Literature DB >> 27676162

Comparison between IEGM-based approach and echocardiography in AV/PV and VV delay optimization in CRT-D recipients (Quicksept study).

Massimo Giammaria1, Gianluca Quirino2, Enrico Cecchi3, Gaetano Senatore4, Paolo Pistelli5, Mario Bocchiardo6, Roberto Mureddu7, Paolo Diotallevi8, Eraldo Occhetta9, Andrea Magnani9, Mauro Bensoni10, Catia Checchinato11, Valentina Conti12, Sandra Badolati13, Antonio Mazza14.   

Abstract

BACKGROUND: AtrioVentricular (AV) and InterVentricular (VV) delay optimization can improve ventricular function in Cardiac Resynchronization Therapy (CRT) and is usually performed by means of echocardiography. St Jude Medical has developed an automated algorhythm which calculates the optimal AV and VV delays (QuickOpt™) based on Intracardiac ElectroGrams, (IEGM), within 2 min. So far, the efficacy of the algorhythm has been tested acutely with standard lead position at right ventricular (RV) apex. Aim of this project is to evaluate the algorhythm performance in the mid- and long-term with RV lead located in mid-septum.
METHODS: AV and VV delays optimization data were collected in 13 centers using both echocardiographic and QuickOpt™ guidance in CRTD implanted patients provided with this algorhythm. Measurements of the aortic Velocity Time Integral (aVTI) were performed with both methods in a random order at pre-discharge, 6-month and 12-month follow-up.
RESULTS: Fifty-three patients were studied (46 males; age 68 ± 10y; EF 28 ± 7%). Maximum aVTI obtained by echocardiography at different AV delays, were compared with aVTI acquired at AV delays suggested by QuickOpt. The AV Pearson correlations were 0.96 at pre-discharge, 0.95 and 0,98 at 6- and 12- month follow-up respectively. After programming optimal AV, the same approach was used to compare echocardiographic aVTI with aVTI corresponding to the VV values provided by QuickOpt. The VV Pearson Correlation were 0,92 at pre-discharge, 0,88 and 0.90 at 6-month and 12- month follow-up respectively.
CONCLUSIONS: IEGM-based optimization provides comparable results with echocardiographic method (maximum aVTI) used as reference with mid-septum RV lead location.
Copyright © 2016 Indian Heart Rhythm Society. Production and hosting by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Cardiac resinchronization therapy; Echocardiographyc optimization; IEGM based algorythm; Mid-septum stimulation; Optimization algorythm

Year:  2016        PMID: 27676162      PMCID: PMC5832617          DOI: 10.1016/j.ipej.2016.05.001

Source DB:  PubMed          Journal:  Indian Pacing Electrophysiol J        ISSN: 0972-6292


Background

Although Cardiac Resynchronization Therapy (CRT) has become a standard treatment in chronic heart failure patients with Left Ventricular (LV) systolic dysfunction, intraventricular conduction delay and NYHA class II to IV, despite optimal medical therapy [1], about 30–40% of patients still do not to respond to treatment [2], [3], [4], [5], [6], [7], [8]. The role of sensed and paced atrio-ventricular (PV/AV) and inter-ventricular (VV) delays optimization remains controversial. The 2013 ESC guidelines on CRT do not recommend a routine AV and VV optimization, which should be restricted to non-responders and to patients with ischemic heart disease [1], [2], [3], [4], [5], [6], [7], [8], [9], [10]. Optimization has been performed using echocardiography, which, however, is time- and skill-costly and can be hampered by a relevant intra- and inter-observer variability [11], [12], [13]. This fact has prompted a series of alternative methods for optimization, among which Intracardiac ElectroGraMs (IEGM)-based algorithms [14], [15], [16] have gained most attention. The QuickOpt™ (St Jude Medical, ST Paul, MN, USA) is an automatic IEGM-based algorithm incorporated in CRT devices, validated for optimizing AV and VV delays [14], [17], [18]. However, no serial assessment over time has been performed to test the relation between IEGM- and echo-generated optimal values of AV, PV and VV intervals. Furthermore, right ventricular (RV) lead has been predominantly located at the apex, and the QuickOpt™ reliability in “non-apical” RV pacing sites has not been highlighted yet. The QuickSept study has been designed to evaluate the efficacy of QuickOpt™ algorithm in a population of patients implanted with a CRT system with defibrillator capabilities (CRT-D) with the RV lead in a mid-septal position, at Hospital discharge at six-month and at one-year follow-up, by comparing AV and VV intervals calculated by IEGM with those determined by echocardiographic (ECHO) measurements.

