Evaluating the impact of scan settings on automatic tube current modulation in CT using a novel phantom.

Deborah Merzan1, Patrik Nowik1, Gavin Poludniowski1, Robert Bujila1,2.   

Abstract

OBJECTIVE: The aim of this study was to make a comprehensive evaluation of how variable scan settings can affect the performance of automatic tube current modulation (ATCM) in recent CT scanners from the four major manufacturers.
METHODS: A phantom was designed and manufactured for the purpose of evaluating ATCM. The phantom was scanned with four categories of systematically varied settings (scan projection radiograph, technique and reconstruction parameters and phantom miscentring). The performance of ATCM, in terms of applied tube current and noise uniformity, for the scans with varied settings was compared with a reference scan using subjective and quantitative approaches.
RESULTS: The ATCM implemented by each manufacturer is based on different principles and any affect to the performance of the ATCM, when varying scan settings, will manifest differently among the vendors. The results are summarized in four tables corresponding to the categories of varied settings.
CONCLUSION: The developed phantom proved useful for evaluating the ATCM. It is important to understand how different implementations (vendor specific) of ATCM perform in order to make informed decisions about the selection of scan settings when designing protocols. The resulting tables can serve as a reference for understanding the different implementations of ATCM and highlight settings that should be taken into consideration when adjusting an imaging protocol. Advances in knowledge: The results from this work can serve as a reference for how changes in geometry or scan settings can affect the performance of ATCM, in terms of tube current and noise.

Mesh:

Year:  2016        PMID: 27845559      PMCID: PMC5605012          DOI: 10.1259/bjr.20160308

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  16 in total

1.  [Evaluation and routine testing in medical imaging departments - Part3-3: Acceptance tests - imaging performance of X-ray equipment for digital subtraction angiography (DSA) JIS Z 4752-3-3:2003].

Authors:  Fuminori Shinohara; Masaharu Abe; Shigeru Miyaxzaki
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2003-05

2.  Comparison of different phantom designs for CT scanner automatic tube current modulation system tests.

Authors:  S Sookpeng; C J Martin; D J Gentle
Journal:  J Radiol Prot       Date:  2013-09-11       Impact factor: 1.394

3.  The influences of tube voltage and scan direction on combined tube current modulation: a phantom study.

Authors:  Hyun Woo Goo; Dong Soo Suh
Journal:  Pediatr Radiol       Date:  2006-04-27

4.  Impact of operator-selected image noise index and reconstruction slice thickness on patient radiation dose in 64-MDCT.

Authors:  Kalpana M Kanal; Brent K Stewart; Orpheus Kolokythas; William P Shuman
Journal:  AJR Am J Roentgenol       Date:  2007-07       Impact factor: 3.959

5.  [Performance evaluation for CT-AEC(CT automatic exposure control)systems].

Authors:  Yoshihisa Muramatsu; Shu Ikeda; Kazuaki Osawa; Ryo Sekine; Nobuyuki Niwa; Masami Terada; Nicholas Keat; Shigeru Miyazaki
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2007-05-20

6.  Utilizing a simple CT dosimetry phantom for the comprehension of the operational characteristics of CT AEC systems.

Authors:  Ioannis A Tsalafoutas; Athanasios Varsamidis; Stella Thalassinou; Efstathios P Efstathopoulos
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

Review 7.  Radiation dose modulation techniques in the multidetector CT era: from basics to practice.

Authors:  Chang Hyun Lee; Jin Mo Goo; Hyun Ju Ye; Sung-Joon Ye; Chang Min Park; Eun Ju Chun; Jung-Gi Im
Journal:  Radiographics       Date:  2008 Sep-Oct       Impact factor: 5.333

8.  A methodology for image quality evaluation of advanced CT systems.

Authors:  Joshua M Wilson; Olav I Christianson; Samuel Richard; Ehsan Samei
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

9.  Automatic exposure control in multichannel CT with tube current modulation to achieve a constant level of image noise: experimental assessment on pediatric phantoms.

Authors:  Hervé J Brisse; Ludovic Madec; Geneviève Gaboriaud; Thomas Lemoine; Alexia Savignoni; Sylvia Neuenschwander; Bernard Aubert; Jean-Claude Rosenwald
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

10.  Investigation of the influence of image reconstruction filter and scan parameters on operation of automatic tube current modulation systems for different CT scanners.

Authors:  Supawitoo Sookpeng; Colin J Martin; David J Gentle
Journal:  Radiat Prot Dosimetry       Date:  2014-08-08       Impact factor: 0.972

View more
  4 in total

1.  Organ doses from CT localizer radiographs: Development, validation, and application of a Monte Carlo estimation technique.

Authors:  Jocelyn Hoye; Shobhit Sharma; Yakun Zhang; Wanyi Fu; Francesco Ria; Anuj Kapadia; W Paul Segars; Joshua Wilson; Ehsan Samei
Journal:  Med Phys       Date:  2019-09-16       Impact factor: 4.071

2.  The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT equipments.

Authors:  Touko Kaasalainen; Teemu Mäkelä; Mika Kortesniemi
Journal:  Eur J Radiol Open       Date:  2018-12-17

3.  0° vs. 180° CT localiser: The effect of vertical off-centring, phantom positioning and tube voltage on dose optimisation in multidetector computed tomography.

Authors:  Yazan Al-Hayek; Xiaoming Zheng; Rob Davidson; Christopher Hayre; Dana Al-Mousa; Campbell Finlay; Kelly Spuur
Journal:  J Med Radiat Sci       Date:  2021-08-17

Review 4.  Best practice for the nuclear medicine technologist in CT-based attenuation correction and calcium score for nuclear cardiology.

Authors:  Luca Camoni; Andrea Santos; Marieclaire Attard; Marius Ovidiu Mada; Agata Karolina Pietrzak; Sonja Rac; Sebastijan Rep; Christelle Terwinghe; Pedro Fragoso Costa
Journal:  Eur J Hybrid Imaging       Date:  2020-07-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.