Dalton T Snyder1, R Graham Cooks2.
Abstract
The use of successive resonances for ion ejection is demonstrated here as a method of scanning quadrupole ion traps with improvement in both resolution and sensitivity compared with single frequency resonance ejection. The conventional single frequency resonance ejection waveform is replaced with a dual-frequency waveform. The two included frequencies are spaced very closely and their relative amplitudes are adjusted so that the first frequency that ions encounter excites them to higher amplitudes where space charge effects are less prominent, thereby giving faster and more efficient ejection when the ions come into resonance with the second frequency. The method is applicable at any arbitrary frequency, unlike double and triple resonance methods. However, like double and triple resonance ejection, ejection using successive resonances requires the rf and AC waveforms to be phase-locked in order to retain mass accuracy and mass precision. The improved performance is seen in mass spectra acquired by rf amplitude scans (resonance ejection) as well as by secular frequency scans. Graphical Abstract ᅟ.Entities:
Keywords: Arbitrary frequency; Double resonance ejection; Mass resolution; Quadrupole ion trap; Scan modes; Successive resonance ejection