Materials and methods

Population

Between January 2011 and January 2012, in 13 Italian Laboratories 53 patients were implanted with a CRT-D system according to current guidelines indications [19]. All patients gave written informed consent. Baseline patients’ features are shown in Table 1. Main inclusion criteria were: stable and safe placement of an active–fixation RV lead on mid-interventricular septum; achievement of an efficacious LV intravenous pacing from a Coronary Sinus (CS) branch. St Jude Medical (Saint Paul, Minnesota MN, USA) manufactured CRT-D systems with QuickOpt™ algorithm were implanted: models CD3211-36 Promote Plus, CD3215-36 Promote Accel, CD3239-40 Promote Quadra, 3251-40 Unify Quadra and 3235-40 Unify, with the single or dual-coil active fixation leads Durata mod. 7120Q and 7122Q. Enrolment was accomplished at pre-discharge from hospital after recovery from implant. At hospital discharge and after at each follow-up, AV, PV and VV intervals were programmed as designated by the ECHO assessment.
Table 1

Baseline features of study population (n = 53). LVEF: Left Ventricular Ejection Fraction; ESV: End Systolic Volume; EDV: End Diastolic Volume.

Baseline featuresMean ± SD or %
Gender (M)46 (86.8%)
Age (years)68 ± 10
LVEF (%)28 ± 7
ESV (ml)148 ± 70
EDV (ml)198 ± 79
Ischemic cardiomyopathy47.2%
Dilated cardiomyopathy39.6%
Hypertensive cardiomyopathy3.8%
Myocardial non-compaction1.9%
Etiology not reported7.5%
Baseline features of study population (n = 53). LVEF: Left Ventricular Ejection Fraction; ESV: End Systolic Volume; EDV: End Diastolic Volume.

Aims

Primary end-point was to evaluate the relationship between the series of aortic flow Velocity Time Integral (aVTI) values calculated by the two methods at the PV, AV, and VV interval settings recommended by both the QuickOpt™ and the standard ECHO optimization in CRT-D patients, at three index times: pre-discharge, at 6-month and at 12-month follow-ups. Secondary end-point was to define the correlation between the optimal AV, PV and VV intervals defined by ECHO, using aVTI measurements and by the QuickOpt™ algorithm, IEGM-based.

Right ventricular and left ventricular lead positioning

Defibrillation lead was implanted in the mid interventricular septum in all patients. CS lead was implanted using current conventional technique.

Echocardiographic measurements

ECHO and Doppler measurements were performed at hospital discharge, at 6-month and 12-month follow-ups. Transducer and sample volume were maintained in the same position as far as possible during the entire series of AV, PV or VV programming [20]. Acquisition of aortic Doppler flow velocities was performed digitally. The optimal AV, PV, and VV delay by the ECHO method was defined as the delay associated with the largest average aVTI. Aortic VTI measurements were obtained in accordance with the American Society of Echocardiography guidelines [12]. A random sample (average 2 ECHO for each Laboratory at discharge) of echocardiographic exams pre-discharge, 6 months and one year follow-up was evaluated off-line for each center by a Core-lab. The reproducibility in aVTI measurements was determined as the inter- and intra-center (that is: between echocardiographers) using variability coefficient (CV).

aVTI and IEGM acquisitions

AV and VV interval optimization was carried out by each participating center by calculating the aVTI on pulsed Doppler of transaortic flow. Using the VTI method the best matching of AV and VV intervals was determined at each patient’s evaluation. Transducer and sample volume were maintained in the same position as far as possible during the entire series of AV/PV or VV programming [12], [20]. Acquisition of aortic Doppler flow velocities was performed digitally. With patients in a stable clinical condition, three aVTI measurements for each combination of AV, PV and VV (see Table 2) programmed values were performed 2 min after the programming of delays, excluding both premature beats and post-premature contraction beats. From the sampled aortic flow, pre-ejections components were excluded, which corresponded to atrial and isovolumetric ventricular contractions. The sequence of acquisitions of aVTI with different AV and VV intervals was single-blind and randomly decided by the attending electrophysiologist, in order to keep the echocardiographer unaware of the programmed delay.
Table 2

aVTI measures were carried out with different AV, PV and VV delays. The table shows the delays used during Echocardiographic measurements.

