Investigation of the influence of image reconstruction filter and scan parameters on operation of automatic tube current modulation systems for different CT scanners.

Supawitoo Sookpeng1, Colin J Martin2, David J Gentle3.   

Abstract

Variation in the user selected CT scanning parameters under automatic tube current modulation (ATCM) between hospitals has a substantial influence on the radiation doses and image quality for patients. The aim of this study was to investigate the effect of changing image reconstruction filter and scan parameter settings on tube current, dose and image quality for various CT scanners operating under ATCM. The scan parameters varied were pitch factor, rotation time, collimator configuration, kVp, image thickness and image filter convolution (FC) used for reconstruction. The Toshiba scanner varies the tube current to achieve a set target noise. Changes in the FC setting and image thickness for the first reconstruction were the major factors affecting patient dose. A two-step change in FC from smoother to sharper filters doubles the dose, but is counterbalanced by an improvement in spatial resolution. In contrast, Philips and Siemens scanners maintained tube current values similar to those for a reference image and patient, and the tube current only varied slightly for changes in individual CT scan parameters. The selection of a sharp filter increased the image noise, while use of iDose iterative reconstruction reduced the noise. Since the principles used by CT manufacturers for ATCM vary, it is important that parameters which affect patient dose and image quality for each scanner are made clear to operator to aid in optimisation.
© The Author 2014. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Mesh:

Year:  2014        PMID: 25107439     DOI: 10.1093/rpd/ncu236

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  4 in total

1.  Evaluating the impact of scan settings on automatic tube current modulation in CT using a novel phantom.

Authors:  Deborah Merzan; Patrik Nowik; Gavin Poludniowski; Robert Bujila
Journal:  Br J Radiol       Date:  2016-11-15       Impact factor: 3.039

2.  The investigation of dose and image quality of chest computed tomography using different combinations of noise index and adaptive statistic iterative reconstruction level.

Authors:  Supawitoo Sookpeng; Colin J Martin; Chitsanupong Butdee
Journal:  Indian J Radiol Imaging       Date:  2019 Jan-Mar

3.  A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre-A phantom study.

Authors:  Patrizio Barca; Fabio Paolicchi; Giacomo Aringhieri; Federica Palmas; Daniela Marfisi; Maria Evelina Fantacci; Davide Caramella; Marco Giannelli
Journal:  PLoS One       Date:  2021-01-14       Impact factor: 3.240

4.  The effect of variations in CT scan protocol on femoral finite element failure load assessment using phantomless calibration.

Authors:  Ali Ataei; Jelle Eikhout; Ruud G H van Leeuwen; Esther Tanck; Florieke Eggermont
Journal:  PLoS One       Date:  2022-03-18       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.