| Literature DB >> 27676162 |
Massimo Giammaria1, Gianluca Quirino2, Enrico Cecchi3, Gaetano Senatore4, Paolo Pistelli5, Mario Bocchiardo6, Roberto Mureddu7, Paolo Diotallevi8, Eraldo Occhetta9, Andrea Magnani9, Mauro Bensoni10, Catia Checchinato11, Valentina Conti12, Sandra Badolati13, Antonio Mazza14.
Abstract
BACKGROUND: AtrioVentricular (AV) and InterVentricular (VV) delay optimization can improve ventricular function in Cardiac Resynchronization Therapy (CRT) and is usually performed by means of echocardiography. St Jude Medical has developed an automated algorhythm which calculates the optimal AV and VV delays (QuickOpt™) based on Intracardiac ElectroGrams, (IEGM), within 2 min. So far, the efficacy of the algorhythm has been tested acutely with standard lead position at right ventricular (RV) apex. Aim of this project is to evaluate the algorhythm performance in the mid- and long-term with RV lead located in mid-septum.Entities:
Keywords: Cardiac resinchronization therapy; Echocardiographyc optimization; IEGM based algorythm; Mid-septum stimulation; Optimization algorythm
Year: 2016 PMID: 27676162 PMCID: PMC5832617 DOI: 10.1016/j.ipej.2016.05.001
Source DB: PubMed Journal: Indian Pacing Electrophysiol J ISSN: 0972-6292
Baseline features of study population (n = 53). LVEF: Left Ventricular Ejection Fraction; ESV: End Systolic Volume; EDV: End Diastolic Volume.
| Baseline features | Mean ± SD or % |
|---|---|
| Gender (M) | 46 (86.8%) |
| Age (years) | 68 ± 10 |
| LVEF (%) | 28 ± 7 |
| ESV (ml) | 148 ± 70 |
| EDV (ml) | 198 ± 79 |
| Ischemic cardiomyopathy | 47.2% |
| Dilated cardiomyopathy | 39.6% |
| Hypertensive cardiomyopathy | 3.8% |
| Myocardial non-compaction | 1.9% |
| Etiology not reported | 7.5% |
aVTI measures were carried out with different AV, PV and VV delays. The table shows the delays used during Echocardiographic measurements.
| aVTI | |
| QuickOpt AV/PV | QuickOpt VV |
| QuickOpt AV/PV ± 20 | QuickOpt VV ± 10 |
| QuickOpt AV/PV ± 40 | QuickOpt VV ± 20 |
| QuickOpt AV/PV ± 60 (max AV/PV = 200 ms) | QuickOpt VV ± 40 |
aVTI: aortic Velocity Time Integral.
AV/PV_Atrio-Ventricular Delay.
Left and Right Leads electrical measurements at pre-discharge, 6-month (6 m FU) and 12-month follow-Ups (12 m FU). No failure in sensing, pacing and defibrillation therapy was found during the Study.
| Pre-discharge (mean ± SD) | 6 m FU (mean ± SD) | 12 m FU (mean ± SD) | |
|---|---|---|---|
| RV | 10.88 ± 4.76 | 10.05 ± 5.03 | 9.91 ± 5.28 |
| RV Pacing threshold (V × 0.5 ms) | 0.56 ± 0.51 | 0.69 ± 0.42 | 0.68 ± 0.46 |
| LV | 1.16 ± 0.6 | 1.5 ± 1 | 1.5 ± 1.18 |
RV: Right ventricle.
LV:Left Ventricle.
aVTIs obtained with optimal AV/PV intervals calculated with Echo-based and IEGM-based QuickOPt method at pre-discharge, 6-monthfollow-Up (6 m FU) and 12-month follow-Up (12 m FU); Pearson correlation analysis showed a good correlation between aVTIs obtained with both methods; Median and quartiles (25, 75) of AV/PV Optimal Delay; there aren’t significant differences between the two methods.
