Accuracy of a portable multisensor body monitor for predicting resting energy expenditure in older people: a comparison with indirect calorimetry.

Stefanie Heiermann1, Kerstin Khalaj Hedayati, Manfred J Müller, Manuela Dittmar.   

Abstract

BACKGROUND: Accurate and comfortable methods are needed to determine resting energy expenditure (REE) in older people who are characterized by a lowered metabolic rate. The portable SenseWear® armband (SWA) body monitor, worn on the right upper arm, can easily be used by this age group in an ambulatory manner.
OBJECTIVE: The purpose of this study was to evaluate the reliability and accuracy of the SWA armband in determining REE in healthy, normal-weight older people.
METHODS: Participants were 49 older, community-dwelling volunteers aged 60-87 years. Reliability was tested in all subjects. Accuracy was analyzed in 32 subjects by comparing REE estimated by SWA against REE measured by indirect calorimetry as a criterion method. Data were simultaneously and continuously recorded for 20 min in the morning. In the same subjects, REE determined by SWA from night recording was evaluated for accuracy against indirect calorimetry. The agreement between methods was assessed by the Bland-Altman procedure.
RESULTS: (1) REE measured by SWA was reliable when comparing 2 consecutive measurements (typical error: 1.9%). (2) REE calculated from morning recording (1,543 ± 181 kcal/24 h) was higher than REE from night recording (1,564 ± 192 kcal/24 h; p = 0.018). (3) Compared with REE by indirect calorimetry (1,377 ± 228 kcal/24 h) for accuracy, the SWA overestimated mean REE in the older people by 12 and 14%, respectively, for morning and night recording (each p < 0.001).
CONCLUSIONS: The SWA provides a reliable estimate of REE in healthy older subjects and has the advantage of easy handling. The 20-min recording time, which was recommended by the manufacturer, can be applied. However, the SWA overestimates REE in this group, possibly due to age-related changes in skin conductance and thermoregulation, both being measured by sensors of the SWA armband. This requires improving the SWA by developing better fitting algorithms for predicting REE in older people.
Copyright © 2011 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2010        PMID: 21196692     DOI: 10.1159/000322109

Source DB:  PubMed          Journal:  Gerontology        ISSN: 0304-324X            Impact factor:   5.140


  4 in total

1.  Evaluation of a non-invasive multisensor accelerometer for calculating energy expenditure in ventilated intensive care patients compared to indirect calorimetry and predictive equations.

Authors:  Janine Krüger; Matthias Kraft; Matthias Gründling; Sigrun Friesecke; Simone Gärtner; Lena J Vogt; Nicole Schüler; Antje Steveling; Markus M Lerch; Ali A Aghdassi
Journal:  J Clin Monit Comput       Date:  2016-09-14       Impact factor: 2.502

Review 2.  Validity of activity monitors in health and chronic disease: a systematic review.

Authors:  Hans Van Remoortel; Santiago Giavedoni; Yogini Raste; Chris Burtin; Zafeiris Louvaris; Elena Gimeno-Santos; Daniel Langer; Alastair Glendenning; Nicholas S Hopkinson; Ioannis Vogiatzis; Barry T Peterson; Frederick Wilson; Bridget Mann; Roberto Rabinovich; Milo A Puhan; Thierry Troosters
Journal:  Int J Behav Nutr Phys Act       Date:  2012-07-09       Impact factor: 6.457

3.  Energy expenditure evaluation in humans and non-human primates by SenseWear Armband. Validation of energy expenditure evaluation by SenseWear Armband by direct comparison with indirect calorimetry.

Authors:  Francesca Casiraghi; Raweewan Lertwattanarak; Livio Luzi; Alberto O Chavez; Alberto M Davalli; Terry Naegelin; Anthony G Comuzzie; Patricia Frost; Nicolas Musi; Franco Folli
Journal:  PLoS One       Date:  2013-09-19       Impact factor: 3.240

4.  Low validity of the Sensewear Pro3 activity monitor compared to indirect calorimetry during simulated free living in patients with osteoarthritis of the hip.

Authors:  Andreas Hermann; Mathias Ried-Larsen; Andreas Kryger Jensen; René Holst; Lars Bo Andersen; Søren Overgaard; Anders Holsgaard-Larsen
Journal:  BMC Musculoskelet Disord       Date:  2014-02-19       Impact factor: 2.362

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.