aVTIa measures with DDD atrial rate >10bpm spontaneous rate:
QuickOpt AV/PVbQuickOpt VV
QuickOpt AV/PV ± 20QuickOpt VV ± 10
QuickOpt AV/PV ± 40QuickOpt VV ± 20
QuickOpt AV/PV ± 60 (max AV/PV = 200 ms)QuickOpt VV ± 40

aVTI: aortic Velocity Time Integral.

AV/PV_Atrio-Ventricular Delay.

aVTI measures were carried out with different AV, PV and VV delays. The table shows the delays used during Echocardiographic measurements. aVTI: aortic Velocity Time Integral. AV/PV_Atrio-Ventricular Delay.

Statistical analysis

Categorical variables describing the patient population are expressed as absolute numbers and percentages, while continuous variables are shown as means [with standard deviations (SD)] or medians (with quartiles) for continuous variables. Nonparametric Wilcoxon-Mann-Whitney and Wilcoxon signed rank (for paired data) tests were used for non-normally distributed variables. The primary endpoint was assessed by means the linear correlation analysis by the Pearson product-moment correlation coefficient to assess the agreement between the ECHO-based and the IEGM-based aVTIs for each of the AV/PV, and VV delay determinations. The secondary endpoint was assessed by means the Bland–Altman [21] plot method, to test whether the overlap between the series of values given by the two techniques was congruent. It was deemed as clinically acceptable to have a difference not wider than 30 msec in measurements of AV and VV intervals. Intra and inter-observer reproducibility of echocardiographic measurements were reported as CV calculated using the formula: CV = ¼ (SD/arithmetic mean of measurements)/100, where SD is the standard deviation of residuals (measurement 1 - measurement 2) [21]. CV value less than 5% was considered to indicate a good reproducibility. The software Microsoft Office Excel 2007 was employed for data storage and analysis.

Results

Electrical results

Fifty-three patients were enrolled. Rv lead was in septal position in all patients. Atrial lead was implanted in appendage in 98% of the patients. LV lead was implanted in 26 (49%) patients in a lateral vein, in 1(2%) in an antero-lateral vein, in 4 (7%) in a middle vein, in 10 (20%) in a postero vein and in 12 (22%) in a postero-lateral vein. Leads electrical parameters were acceptable after implant procedure and stable at follow-up (Table 3). Neither RV lead dislodgement nor failures in electrical therapies delivery nor arrhythmic death occurred.
Table 3

Left and Right Leads electrical measurements at pre-discharge, 6-month (6 m FU) and 12-month follow-Ups (12 m FU). No failure in sensing, pacing and defibrillation therapy was found during the Study.

Pre-discharge (mean ± SD)6 m FU (mean ± SD)12 m FU (mean ± SD)
RVa wave amplitude (mV)10.88 ± 4.7610.05 ± 5.039.91 ± 5.28
RV Pacing threshold (V × 0.5 ms)0.56 ± 0.510.69 ± 0.420.68 ± 0.46
LVb Pacing threshold (V × 0.5 ms)1.16 ± 0.61.5 ± 11.5 ± 1.18

RV: Right ventricle.

LV:Left Ventricle.

Left and Right Leads electrical measurements at pre-discharge, 6-month (6 m FU) and 12-month follow-Ups (12 m FU). No failure in sensing, pacing and defibrillation therapy was found during the Study. RV: Right ventricle. LV:Left Ventricle.

