Reducing the dosimetric impact of positional errors in field junctions for craniospinal irradiation using VMAT.

Andrej Strojnik1, Ignasi Méndez1, Primož Peterlin1.   

Abstract

AIM: To improve treatment plan robustness with respect to small shifts in patient position during the VMAT treatment by ensuring a linear ramp-like dose profile in treatment field overlap regions.
BACKGROUND: Craniospinal irradiation (CSI) is considered technically challenging because the target size exceeds the maximal field size, which necessitates using abutted or overlapping treatment fields. Volumetric modulated arc therapy (VMAT) is increasingly being examined for CSI, as it offers both better dose homogeneity and better dose conformance while also offering a possibility to create field junctions which are more robust towards small shifts in patient position during the treatment.
MATERIALS AND METHODS: A VMAT treatment plan with three isocenters was made for a test case patient. Three groups of overlapping arc field pairs were used; one for the cranial and two for the spinal part. In order to assure a ramp-like dose profile in the field overlap region, the upper spinal part was optimised first, with dose prescription explicitly enforcing a ramp-like dose profile. The cranial and lower spinal part were done afterwards, taking into account the dose contribution of the upper spinal fields.
RESULTS: Using simple geometrical reasoning, we demonstrated that hot- and cold spots which arise from small displacement of one treatment field relative to the other treatment field can be reduced by taking two precautions: (a) widening the field overlap region, and (b) reducing the field gradient across the overlap region. The function with the smallest maximal gradient is a linear ramp. We present a treatment planning technique which yields the desired dose profile of the two contributing fields, and minimises dosimetric dependence on minor positional errors in patient set-up.

Entities:  

Keywords:  Craniospinal irradiation; Set-up error; Treatment field junction; VMAT

Year:  2016        PMID: 27601956      PMCID: PMC5002014          DOI: 10.1016/j.rpor.2016.03.002

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  25 in total

1.  A simple method of radiation treatment of craniospinal fields with patient supine.

Authors:  R B Hawkins
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-01-01       Impact factor: 7.038

2.  A three-isocenter jagged-junction IMRT approach for craniospinal irradiation without beam edge matching for field junctions.

Authors:  Fred Cao; Ramani Ramaseshan; Robert Corns; Sheryl Harrop; Nimet Nuraney; Peter Steiner; Stephanie Aldridge; Mitchell Liu; Hannah Carolan; Alex Agranovich; Anand Karvat
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-03-19       Impact factor: 7.038

3.  Intensity-modulated radiotherapy for craniospinal irradiation: target volume considerations, dose constraints, and competing risks.

Authors:  William Parker; Edith Filion; David Roberge; Carolyn R Freeman
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-09-01       Impact factor: 7.038

4.  On multichannel film dosimetry with channel-independent perturbations.

Authors:  I Méndez; P Peterlin; R Hudej; A Strojnik; B Casar
Journal:  Med Phys       Date:  2014-01       Impact factor: 4.071

5.  Task Group 142 report: quality assurance of medical accelerators.

Authors:  Eric E Klein; Joseph Hanley; John Bayouth; Fang-Fang Yin; William Simon; Sean Dresser; Christopher Serago; Francisco Aguirre; Lijun Ma; Bijan Arjomandy; Chihray Liu; Carlos Sandin; Todd Holmes
Journal:  Med Phys       Date:  2009-09       Impact factor: 4.071

6.  Plan robustness in field junction region from arcs with different patient orientation in total marrow irradiation with VMAT.

Authors:  Pietro Mancosu; Piera Navarria; Luca Castagna; Giacomo Reggiori; Antonella Stravato; Anna Gaudino; Barbara Sarina; Stefano Tomatis; Marta Scorsetti
Journal:  Phys Med       Date:  2015-06-09       Impact factor: 2.685

7.  Intensity-modulated radiation therapy and volumetric-modulated arc therapy for adult craniospinal irradiation--a comparison with traditional techniques.

