| Literature DB >> 28900367 |
Peifeng Ke1, Jiawei Liu1,2, Yan Chao1, Xiaobin Wu1, Yujuan Xiong1, Li Lin1, Zemin Wan1, Xinzhong Wu1, Jianhua Xu1, Junhua Zhuang1, Xianzhang Huang1.
Abstract
INTRODUCTION: Thalassemia could interfere with some assays for haemoglobin A1c (HbA1c) measurement, therefore, it is useful to be able to screen for thalassemia while measuring HbA1c. We used Capillarys 2 Flex Piercing (Capillarys 2FP) HbA1c programme to simultaneously measure HbA1c and screen for thalassemia.Entities:
Keywords: capillary electrophoresis (CE); haemoglobin A1c; haemoglobin A2; thalassemia
Mesh:
Substances:
Year: 2017 PMID: 28900367 PMCID: PMC5575649 DOI: 10.11613/BM.2017.030704
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Figure 1Haemoglobin A1c analysis by Capillarys 2 Flex Piercing. (A) Normal control. (B) HbH samples - HbA1c cannot be measured (in parenthesis) due to overlapping with the HbBart’s fraction.
Figure 2Bland-Altman plot showing the differences in haemoglobin A1c results between Capillarys 2 Flex Piercing (Capillarys 2FP) and Premier Hb9210 systems in thalassaemia samples (NGSP units, %). Solid line (mean) – mean difference. Dotted lines - 95% confidence interval of mean difference. Dashed lines (± 1.96 SD) - ± 1.96 standard deviation of mean difference.
HbA1c concentrations obtained by IFCC and Capillarys 2FP
| 1 | -α3.7/aa | 39 | 5.7 | 41 | 5.9 | 3.5 |
| 2 | -α4.2/aa | 40 | 5.8 | 42 | 6.0 | 3.4 |
| 3 | --SEA/aa | 30 | 4.9 | 28 | 4.7 | - 4.1 |
| 4 | --SEA/aa | 29 | 4.8 | 29 | 4.8 | 0.0 |
| 5 | -α3.7/--SEA | 21 | 4.1 | 23 | 4.3 | 4.9 |
| 6 | -α3.7/--SEA | 31 | 5.0 | 32 | 5.1 | 2.0 |
| 7 | CD41-42 | 31 | 5.0 | 30 | 4.9 | - 2.0 |
| 8 | CD41-42 | 38 | 5.6 | 40 | 5.8 | 3.6 |
| HbA1c - haemoglobin A1c. IFCC - International Federation of Clinical Chemistry and Laboratory Medicine. Capillarys 2FP - Capillarys 2 Flex Piercing. NGSP - National Glycohemoglobin Standardization Programme. The relative bias was calculated in NGSP units, as the Capillarys 2FP value against the IFCC reference method value. The proficiency testing acceptance limit ± 6% of CAP was set as the accuracy limit ( | ||||||
Figure 3The values of haemoglobin A2 (HbA2) were analyzed by Capillarys 2 Flex Piercing for normal controls and thalassaemia. (A) The box-plots show the frequency of HbA2 in normal control, α-thalassaemia silent/trait and β-thalassaemia trait. The horizontal line in each box represents the median values. The upper and lower limits of each box correspond to the 25th and 75th percentile values. The highest and lowest horizontal bars represent the minimum and maximum values. (B) For screening α-thalassaemia silent/trait, the receiver operating characteristic curve (ROC) shows an area under curve (AUC) of 0.80 (95% confidence interval, 0.78 - 0.84) and cutoff value ≤ 2.2% (sensitivity 76.9%, specificity 73.7%). (C) For screening β-thalassaemia trait, ROC shows AUC of 1.00 (95% confidence interval, 0.99 - 1.00) and cutoff value > 2.8% (sensitivity 100%, specificity 100%).