Comparison between aVTI after optimizing PV and AV intervals by echocardiography and by QuickOpt™

Optimal mean PV and AV values, obtained from the best aVTI which were calculated by both IEGM-based and echocardiographic method, showed a good correlation at hospital discharge and at 6-month and 12-month follow-up. Correlation coefficients were always above 94% (Table 4). The close linear relationship between the values of maximum aVTI calculated with both methods at discharge and after 12 months is showed in Fig. 1.
Table 4

aVTIs obtained with optimal AV/PV intervals calculated with Echo-based and IEGM-based QuickOPt method at pre-discharge, 6-monthfollow-Up (6 m FU) and 12-month follow-Up (12 m FU); Pearson correlation analysis showed a good correlation between aVTIs obtained with both methods; Median and quartiles (25, 75) of AV/PV Optimal Delay; there aren’t significant differences between the two methods.

aVTI (cm) Mean ± SDPearson coefficient valueOptimal delays median; Interquartile 25; 75 (ms)P value
AV delay pre-dischargeMax Echo20.8 ± 6.70.96170[130; 190]P = 0.9
IEGM19.2 ± 9.7150 [150; 170]
PV delay pre-dischargeMax Echo23 ± 80.94120[85; 150]P = 0.4
IEGM21.2 ± 7.3110 [100; 130]
AV delay 6 m FUMax Echo22 ± 60.95160[142.5; 190]P = 0.67
IEGM20 ± 6160[150; 170]
PV delay 6 m FUMax Echo25 ± 80.94120[100; 140]P = 0.59
IEGM23 ± 7110[100; 120]
AV delay 12 m FUMax Echo22.9 ± 8.80.98165[150; 190]P = 0.38
IEGM21.7 ± 8160[150; 170]
PV delay 12 m FUMax Echo26 ± 100.97110[80; 140]P = 0.69
IEGM24 ± 9110[100; 120]
Fig. 1

Correlation between aVTI (cm) calculated with optimal AV/PV intervals optimized by Echo and QuickOpt™ method’s at pre-discharge and at 12 month Follow up (upper line). The correlation between the two methods is always good.

Correlation between aVTI (cm) calculated with optimal AV/PV intervals optimized by Echo and QuickOpt™ method’s at pre-discharge and at 12 month Follow up (upper line). The correlation between the two methods is always good. aVTIs obtained with optimal AV/PV intervals calculated with Echo-based and IEGM-based QuickOPt method at pre-discharge, 6-monthfollow-Up (6 m FU) and 12-month follow-Up (12 m FU); Pearson correlation analysis showed a good correlation between aVTIs obtained with both methods; Median and quartiles (25, 75) of AV/PV Optimal Delay; there aren’t significant differences between the two methods.

Comparison between aVTI after optimizing interventricular (VV) intervals by echocardiography and by QuickOpt™

At the optimized VV intervals, the values of aVTI calculated by both IEGM and echocardiographic method showed a good correlation, with a coefficient above 88% at the three index times (Table 5). The relationship between the maximum aVTI calculated with both methods at discharge and at 12-month follow-up is showed in Fig. 2.
Table 5

aVTIs obtained with optimal VV intervals calculated with Echo-based and IEGM-based QuickOpt™ method at pre-discharge, 6-month follow-Up (6 m FU) and 12.month follow-Up (12 m FU). Pearson correlation analysis showed a good correlation between aVTIs obtained with both methods.

aVTI (cm) Mean ± SDPearson
VV delay pre-dischargeMax Echo23.2 ± 80.92
IEGM20.8 ± 7.1
VV Delay 6 m FUMax Echo24.9 ± 7.50.88
IEGM21 ± 6.7
VV Delay 12 m FUMax Echo25.9 ± 100.91
IEGM22 ± 8
Fig. 2

Correlation between aVTI (cm) calculated at the optimized VV intervals obtained by Echo and QuickOpt™ method’s at pre-discharge and at 12-month follow-up (upper line). The correlation between the two methods was always good.

Correlation between aVTI (cm) calculated at the optimized VV intervals obtained by Echo and QuickOpt™ method’s at pre-discharge and at 12-month follow-up (upper line). The correlation between the two methods was always good. aVTIs obtained with optimal VV intervals calculated with Echo-based and IEGM-based QuickOpt™ method at pre-discharge, 6-month follow-Up (6 m FU) and 12.month follow-Up (12 m FU). Pearson correlation analysis showed a good correlation between aVTIs obtained with both methods. As expected the overall difference between the best aVTI and the worst aVTI is significant for AV (p < 0.001), PV (p < 0.001) and VV (p < 0.001). Different Delay produced significant different in VTI Values.

