Literature DB >> 27815591

Does vitamin E-blended polyethylene reduce wear in primary total hip arthroplasty: a blinded randomised clinical trial.

Caroline Scemama1, Philippe Anract2, Valérie Dumaine2, Antoine Babinet2, Jean Pierre Courpied2, Moussa Hamadouche2.   

Abstract

PURPOSE: Some data indicate that first-generation highly cross-linked polyethylene (HXLPE) can oxidise in vivo and is associated with reduced mechanical properties. To overcome these limitations, a natural anti-oxidant vitamin E has been added to HXLPE to preserve the mechanical properties and decrease oxidative degradation whilst conserving high wear resistance. We hypothesised that after a minimal three years of follow-up the use of vitamin E-blended HXLPE would result in lower radiographic wear when compared with ultra-high molecular weight polyethylene (UHMWPE).
METHODS: One hundred patients were randomised to receive hybrid total hip arthroplasty (THA) using a monoblock cementless acetabular component made either of UHMWPE or vitamin E-blended HXLPE. All other parameters were identical in both groups. Complete follow-up was available for 74 of these patients. Femoral head penetration was measured using a validated computer-assisted method.
RESULTS: The median creep measured 0.111 mm (range, -0.576 - +0.444 mm) in the vitamin E-blended group versus 0.170 mm (range, -0.861 - +0.884 mm) in the UHMWPE group (difference of medians, 0.059; p = 0.046). The median steady state penetration rate was -0.008 mm/year (range, -0.88 - +0.64 mm/year) in the vitamin E-blended group versus 0.133 mm/year (range, -0.84 - +0.85 mm/year) in the UHMWPE group (difference of medians 0.141, p = 0.043).
CONCLUSIONS: This study demonstrated that femoral head penetration was lower when using vitamin E-blended HXLPE when compared with UHMWPE, with a steady-state penetration rate far below the osteolysis threshold. Longer-term follow-up is needed to warrant whether wear reduction will generate less occurrence of osteolysis and aseptic loosening.

Entities:  

Keywords:  HXLPE; Polyethylene; Total hip arthroplasty; UHMWPE; Vitamin E; Wear

Mesh:

Substances:

Year:  2016        PMID: 27815591     DOI: 10.1007/s00264-016-3320-2

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  26 in total

1.  The effect of alpha-tocopherol on the oxidation and free radical decay in irradiated UHMWPE.

Authors:  Ebru Oral; Shannon L Rowell; Orhun K Muratoglu
Journal:  Biomaterials       Date:  2006-08-02       Impact factor: 12.479

2.  In vivo oxidation of retrieved cross-linked ultra-high-molecular-weight polyethylene acetabular components with residual free radicals.

Authors:  Keith K Wannomae; Shayan Bhattacharyya; Andrew Freiberg; Daniel Estok; William H Harris; Orhun Muratoglu
Journal:  J Arthroplasty       Date:  2006-07-17       Impact factor: 4.757

3.  A double-blind, prospective, randomized controlled trial comparing highly cross-linked and conventional polyethylene in primary total hip arthroplasty.

Authors:  George T Calvert; Peter A Devane; Jan Fielden; Kathryn Adams; J Geoffrey Horne
Journal:  J Arthroplasty       Date:  2008-06-10       Impact factor: 4.757

4.  Determination of polyethylene wear in total hip replacements with use of digital radiographs.

Authors:  J M Martell; S Berdia
Journal:  J Bone Joint Surg Am       Date:  1997-11       Impact factor: 5.284

5.  Femoral Head Penetration of Vitamin E-Infused Highly Cross-Linked Polyethylene Liners: A Randomized Radiostereometric Study of Seventy Hips Followed for Two Years.

Authors:  Bita Shareghi; Per-Erik Johanson; Johan Kärrholm
Journal:  J Bone Joint Surg Am       Date:  2015-08-19       Impact factor: 5.284

6.  Poor performance of Enduron polyethylene liner in total hip arthroplasty: a minimum ten-year follow up and ultra-morphological analysis of wear particles.

Authors:  Yufei Yan; Hao Chen; Jianmin Feng; Kaizhe Chen; Kaidi Zhou; Weixiang Hong; Yi Wang; Zhihong Liu; Jiong Zhang; Qingming Yang; Lei Guo; Chuan He
Journal:  Int Orthop       Date:  2016-05-13       Impact factor: 3.075

7.  In vivo wear and migration of highly cross-linked polyethylene cups a radiostereometry analysis study.

Authors:  Stephan Röhrl; Bo Nivbrant; Li Mingguo; Ben Hewitt
Journal:  J Arthroplasty       Date:  2005-06       Impact factor: 4.757

Review 8.  History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene.

