Abstract
PURPOSE: To evaluate the combination of a pressure-indicating sensor film with hydrogel-forming microneedle arrays, as a method of feedback to confirm MN insertion in vivo.Entities:
Keywords: application; feedback; microneedles; pressure-indicating sensor film; transdermal
Mesh:
Substances:
Year: 2016 PMID: 27633885 PMCID: PMC5093199 DOI: 10.1007/s11095-016-2032-z
Source DB: PubMed Journal: Pharm Res ISSN: 0724-8741 Impact factor: 4.200
Fig. 1Images showing the application of thumb pressure by the subject to the MN patch on their ventral forearm (a) and the resulting colour change of the PISF (b).
Fig. 2MN penetration depth in excised neonatal porcine skin applied using the Texture Analyser at varying forces and manual thumb pressure (Means ± S.D., n = 5).
Mean Penetration Depths (n = 5) and Representative Images of the Receiver Layer of the PISF and the Interactive 3D Surface Plot Generated, Following MN Insertion into Excised Neonatal Porcine Skin Using the Texture Analyser at 20 N, 30 N, 50 N and Manual Thumb Application
MN Penetration Depth when Self-Applied in Human Subjects, with and Without the Incorporation of a PISF (Means ± S.D., n = 10)
| MN penetration depth (μm) | |
|---|---|
| With PISF | 245.2 ± 65.8 |
| Without PISF | 228.1 ± 53.7 |
Mean Penetration Depths and Representative Images of the Receiver Layer of the PISF and the Interactive 3D Surface Plot Generated for Three Volunteers, Following Self-Application of MN Arrays into the Ventral Forearm
Human Subjects’ Rating of MNs and the Application Procedure, According to a Likert Scale (n = 10)
| Scenario | Very satisfied | Satisfied | Neither satisfied or dissatisfied | Dissatisfied | Very dissatisfied |
|---|---|---|---|---|---|
| Application procedure | 14 | 6 | 0 | 0 | 0 |
| Applied dose would enter the body | 9 | 8 | 3 | 0 | 0 |
| Works reliably | 9 | 8 | 3 | 0 | 0 |
| Ease of application | 14 | 5 | 0 | 1 | 0 |
| Counselling & patient information leaflet | 18 | 1 | 1 | 0 | 0 |
Fig. 3Subjects’ opinions regarding MN technology in general (a) and preference for MN arrays with or without PISF (b), following MN application.