Secondary end point

Comparison between optimal PV/AV intervals measured by echocardiographic method and by QuickOpt™

Using Bland-Altman plot analysis, in any single patient the clinically acceptable correlation between echo- and IEGM-generated PV/AV optimized measures showed a progressive improvement. Cases where differences in mean’s variations were less than 30 ms were 52% at discharge and became 61% and 65% at six- and twelve-month follow-up (Fig. 3).
Fig. 3

Bland-Altman plot of differences in optimal atrioventricular interval measured by echocardiography and by QuickOpt at Phd, 6 and 12-month follow-up. Grey zone indicate average, upper and lower limit of agreement (1.13 msec ± 31 msec; IC 95% −61 msec + 63 msec). Colored area indicates clinically acceptable margins of difference between the two tecniques (0 ± 30 msec). Although there was no statistical difference between the two methods, the clinically acceptable margins of difference between them (0 ± 30 msec) showed a low concordance (64%).

Bland-Altman plot of differences in optimal atrioventricular interval measured by echocardiography and by QuickOpt at Phd, 6 and 12-month follow-up. Grey zone indicate average, upper and lower limit of agreement (1.13 msec ± 31 msec; IC 95% −61 msec + 63 msec). Colored area indicates clinically acceptable margins of difference between the two tecniques (0 ± 30 msec). Although there was no statistical difference between the two methods, the clinically acceptable margins of difference between them (0 ± 30 msec) showed a low concordance (64%).

Comparison between optimal VV intervals measured by echocardiographic method and by QuickOpt™

The analysis of values given by the two techniques, performed using the Bland–Altman plot, demonstrated that the clinically acceptable congruence between them was substantially stable during the follow-up. The corresponding difference less than 30 ms between the two techniques were 56% at discharge, and 51% and 48% at 6- and 12-month follow-up respectively (Fig. 4).
Fig. 4

Bland-Altman plot of differences in optimal VV interval measured by echocardiography and by QuickOpt at Phd, 6 and 12-month follow-up. Grey zone indicate average, upper and lower limit of agreement (25 msec ± 33 msec; IC 95% −40 msec + 90 msec). Colored area indicates clinically acceptable margins of difference between the two tecniques (0 ± 30 msec). Although there was no statistical difference between the two methods, the clinically acceptable margins of difference between them (0 ± 30 msec) showed a low concordance (56%).

Bland-Altman plot of differences in optimal VV interval measured by echocardiography and by QuickOpt at Phd, 6 and 12-month follow-up. Grey zone indicate average, upper and lower limit of agreement (25 msec ± 33 msec; IC 95% −40 msec + 90 msec). Colored area indicates clinically acceptable margins of difference between the two tecniques (0 ± 30 msec). Although there was no statistical difference between the two methods, the clinically acceptable margins of difference between them (0 ± 30 msec) showed a low concordance (56%).

Intra and inter-observer reproducibility of echocardiographic measurements

Intra and inter-observer reproducibility of aVTI measurements, reported as coefficient of variation (CV), are displayed in Table 6. The aVTI determined at LVOT measurement resulted reproducible, being both intra- and inter-observer CV less than 5% in all follow-ups.
Table 6

Variability coefficient (mean), its standard deviation (+SD) and its range, between both inter- and intra-center echocardiographic measurements of aVTI. All values are less than 5%, demonstrating a good reproducibility between echocardiographic examinations.

CV%SDRange
Pre-Discharge4.7+1.052.9–6.6
6-Month follow-up4.8+1.152.7–7.4
12-Month follow-up4.3+0.63.2–5.5

CV: Variability coefficient.

Variability coefficient (mean), its standard deviation (+SD) and its range, between both inter- and intra-center echocardiographic measurements of aVTI. All values are less than 5%, demonstrating a good reproducibility between echocardiographic examinations. CV: Variability coefficient.

