Uppermost blood levels of the right and left atria in the supine position: implication for measuring central venous pressure and pulmonary artery wedge pressure.

Jeong-Hwa Seo1, Chul-Woo Jung, Jae-Hyon Bahk.   

Abstract

BACKGROUND: To eliminate the influence of hydrostatic pressure, proper transducer positions for central venous pressure and pulmonary artery wedge pressure are at the uppermost blood levels of right atrium (RA) and left atrium (LA). This study was performed to investigate accurate reference levels of central venous pressure and pulmonary artery wedge pressure in the supine position.
METHODS: Chest computed tomography images of 96 patients without history of cardiothoracic surgery, heart disease, or cardiothoracic anatomical abnormality were retrospectively reviewed. The anteroposterior (AP) diameter of the thorax and the vertical distances from the skin on the back to the most anterior portion of RA (RA height) and LA (LA height) were measured. Their ratios were abbreviated, respectively, as RA height/AP diameter and LA height/AP diameter. Data are expressed as mean +/- SD (range).
RESULTS: There was a significant difference [4.6 +/- 1.0 (1.6-6.4) cm; P < 0.001] between RA and LA heights. AP diameter was positively correlated with RA and LA heights (R = 0.839 and 0.700, respectively; P < 0.001). There was also a significant difference between RA height/AP diameter [0.83 +/- 0.03 (0.71-0.91)] and LA height/AP diameter [0.62 +/- 0.04 (0.52-0.72)] (P < 0.001).
CONCLUSION: In the supine position, a central venous pressure transducer should be positioned approximately 4.6 cm higher than a pulmonary artery wedge pressure transducer. The external reference level for central venous pressure seems to be at approximately four fifths of the AP diameter of the thorax from the back, and that for pulmonary artery wedge pressure seems to be at approximately three fifths of the AP diameter.

Entities:  

Mesh:

Year:  2007        PMID: 17667570     DOI: 10.1097/01.anes.0000270720.67307.fa

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  3 in total

1.  Optimal level of the reference transducer for central venous pressure and pulmonary artery occlusion pressure monitoring in supine, prone, and sitting position.

Authors:  In-Kyung Song; Han-Seul Park; Ji-Hyun Lee; Eun-Hee Kim; Hee-Soo Kim; Jae-Hyon Bahk; Jin-Tae Kim
Journal:  J Clin Monit Comput       Date:  2016-03-21       Impact factor: 2.502

2.  Evaluation of external reference levels for central venous pressure measurements of severely obese patients in the supine position.

Authors:  Kazuyuki Mizunoya; Hitoshi Saito; Yuji Morimoto
Journal:  J Anesth       Date:  2018-05-28       Impact factor: 2.078

3.  Optimal Transducer Level for Atrial and Pulmonary Arterial Pressure Measurement in Patients with Functional Single Ventricle.

Authors:  Yong-Hee Park; Da-Hye Yoo; Eun-Hee Kim; In-Kyung Song; Ji-Hyun Lee; Hee-Soo Kim; Woong-Han Kim; Jin-Tae Kim
Journal:  Pediatr Cardiol       Date:  2016-10-01       Impact factor: 1.655

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.