Linear mass scans in quadrupole ion traps using the inverse Mathieu q scan.

Dalton T Snyder1, Christopher J Pulliam1, R Graham Cooks2.   

Abstract

RATIONALE: Secular frequency scanning is a method of mass selectively scanning ions out of a quadrupole ion trap by linearly ramping the frequency of the resonance ejection signal through ion secular frequencies at constant rf amplitude and frequency. The method is electronically much simpler than resonance ejection but it requires a complex nonlinear calibration procedure to correlate mass-to-charge with time.
METHODS: A method of secular frequency scanning in quadrupole ion traps is described in which mass-to-charge is linear with time. This method, termed an "inverse Mathieu q scan", contrasts with linear frequency sweeping which requires a complex nonlinear mass calibration procedure. In the current method, mass scans are forced to be linear with time by scanning the frequency of the supplementary ac so that there is an inverse relationship between the ejected ion's Mathieu q parameter and time.
RESULTS: In all cases, excellent mass spectral linearity is observed. The rf amplitude is shown to control both the scan range and the scan rate, whereas the ac amplitude and scan rate influence the mass resolution. The scan rate depends linearly on the rf amplitude, a unique feature of this scan. Although changes in either rf or ac amplitude affect the positions of peaks in time, they do not change the mass calibration procedure since this only requires a simple linear fit of m/z vs time. Space charge effects are shown to give rise to significant changes in resolution as well as to mass shifts.
CONCLUSIONS: A method of secular frequency scanning which provides a linear mass scale has been demonstrated. The inverse Mathieu q scan offers a significant increase in mass range and power savings while maintaining access to linearity, paving the way for a mass spectrometer based completely on ac waveforms for ion isolation, ion activation, and ion ejection.
Copyright © 2016 John Wiley & Sons, Ltd. Copyright © 2016 John Wiley & Sons, Ltd.

Year:  2016        PMID: 27496220     DOI: 10.1002/rcm.7710

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  5 in total

1.  Ion trap operational modes for ion/ion reactions yielding high mass-to-charge product ions.

Authors:  Kenneth W Lee; Gregory S Eakins; Mark S Carlsen; Scott A McLuckey
Journal:  Int J Mass Spectrom       Date:  2020-02-21       Impact factor: 1.986

2.  Successive Resonances for Ion Ejection at Arbitrary Frequencies in an Ion Trap.

Authors:  Dalton T Snyder; R Graham Cooks
Journal:  J Am Soc Mass Spectrom       Date:  2016-09-06       Impact factor: 3.109

3.  Implementation of Precursor and Neutral Loss Scans on a Miniature Ion Trap Mass Spectrometer and Performance Comparison to a Benchtop Linear Ion Trap.

Authors:  Dalton T Snyder; Lucas J Szalwinski; Ryan Hilger; R Graham Cooks
Journal:  J Am Soc Mass Spectrom       Date:  2018-03-13       Impact factor: 3.109

4.  Precursor and Neutral Loss Scans in an RF Scanning Linear Quadrupole Ion Trap.

Authors:  Dalton T Snyder; Lucas J Szalwinski; Robert L Schrader; Valentina Pirro; Ryan Hilger; R Graham Cooks
Journal:  J Am Soc Mass Spectrom       Date:  2018-03-09       Impact factor: 3.109

5.  Single Analyzer Precursor Ion Scans in a Linear Quadrupole Ion Trap Using Orthogonal Double Resonance Excitation.

Authors:  Dalton T Snyder; R Graham Cooks
Journal:  J Am Soc Mass Spectrom       Date:  2017-06-02       Impact factor: 3.109

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.