Use of self-expanding stents for better intracranial flow diverter wall apposition.

Anna Luisa Kühn1, Ajay K Wakhloo1, Matthew J Gounis1, Peter Kan2, Katyucia de Macedo Rodrigues1, J Diego Lozano1, Miklos G Marosfoi1, Mary Perras1, Christopher Brooks1, Mary C Howk1, David E Rex1, Francesco Massari1, Ajit S Puri1.   

Abstract

Background Flow diverter (FD) malapposition is associated with stroke-related complications. We document the use of self-expanding nitinol stents to remove/reduce the ledge of a FD deployed for aneurysm treatment. Methods We identified five patients who were treated with the Pipeline embolization device (PED) in conjunction with a Neuroform EZ stent for inadequate wall apposition of the ends of the FD at our institution between May 2014 and July 2015. Among other parameters, angiographic results, cone-beam computed tomography assessment of wall apposition and patient clinical outcome were evaluated. Results Incomplete device end apposition was seen in three cases, and precarious positioning of the distal end of the PED over the aneurysm neck was seen in two cases. In all five cases, successful treatment with good wall apposition and proper pinning of the PED distal edge was achieved using an additional Neuroform EZ stent. Appropriate aneurysm neck coverage and flow stagnation was seen in all cases. The combination of high radial outward force and open-cell design permits the Neuroform EZ stent to jail the malappositioned edges of the FD while maintaining good vessel-wall apposition itself and prevent migration of the PED. Short-term follow-up angiography showed device patency and complete aneurysm obliteration in all cases. Conclusions Preliminary results of this small case series suggest that the Neuroform EZ stent allows for effective treatment of FD malapposition in selected patients amenable for this endovascular approach. Long-term and larger cohort studies are needed to validate these results.

Entities:  

Keywords:  Aneurysm; flow diverter; malapposition; open-cell stent; radial force

Mesh:

Substances:

Year:  2016        PMID: 27956518      PMCID: PMC5433601          DOI: 10.1177/1591019916681981

Source DB:  PubMed          Journal:  Interv Neuroradiol        ISSN: 1591-0199            Impact factor:   1.610


  10 in total

1.  Angiographic CT in cerebrovascular stenting.

Authors:  Götz Benndorf; Charles M Strother; Benjamin Claus; Ramin Naeini; Hesham Morsi; Richard Klucznik; Michael E Mawad
Journal:  AJNR Am J Neuroradiol       Date:  2005-08       Impact factor: 3.825

Review 2.  Endovascular treatment of intracranial aneurysms: current status.

Authors:  Laurent Pierot; Ajay K Wakhloo
Journal:  Stroke       Date:  2013-07       Impact factor: 7.914

3.  Use of concentric Solitaire stent to anchor Pipeline flow diverter constructs in treatment of shallow cervical carotid dissecting pseudoaneurysms.

Authors:  Jason P Rahal; Venkata S Dandamudi; Robert S Heller; Mina G Safain; Adel M Malek
Journal:  J Clin Neurosci       Date:  2013-11-11       Impact factor: 1.961

4.  Utility of VasoCT in the treatment of intracranial aneurysm with flow-diverter stents.

Authors:  Osman Kizilkilic; Naci Kocer; George Emmanuel Metaxas; Drazenko Babic; Robert Homan; Civan Islak
Journal:  J Neurosurg       Date:  2012-05-04       Impact factor: 5.115

5.  Flow-diverter stent for the endovascular treatment of intracranial aneurysms: a prospective study in 29 patients with 34 aneurysms.

Authors:  Boris Lubicz; Laurent Collignon; Gaï Raphaeli; Jean-Pierre Pruvo; Michaël Bruneau; Olivier De Witte; Xavier Leclerc
Journal:  Stroke       Date:  2010-08-26       Impact factor: 7.914

6.  Delayed complications after flow-diverter stenting: reactive in-stent stenosis and creeping stents.

Authors:  José E Cohen; John Moshe Gomori; Samuel Moscovici; Ronen R Leker; Eyal Itshayek
Journal:  J Clin Neurosci       Date:  2013-12-07       Impact factor: 1.961

7.  Incomplete stent apposition in Enterprise stent-mediated coiling of aneurysms: persistence over time and risk of delayed ischemic events.

Authors:  Robert Heller; Daniel R Calnan; Michael Lanfranchi; Neel Madan; Adel M Malek
Journal:  J Neurosurg       Date:  2013-03-15       Impact factor: 5.115

8.  Spontaneous delayed migration/shortening of the pipeline embolization device: report of 5 cases.

Authors:  N Chalouhi; S I Tjoumakaris; L F Gonzalez; D Hasan; P J Pema; G Gould; R H Rosenwasser; P M Jabbour
Journal:  AJNR Am J Neuroradiol       Date:  2013-06-27       Impact factor: 3.825

9.  Delayed migration of a pipeline embolization device.

Authors:  Nohra Chalouhi; Sudhakar R Satti; Stavropoula Tjoumakaris; Aaron S Dumont; L Fernando Gonzalez; Robert Rosenwasser; Pascal Jabbour
Journal:  Neurosurgery       Date:  2013-06       Impact factor: 4.654

10.  Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia.

Authors:  Thomas F Flood; Imramsjah M J van der Bom; Lara Strittmatter; Ajit S Puri; Gregory M Hendricks; Ajay K Wakhloo; Matthew J Gounis
Journal:  J Neurointerv Surg       Date:  2014-01-30       Impact factor: 5.836

  10 in total
  5 in total

1.  Filament inadequate wall apposition of the different ends of flow diverters in the abdominal aorta of rabbits.

Authors:  Kuizhong Wang; Peng Jin; Peigang Lu; Qiang Liu; Bo Li; Zhenqiang Hao
Journal:  Neuroradiol J       Date:  2019-11-27

2.  Further Development of Combined Techniques Using Stent Retrievers, Aspiration Catheters and BGC : The PROTECTPLUS Technique.

Authors:  Christian Maegerlein; Maria Teresa Berndt; Sebastian Mönch; Kornelia Kreiser; Tobias Boeckh-Behrens; Manuel Lehm; Silke Wunderlich; Claus Zimmer; Benjamin Friedrich
Journal:  Clin Neuroradiol       Date:  2018-11-09       Impact factor: 3.649

3.  The Exchange-Free Technique: A Novel Technique for Enhancing Surpass Flow Diverter Placement.

Authors:  Osman Ocal; Anıl Arat
Journal:  Asian J Neurosurg       Date:  2020-08-28

4.  Placement of a Stent within a Flow Diverter Improves Aneurysm Occlusion Rates.

Authors:  O Ocal; A Peker; S Balci; A Arat
Journal:  AJNR Am J Neuroradiol       Date:  2019-10-03       Impact factor: 3.825

5.  Acute thrombus formation on phosphorilcholine surface modified flow diverters.

Authors:  Miklos Marosfoi; Frederic Clarencon; Erin T Langan; Robert M King; Olivia W Brooks; Takamisu Tamura; John M Wainwright; Matthew J Gounis; Srinivasan Vedantham; Ajit S Puri
Journal:  J Neurointerv Surg       Date:  2017-07-08       Impact factor: 5.836

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.