Literature DB >> 17338767

Acute evaluation of programmer-guided AV/PV and VV delay optimization comparing an IEGM method and echocardiogram for cardiac resynchronization therapy in heart failure patients and dual-chamber ICD implants.

James H Baker1, John McKenzie, Scott Beau, G Stephen Greer, James Porterfield, Matthew Fedor, Steven Greenberg, Emile G Daoud, Raffaele Corbisiero, J Russell Bailey, Linda Porterfield.   

Abstract

INTRODUCTION: Intracardiac delay optimization of biventricular and dual-chamber pacing devices currently relies on time-consuming echocardiographic measurements. A novel intracardiac electrogram (IEGM) method for atrioventricular (AV/PV) and interventricular (VV) delay optimization was developed, which can be performed during routine device follow-up. METHODS AND
RESULTS: In this prospective, nonrandomized, multi-center trial, patients previously implanted with St. Jude Medical cardiac resynchronization therapy defibrillator (CRT-D) devices or dual-chamber implantable cardioverter defibrillators (ICDs) underwent standard AV/PV and/or VV delay optimization guided by Doppler echocardiogram measurements of the maximum aortic velocity time integral (aortic VTI). Aortic VTI measurements applying the IEGM method recommended delays were then obtained in all patients. Fifty-eight patients (age: 68 +/- 11 years; 81% male; 74% ischemic) and 57 patients (age: 71 +/- 10 years; 74% male; 71% ischemic) were enrolled for AV/PV and VV delay evaluation, respectively. An independent core lab determined the maximum aortic VTIs. Data analysis of the AV, PV, and VV delays demonstrated the concordance correlation coefficient (CCC) between the standard method aortic VTI values and the IEGM method aortic VTI values was 97.5%, 96.1%, and 96.6%, respectively. All analyses demonstrated that the CCC > 90% (P < 0.05).
CONCLUSION: The automated programmer-based IEGM method provides a reliable and simpler alternative to standard techniques for the optimization of AV/PV and VV delay settings in patients with CRT-D devices and dual-chamber ICDs.

Entities:  

Mesh:

Year:  2007        PMID: 17338767     DOI: 10.1111/j.1540-8167.2006.00671.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  19 in total

1.  AV interval optimization using pressure volume loops in dual chamber pacemaker patients with maintained systolic left ventricular function.

Authors:  Frank Eberhardt; Thorsten Hanke; Joern Fitschen; Matthias Heringlake; Frank Bode; Heribert Schunkert; Uwe K H Wiegand
Journal:  Clin Res Cardiol       Date:  2012-03-09       Impact factor: 5.460

2.  Increasing knowledge and changing views in cardiac resynchronization therapy.

Authors:  Laszlo Buga; John G F Cleland
Journal:  Heart Fail Rev       Date:  2012-11       Impact factor: 4.214

3.  Comparison of echocardiography and device based algorithm for atrio-ventricular delay optimization in heart block patients.

Authors:  Rajesh Vijayvergiya; Ankur Gupta
Journal:  World J Cardiol       Date:  2015-11-26

4.  [Optimized CRT programming: relevance and practical application].

Authors:  F Bode; F Schütte
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2009-09

5.  Echocardiography versus intracardiac electrocardiography-based optimization for cardiac resynchronization therapy : a comparative clinical long-term trial.

Authors:  C J Jensen; A Liadski; M Bell; C K Naber; O Bruder; G V Sabin; B Küpper; H Wieneke
Journal:  Herz       Date:  2011-10       Impact factor: 1.443

Review 6.  How to improve outcomes: should we put more emphasis on programming and medical care and less on patient selection?

Authors:  Laszlo Buga
Journal:  Heart Fail Rev       Date:  2012-11       Impact factor: 4.214

7.  The reliability of cardiogenic impedance and correlation with echocardiographic and plethysmographic parameters for predicting CRT time intervals post implantation.

Authors:  Elena Sciaraffia; Matthew R Ginks; John Gustafsson; Andreas Karlsson; C Aldo Rinaldi; Carina Blomström Lundqvist
Journal:  J Interv Card Electrophysiol       Date:  2013-04-27       Impact factor: 1.900

Review 8.  Doppler echocardiographic methods for optimization of the atrioventricular delay during cardiac resynchronization therapy.

Authors:  Alan D Waggoner; Lisa de las Fuentes; Victor G Davila-Roman
Journal:  Echocardiography       Date:  2008-10       Impact factor: 1.724

9.  Application of CRT-D in a Marfan syndrome patient with chronic heart failure accompanied by ventricular tachycardia and ventricular fibrillation.

Authors:  Hai-ying Xu; Yi-zhou Xu; Feng Ling; Zhong Yu; Jun Yang; Xu Duan; Bei Wang; Jin-yu Huang
Journal:  J Zhejiang Univ Sci B       Date:  2013-08       Impact factor: 3.066

10.  Acute changes in electromechanical parameters during different pacing configurations using a quadripolar left ventricular lead.

Authors:  Cinzia Valzania; Maria J Eriksson; Mauro Biffi; Giuseppe Boriani; Fredrik Gadler
Journal:  J Interv Card Electrophysiol       Date:  2013-07-03       Impact factor: 1.900

View more