Comparison of different phantom designs for CT scanner automatic tube current modulation system tests.

S Sookpeng1, C J Martin, D J Gentle.   

Abstract

Modern CT scanners modulate tube current during scans according to patient size, shape and attenuation. However, the ATCM (automatic tube current modulation) systems for different CT manufacturers work on different principles. Although the systems are used for the majority of patients and examinations, there is no standard phantom for routine quality control of CT scanner ATCM operation. The ideal phantom for testing these systems should be capable of evaluating how tube current and image quality as well as dose vary according to changes in patient size and shape. For this study, a conical phantom designed by ImPACT has been compared with two phantoms made from elliptical sections with varying dimensions. The concept of the designs is to reflect the ATCM performance for the varying shapes and dimensions along the length of the human body. The first phantom comprises five elliptical sections with a wide range of different dimensions and the second has three sections that are more similar in size. The phantoms have been used to test ATCM systems for Philips, Siemens, GE and Toshiba scanners. Although the results of the tube current modulation patterns were similar for all CT scanners, the abrupt changes in attenuation for the first sectional phantom provoked an abnormal ATCM response for the GE and Toshiba scanners. The second sectional phantom was developed from the results of the first, and was more effective for ATCM system testing and could be used for dose and image quality assessment in standard positions. However, the ImPACT conical phantom provided the best overall assessment of performance in terms of tube current modulations and noise pattern.

Entities:  

Mesh:

Year:  2013        PMID: 24025482     DOI: 10.1088/0952-4746/33/4/735

Source DB:  PubMed          Journal:  J Radiol Prot        ISSN: 0952-4746            Impact factor:   1.394


  6 in total

1.  Practical dose point-based methods to characterize dose distribution in a stationary elliptical body phantom for a cone-beam C-arm CT system.

Authors:  Jang-Hwan Choi; Dragos Constantin; Arundhuti Ganguly; Erin Girard; Richard L Morin; Robert L Dixon; Rebecca Fahrig
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

2.  Evaluating the impact of scan settings on automatic tube current modulation in CT using a novel phantom.

Authors:  Deborah Merzan; Patrik Nowik; Gavin Poludniowski; Robert Bujila
Journal:  Br J Radiol       Date:  2016-11-15       Impact factor: 3.039

3.  Low tube voltage and low contrast material volume cerebral CT angiography.

Authors:  Song Luo; Long Jiang Zhang; Felix G Meinel; Chang Sheng Zhou; Li Qi; Andrew D McQuiston; U Joseph Schoepf; Guang Ming Lu
Journal:  Eur Radiol       Date:  2014-05-04       Impact factor: 5.315

4.  Dosimetric changes with computed tomography automatic tube-current modulation techniques.

Authors:  Sofia Spampinato; Anna Maria Gueli; Pietro Milone; Luigi Angelo Raffaele
Journal:  Radiol Phys Technol       Date:  2018-04-06

5.  Automatic current selection with iterative reconstruction reduces effective dose to less than 1 mSv in low-dose chest computed tomography in persons with normal BMI.

Authors:  Li-Guo Chen; Ping-An Wu; Min-Huei Sheu; Hsing-Yang Tu; Li-Chuan Huang
Journal:  Medicine (Baltimore)       Date:  2019-07       Impact factor: 1.817

6.  A Projection Quality-Driven Tube Current Modulation Method in Cone-Beam CT for IGRT: Proof of Concept.

Authors:  Kuo Men; Jianrong Dai
Journal:  Technol Cancer Res Treat       Date:  2017-11-12
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.