Authors:  Steven M Kurtz; Heather A Gawel; Jasmine D Patel
Journal:  Clin Orthop Relat Res       Date:  2011-08       Impact factor: 4.176

9.  A comparison of the penetration rate of two polyethylene acetabular liners of different levels of cross-linking. A prospective randomised trial.

Authors:  P Triclot; G Grosjean; F El Masri; J P Courpied; M Hamadouche
Journal:  J Bone Joint Surg Br       Date:  2007-11

10.  A randomised controlled trial comparing highly cross-linked and contemporary annealed polyethylene after a minimal eight-year follow-up in total hip arthroplasty using cemented acetabular components.

Authors:  J Langlois; F Atlan; C Scemama; J P Courpied; M Hamadouche
Journal:  Bone Joint J       Date:  2015-11       Impact factor: 5.082

View more
  12 in total

Review 1.  Acetabular reinforcement rings associated with allograft for severe acetabular defects.

Authors:  Emmanuel Gibon; Luc Kerboull; Jean-Pierre Courpied; Moussa Hamadouche
Journal:  Int Orthop       Date:  2018-09-13       Impact factor: 3.075

2.  Reasons for Revision, Oxidation, and Damage Mechanisms of Retrieved Vitamin E-Stabilized Highly Crosslinked Polyethylene in Total Knee Arthroplasty.

Authors:  Hannah Spece; Jaclyn T Schachtner; Daniel W MacDonald; Gregg R Klein; Michael A Mont; Gwo-Chin Lee; Steven M Kurtz
Journal:  J Arthroplasty       Date:  2019-07-18       Impact factor: 4.757

3.  Implant survival of 2,723 vitamin E-infused highly crosslinked polyethylene liners in total hip arthroplasty: data from the Finnish Arthroplasty Register.

Authors:  Matias Hemmilä; Inari Laaksonen; Markus Matilainen; Antti Eskelinen; Jaason Haapakoski; Ari-Pekka Puhto; Jukka Kettunen; Konsta Pamilo; Keijo T Mäkelä
Journal:  Acta Orthop       Date:  2021-02-01       Impact factor: 3.717

Review 4.  Ultra-High Molecular Weight Polyethylene: Influence of the Chemical, Physical and Mechanical Properties on the Wear Behavior. A Review.

Authors:  Pierangiola Bracco; Anuj Bellare; Alessandro Bistolfi; Saverio Affatato
Journal:  Materials (Basel)       Date:  2017-07-13       Impact factor: 3.623

5.  Quantitative Measurements of Backside Wear in Acetabular Hip Joint Replacement: Conventional Polyethylene Versus Cross-Linked Polyethylene.

Authors:  Steffen Braun; Sebastian Jaeger; Robert Sonntag; Stefan Schroeder; J Philippe Kretzer
Journal:  Materials (Basel)       Date:  2020-04-15       Impact factor: 3.623

6.  Is Procalcitonin (PCT) a reliable biomarker for preoperative diagnosing of low grade periprosthetic joint infection? A prospective study.

Authors:  André Busch; Marcus Jäger; Harald Engler; Marcel Haversath; Charlotte Bielefeld; Stefan Landgraeber; Alexander Wegner
Journal:  BMC Musculoskelet Disord       Date:  2020-04-20       Impact factor: 2.362

7.  The concept of a cementless isoelastic monoblock cup made of highly cross-linked polyethylene infused with vitamin E: radiological analyses of migration and wear using EBRA and clinical outcomes at mid-term follow-up.

Authors:  Yama Afghanyar; Sebastian Joser; Jonas Tecle; Philipp Drees; Jens Dargel; Philipp Rehbein; Karl Philipp Kutzner
Journal:  BMC Musculoskelet Disord       Date:  2021-01-23       Impact factor: 2.362

8.  A vitamin E blended highly cross-linked polyethylene acetabular cup results in less wear: 6-year results of a randomized controlled trial in 199 patients.

Authors:  Julie R A Massier; Joost H J Van Erp; Thom E Snijders; Arthur DE Gast
Journal:  Acta Orthop       Date:  2020-08-24       Impact factor: 3.717

9.  2-year results of an RCT of 2 uncemented isoelastic monoblock acetabular components: lower wear rate with vitamin E blended highly cross-linked polyethylene compared to ultra-high molecular weight polyethylene.

Authors:  Joost H J van Erp; Julie R A Massier; Jelle J Halma; Thom E Snijders; Arthur de Gast
Journal:  Acta Orthop       Date:  2020-02-26       Impact factor: 3.717

Review 10.  Overview of Randomized Controlled Trials in Primary Total Hip Arthroplasty (34,020 Patients): What Have We Learnt?

Authors:  Hosam E Matar; Simon R Platt; Tim N Board; Martyn L Porter
Journal:  J Am Acad Orthop Surg Glob Res Rev       Date:  2020-08
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.