Discussion

In a selected population of CRT-D patients with a mid-septal placement of RV lead, the Quicksept study showed that aVTI values at the optimized AV/PV and VV intervals as determined by ECHO method and by QuickOpt™ algorithm -based method were quite well correlated, and that this correlation was maintained in long-term follow-ups. Previously published works have validated the IEGM-based algorithm for optimizing AV and VV delays, in an acute and single-time setting and with RV lead at the apex. Most prospective, small sample and non-randomized studies have demonstrated that the QuickOpt™ -generated acute optimization was comparable with the ECHO-based one as far as long-term response to CRT was concerned, and that the QuickOpt™ method resulted faster and easier to perform than ECHO-guided optimization [13], [14], [15], [16], [17]. On the other hand other studies was reached a discordant conclusion, The FREEDOM [23] results presented at Heart Rhythm Society in 2010 showed no statistically significant difference between echo and QuickOpt approaches. In Quicksept, ECHO aVTI was used for ECHO-based optimization of AV and VV intervals. In the Pearson correlation analysis, comparing aVTI values, which were obtained by measuring AV and VV intervals by echocardiography and IEGM-based QuickOpt™ algorithm, the level of correspondence resulted around 90% at all three index times (above 90% at AV evaluation, above 88% at VV evaluation), with a fairly linear correlation. However, a further analysis by means of Bland-Altman plot on ECHO-derived and QuickOpt™ –generated data regarding AV intervals found that, at most, only in 52% of cases at discharge, 61% at 6 months follow up and 65% at 12 months follow-up were the percent difference between measurements supplied by the two methods below 30 msec – the predefined upper limit of clinical insignificance. Moreover, when VV intervals were analyzed, this figure decreased to around half the cases in all the follow-ups time. This finding suggests that optimal aVTIs determined by the two methods were similar even with a remarkable difference in AV and VV intervals, and that, consequently, there could be a wide variations among the programmed settings in the same patients. In measurements of VV intervals by the two methods there was a poor correlation depending on the ventricle which was paced first with 50% of discordance (Fig. 5).
Fig. 5

Correlation of optimal VV intervals measured with ECHO and QuickOpt method. In 50% of cases the ventricle paced first was different. In this analysis 13 patients with VV intervals equal to 0 msec as calculated by either method were excluded. Blue colored area represents the conflict zones between the two methods. In general the larger the advance of the left ventricle, the wider the error between the two methods.

Correlation of optimal VV intervals measured with ECHO and QuickOpt method. In 50% of cases the ventricle paced first was different. In this analysis 13 patients with VV intervals equal to 0 msec as calculated by either method were excluded. Blue colored area represents the conflict zones between the two methods. In general the larger the advance of the left ventricle, the wider the error between the two methods. Discrepancy between echo- and QuickOpt™ -calculated aVTIs being fairly well correlated and corresponding AV and VV intervals being scarcely correlated could be due to the fact that even wide variations of AV and VV delays can elicit small hemodynamic effects When we study AV delays, the hemodynamic effect can also be influenced by the VV delay. The inter-ventricular and intra-ventricular delays are influenced by different factors, too. This might be caused in our opinion by the very short time allowed for the CRT device to work with each predefined setting of AV and VV intervals (in our study 2 min, but in other study only 30 s). A sort of slower hemodynamic adaptation to mechanical consequences of a certain programming could be invoked. It can be also hypothesized that methods for measuring aVTI require a longer period of pacing to produce measurable different hemodynamic effects. On the other hand, 2-min interval allowed before each aVTI measurement represented a reasonable standard utilized in other studies, and longer “adaptive” times would have been barely manageable, due to the remarkable duration of each echocardiographic examination. In conclusion, previous reports on QuickOpt™ performance in optimizing AV and VV intervals in acute settings and with apical placement of RV lead were substantially confirmed by our observations in Quicksept study, which demonstrated the same reliability of QuickOpt™ in mid- and long-term follow-up when RV lead had been placed exclusively in the mid-septum [14], [15], [16], [17], [18], [22].

Study limitations

This study was a small prospective Registry with a limited population: findings need confirming on a wider scale and with a randomized design. The criteria for response to CRT were mostly clinical, but response was not used to further investigate into optimization methods and no comparison between responders and non-responders as regards optimization was attempted. AorticVTI suffers from well-known intrinsic limitations for optimizing AV/VV intervals [24]: other authors have proposed other indices, among which the maximal VTI of mitral inflow has gained most attention [25]. On the other hand, there was no clue that aVTI use hampered either of the methods taken into consideration unequally, and its limits can be supposed not to have altered substantially the results of methods’ comparison.