| aVTI (cm) Mean ± SD | Pearson coefficient value | Optimal delays median; Interquartile 25; 75 (ms) | P value | ||
|---|---|---|---|---|---|
| AV delay pre-discharge | Max Echo | 20.8 ± 6.7 | 0.96 | 170[130; 190] | P = 0.9 |
| IEGM | 19.2 ± 9.7 | 150 [150; 170] | |||
| PV delay pre-discharge | Max Echo | 23 ± 8 | 0.94 | 120[85; 150] | P = 0.4 |
| IEGM | 21.2 ± 7.3 | 110 [100; 130] | |||
| AV delay 6 m FU | Max Echo | 22 ± 6 | 0.95 | 160[142.5; 190] | P = 0.67 |
| IEGM | 20 ± 6 | 160[150; 170] | |||
| PV delay 6 m FU | Max Echo | 25 ± 8 | 0.94 | 120[100; 140] | P = 0.59 |
| IEGM | 23 ± 7 | 110[100; 120] | |||
| AV delay 12 m FU | Max Echo | 22.9 ± 8.8 | 0.98 | 165[150; 190] | P = 0.38 |
| IEGM | 21.7 ± 8 | 160[150; 170] | |||
| PV delay 12 m FU | Max Echo | 26 ± 10 | 0.97 | 110[80; 140] | P = 0.69 |
| IEGM | 24 ± 9 | 110[100; 120] | |||
Fig. 1Correlation between aVTI (cm) calculated with optimal AV/PV intervals optimized by Echo and QuickOpt™ method’s at pre-discharge and at 12 month Follow up (upper line). The correlation between the two methods is always good.
aVTIs obtained with optimal VV intervals calculated with Echo-based and IEGM-based QuickOpt™ method at pre-discharge, 6-month follow-Up (6 m FU) and 12.month follow-Up (12 m FU). Pearson correlation analysis showed a good correlation between aVTIs obtained with both methods.
| aVTI (cm) Mean ± SD | Pearson | ||
|---|---|---|---|
| VV delay pre-discharge | Max Echo | 23.2 ± 8 | 0.92 |
| IEGM | 20.8 ± 7.1 | ||
| VV Delay 6 m FU | Max Echo | 24.9 ± 7.5 | 0.88 |
| IEGM | 21 ± 6.7 | ||
| VV Delay 12 m FU | Max Echo | 25.9 ± 10 | 0.91 |
| IEGM | 22 ± 8 | ||
Fig. 2Correlation between aVTI (cm) calculated at the optimized VV intervals obtained by Echo and QuickOpt™ method’s at pre-discharge and at 12-month follow-up (upper line). The correlation between the two methods was always good.
Fig. 3Bland-Altman plot of differences in optimal atrioventricular interval measured by echocardiography and by QuickOpt at Phd, 6 and 12-month follow-up. Grey zone indicate average, upper and lower limit of agreement (1.13 msec ± 31 msec; IC 95% −61 msec + 63 msec). Colored area indicates clinically acceptable margins of difference between the two tecniques (0 ± 30 msec). Although there was no statistical difference between the two methods, the clinically acceptable margins of difference between them (0 ± 30 msec) showed a low concordance (64%).
Fig. 4Bland-Altman plot of differences in optimal VV interval measured by echocardiography and by QuickOpt at Phd, 6 and 12-month follow-up. Grey zone indicate average, upper and lower limit of agreement (25 msec ± 33 msec; IC 95% −40 msec + 90 msec). Colored area indicates clinically acceptable margins of difference between the two tecniques (0 ± 30 msec). Although there was no statistical difference between the two methods, the clinically acceptable margins of difference between them (0 ± 30 msec) showed a low concordance (56%).
Variability coefficient (mean), its standard deviation (+SD) and its range, between both inter- and intra-center echocardiographic measurements of aVTI. All values are less than 5%, demonstrating a good reproducibility between echocardiographic examinations.
| CV% | SD | Range | |
|---|---|---|---|
| Pre-Discharge | 4.7 | +1.05 | 2.9–6.6 |
| 6-Month follow-up | 4.8 | +1.15 | 2.7–7.4 |
| 12-Month follow-up | 4.3 | +0.6 | 3.2–5.5 |
CV: Variability coefficient.
Fig. 5Correlation of optimal VV intervals measured with ECHO and QuickOpt method. In 50% of cases the ventricle paced first was different. In this analysis 13 patients with VV intervals equal to 0 msec as calculated by either method were excluded. Blue colored area represents the conflict zones between the two methods. In general the larger the advance of the left ventricle, the wider the error between the two methods.