.261 | 0.252 | 0.316 | 0.180 | 0.188 | 0.239 | 0.313 | 0.320 | 0.261 | 0.355 | 0.240 | 0.061 |
SD standard deviation
aThe reduction factors were calculated as “dose at each direction/the highest dose” converting doses at 1 m from the isocenter shown in Table 3
Fig. 3Relationship between head ratio and NCRP-DLP/M ratio in the case of the 0° direction
Ratio of the doses calculated with the Japanese-DLP method and the NCRP-DLP method to the measured dose
| Point (°) | Japanese-DLP methoda/measured dose | NCRP-DLPb/measured dose | NCRP-DLP methodb/Japanese-DLP methoda | ||
|---|---|---|---|---|---|
| Mean | SD | Mean | SD | ||
| 0 | 6.4 | 1.6 | 3.2 | 0.8 | 0.5 |
| 45 | 8.5 | 8.2 | 4.3 | 4.1 | 0.5 |
| 90 | 11.0 | 4.8 | 55 | 24 | 5.0 |
| 135 | 3.3 | 1.1 | 1.7 | 0.6 | 0.5 |
| 180 | 5.2 | 1.6 | 6.5 | 2.0 | 1.3 |
| 225 | 3.8 | 1.2 | 1.9 | 0.6 | 0.5 |
| 270 | 8.4 | 4.5 | 42 | 23 | 5.0 |
| 315 | 9.3 | 8.6 | 4.6 | 4.3 | 0.5 |
| All | 7.0 | 2.8 | 15 | 21 | 2.1 |
SD standard deviation
aJapanese-DLP: Japanese-Dose Length Product method
bNCRP-DLP: National Council on Radiation Protection and Measurements, USA method utilizing Dose Length Product
Comparison of required thickness for shielding of the scanner “CT-13a” calculated by each method as a typical example
| Scanner | Point (°) | Evaluated dose (mSv/3 M)b | Ratio of JCd to Measured | Required shielding ratiof | Required thickness for shielding (Pb, mm)g | JC-Measuredh
| |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| NCRP-DLPc | J Cd | Japanese-DLPe | Measured | NCRP-DLPc | J Cd | Japanese-DLPe | Measured | NCRP-DLPc | J Cd | Japanese- DLPe | Measured | ||||
| CT-13 | 0 | 2.8E+02 | 5.0E+03 | 5.7E+02 | 7.4E+01 | 6.8E+01 | 4.6E−03 | 2.6E−04 | 2.3E−03 | 1.8E−02 | 1.3 | 2.5 | 1.6 | 0.9 | 1.6 |
| 45 | 2.1E+02 | 3.6E+03 | 4.1E+02 | 1.2E+02 | 3.0E+01 | 6.3E−03 | 3.6E−04 | 3.2E−03 | 1.1E−02 | 1.2 | 2.4 | 1.5 | 1.0 | 1.4 | |
| 90 | 3.8E+02 | 6.7E+03 | 7.5E+01 | 4.7E+00 | 1.4E+03 | 3.5E−03 | 1.9E−04 | 1.7E−02 | 2.7E−01 | 1.5 | 2.7 | 0.9 | 0.2 | 2.5 | |
| 135 | 1.5E+02 | 2.7E+03 | 3.0E+02 | 7.2E+01 | 3.7E+01 | 8.6E−03 | 4.9E−04 | 4.3E−03 | 1.8E−02 | 1.1 | 2.3 | 1.4 | 0.9 | 1.4 | |
| 180 | 9.9E+01 | 1.7E+03 | 7.9E+01 | 1.1E+01 | 1.6E+02 | 1.3E−02 | 7.5E−04 | 1.6E−02 | 1.2E−01 | 1.0 | 2.1 | 0.9 | 0.4 | 1.7 | |
| 225 | 9.4E+01 | 1.7E+03 | 1.9E+02 | 4.6E+01 | 3.6E+01 | 1.4E−02 | 7.8E−04 | 6.9E−03 | 2.8E−02 | 0.9 | 2.1 | 1.2 | 0.7 | 1.4 | |
| 270 | 2.3E+02 | 4.1E+03 | 4.6E+01 | 5.8E+00 | 7.1E+02 | 5.6E−03 | 3.2E−04 | 2.8E−02 | 2.2E−01 | 1.3 | 2.4 | 0.7 | 0.2 | 2.2 | |
| 315 | 2.1E+02 | 3.6E+03 | 4.1E+02 | 1.1E+02 | 3.3E+01 | 6.3E−03 | 3.6E−04 | 3.2E−03 | 1.2E−02 | 1.2 | 2.4 | 1.5 | 1.0 | 1.4 | |
| Mean | 7.7E−03 | 4.4E−04 | 1.0E−02 | 8.8E−02 | 1.2 | 2.4 | 1.2 | 0.7 | 1.7 | ||||||
| SD | 3.9E−03 | 2.2E−04 | 9.4E−03 | 1.1E−01 | 0.2 | 0.2 | 0.3 | 0.3 | 0.4 | ||||||
SD standard deviation
aThe detailed of the scanner “CT-13” is described in Table 2
bDose was converted from a week dose to three months dose at evaluated points
cNCRP-DLP: The method using Dose Length Product described in National Council on Radiation Protection and Measurements-Dose Length Product
dJC: Japanese conventional calculation method
eJapanese-DLP: Japanese-Dose Length Product method
fRequired Shielding ratio: Minimum transmission for shielding at the boundary of controlled area [below 1.3 (mSv/3 M)]
gRequired thickness for shielding: Minimum thickness of shielding to achieve the standard for the boundary of controlled area (below 1.3 (mSv/3 M))
hJC-Measured: JC-measured was calculated as “required thickness for shielding by JC”−“required thickness for shielding by measured”
Comparison of shielding calculation methods
| Methods | NCRP-DLPa [ | NCRP-DLPa [ | Japanese-DLPb (present study) | Japanese conventionalc (present study) |
|---|---|---|---|---|
| Number of manufacturers | 3 | 4 | 4 | 4 |
| Number of facilities | 3 | Do not show | 12 | 11 |
| Number of scanners | 3 | 4 | 18 | 16 |
| Research type | Clinical | Rando phantom | Clinical | Clinical |
| Minimum calculated/measured ratio | 0.7 | Do not show | 0.6 | 3.5 |
| Percentage of underestimation except gantry and couch directions | 22.2 | Do not show | 3.4 | 0.0 |
| Reduction ratios of gantry | No | Yes | Yes | No |
| Reduction ratios of self-shield (couch direction) | No | No | Yes | No |
| Major parameter | DLP | DLP | DLP | mAs |
aNCRP-DLP: National Council on Radiation Protection and Measurements, USA method utilizing Dose Length Product
bJapanese-DLP: Japanese-Dose Length Product method
cJapanese conventional: Japanese conventional method