Authors:  Matthew T Studenski; Xinglei Shen; Yan Yu; Ying Xiao; Wenyin Shi; Tithi Biswas; Maria Werner-Wasik; Amy S Harrison
Journal:  Med Dosim       Date:  2012-08-09       Impact factor: 1.482

8.  Evaluation of localization errors for craniospinal axis irradiation delivery using volume modulated arc therapy and proposal of a technique to minimize such errors.

Authors:  Pamela Myers; Sotirios Stathakis; Panayiotis Mavroidis; Carlos Esquivel; Niko Papanikolaou
Journal:  Radiother Oncol       Date:  2013-06-20       Impact factor: 6.280

9.  The need for the moving junction in craniospinal irradiation.

Authors:  A E Kiltie; J M Povall; R E Taylor
Journal:  Br J Radiol       Date:  2000-06       Impact factor: 3.039

10.  Advantage of protons compared to conventional X-ray or IMRT in the treatment of a pediatric patient with medulloblastoma.

Authors:  W H St Clair; J A Adams; M Bues; B C Fullerton; Sean La Shell; H M Kooy; J S Loeffler; N J Tarbell
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-03-01       Impact factor: 7.038

View more
  6 in total

1.  Evaluating the positional uncertainty of intrafraction, adjacent fields, and daily setup with the BrainLAB ExacTrac system in patients who are receiving craniospinal irradiation.

Authors:  Xiaojuan Duan; Yibing Zhou; Hongya Dai; Lirong Zhao; Jindong Qian; Dingqiang Yang; Liwei Zhang; Can Luo; Guanghui Li
Journal:  J Appl Clin Med Phys       Date:  2020-06-03       Impact factor: 2.102

2.  Feasibility of hybrid TomoHelical- and TomoDirect-based volumetric gradient matching technique for total body irradiation.

Authors:  Chae-Seon Hong; Min-Joo Kim; Jihun Kim; Kyung Hwan Chang; Kwangwoo Park; Dong Wook Kim; Min Cheol Han; Hong In Yoon; Jin Sung Kim; Ho Lee
Journal:  Radiat Oncol       Date:  2019-12-19       Impact factor: 3.481

Review 3.  Total Body Irradiation in Haematopoietic Stem Cell Transplantation for Paediatric Acute Lymphoblastic Leukaemia: Review of the Literature and Future Directions.

Authors:  Bianca A W Hoeben; Jeffrey Y C Wong; Lotte S Fog; Christoph Losert; Andrea R Filippi; Søren M Bentzen; Adriana Balduzzi; Lena Specht
Journal:  Front Pediatr       Date:  2021-12-03       Impact factor: 3.418

4.  Modified volumetric modulated arc therapy technique with reduced planning and treatment time for craniospinal irradiation utilising two isocentres.

Authors:  Minh Nguyen; Thu Dang; Liz Deegan; Kelsie Henry; Scott Jones; Andrew Pullar; Robyn Cheuk; Adrian Gibbs; Lucy Sim; Catriona Hargrave
Journal:  J Med Radiat Sci       Date:  2022-04-22

5.  A new strategy for craniospinal axis localization and adaptive dosimetric evaluation using cone beam CT.

Authors:  Kather Hussain Mohamathu Rafic; Christopher Sujith; Balakrishnan Rajesh; Ebenezer Suman Babu S; Peace Balasingh Timothy; B Selvamani; Paul B Ravindran
Journal:  Rep Pract Oncol Radiother       Date:  2019-12-10

6.  Impact of setup errors on multi-isocenter volumetric modulated arc therapy for craniospinal irradiation.

Authors:  Yongqiang Zhou; Yao Ai; Ce Han; Xiaomin Zheng; Jinling Yi; Congying Xie; Xiance Jin
Journal:  J Appl Clin Med Phys       Date:  2020-10-18       Impact factor: 2.243

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.