Author’s disclosure

None.
  23 in total

1.  Cardiac resynchronization in chronic heart failure.

Authors:  William T Abraham; Westby G Fisher; Andrew L Smith; David B Delurgio; Angel R Leon; Evan Loh; Dusan Z Kocovic; Milton Packer; Alfredo L Clavell; David L Hayes; Myrvin Ellestad; Robin J Trupp; Jackie Underwood; Faith Pickering; Cindy Truex; Peggy McAtee; John Messenger
Journal:  N Engl J Med       Date:  2002-06-13       Impact factor: 91.245

2.  2010 focused update of ESC Guidelines on device therapy in heart failure: an update of the 2008 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure and the 2007 ESC Guidelines for cardiac and resynchronization therapy. Developed with the special contribution of the Heart Failure Association and the European Heart Rhythm Association.

Authors:  Kenneth Dickstein; Panos E Vardas; Angelo Auricchio; Jean-Claude Daubert; Cecilia Linde; John McMurray; Piotr Ponikowski; Silvia Giuliana Priori; Richard Sutton; Dirk J van Veldhuisen
Journal:  Eur J Heart Fail       Date:  2010-11       Impact factor: 15.534

Review 3.  Echocardiography-guided biventricular pacemaker optimization: role of echo Doppler in hemodynamic assessment and improvement.

Authors:  Reza Rafie; Tasneem Z Naqvi
Journal:  Expert Rev Cardiovasc Ther       Date:  2012-07

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Echocardiographic examination of atrioventricular and interventricular delay optimization in cardiac resynchronization therapy.

Authors:  Maria Cristina Porciani; Cristina Dondina; Roberto Macioce; Gabriele Demarchi; Paolo Pieragnoli; Nicola Musilli; Andrea Colella; Giuseppe Ricciardi; Antonio Michelucci; Luigi Padeletti
Journal:  Am J Cardiol       Date:  2005-05-01       Impact factor: 2.778

6.  A prospective comparison of echocardiography and device algorithms for atrioventricular and interventricular interval optimization in cardiac resynchronization therapy.

Authors:  Ravindu Kamdar; Evelyn Frain; Fiona Warburton; Laura Richmond; Victoria Mullan; Thomas Berriman; Glyn Thomas; Joanna Tenkorang; Mehul Dhinoja; Mark Earley; Simon Sporton; Richard Schilling
Journal:  Europace       Date:  2010-01       Impact factor: 5.214

Review 7.  Effect of atrioventricular and ventriculoventricular delay optimization on clinical and echocardiographic outcomes of patients treated with cardiac resynchronization therapy: a meta-analysis.

Authors:  Dominique Auger; Ulas Hoke; Jeroen J Bax; Eric Boersma; Victoria Delgado
Journal:  Am Heart J       Date:  2013-04-30       Impact factor: 4.749

8.  Long-term clinical effects of programmer-guided atrioventricular and interventricular delay optimization: Intracardiac electrography versus echocardiography for cardiac resynchronization therapy in patients with heart failure.

Authors:  Dongmei Wang; Haibo Yu; Tian Yun; Hongyun Zang; Guitang Yang; Shouli Wang; Zulu Wang; Quanmin Jing; Yaling Han
Journal:  J Int Med Res       Date:  2013-01-23       Impact factor: 1.671

9.  A critical comparison of echocardiographic measurements used for optimizing cardiac resynchronization therapy: stroke distance is best.

Authors:  Dewi E Thomas; Zaheer R Yousef; Alan G Fraser
Journal:  Eur J Heart Fail       Date:  2009-06-23       Impact factor: 15.534

10.  Optimization of AV and VV delays in the real-world CRT patient population: an international survey on current clinical practice.

Authors:  Daniel Gras; Manish S Gupta; Eric Boulogne; Lisa Guzzo; William T Abraham
Journal:  Pacing Clin Electrophysiol       Date:  2009-03       Impact factor: 1.976

View more
  1 in total

Review 1.  The Role of Echocardiography in the Optimization of Cardiac Resynchronization Therapy: Current Evidence and Future Perspectives.

Authors:  Michael Spartalis; Eleni Tzatzaki; Eleftherios Spartalis; Christos Damaskos; Antonios Athanasiou; Efthimios Livanis; Vassilis Voudris
Journal:  Open Cardiovasc Med J       Date:  2017-12-19
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.