Effect of phase locking AC and RF voltages for high mass analysis in a quadrupole ion trap mass spectrometer.

V M Doroshenko1, R J Cotter.   

Abstract

Experiments on the influence of locking the AC and RF frequency phases on the mass resolution and line position in the mass spectra from a quadrupole ion trap mass spectrometer were carried out at low ejection values of beta z to investigate the possibility of using this effect for high mass analysis. At a typical mass scan rate of 1000 u/s the improvement for the mass resolution in these experiments did not exceed 10-20% at beta z = 0.25 compared with 200-250% at beta z = 0.5. However, without phase locking, the mass resolution at beta z = 0.25 was about two times higher than that at beta z = 0.5. Thus, the absolute values for the mass resolution observed without phase-locked AC and RF frequencies at beta z = 0.25 were about the same as at beta z = 0.5 with phase locking. This observation was explained by a significant negative contribution to the mass resolution of the ion microoscillations along the trajectory at the fundamental RF frequency which is different for the cases of beta z = 0.25 and beta z = 0.5.

Mesh:

Substances:

Year:  1996        PMID: 9004528     DOI: 10.1002/rcm.1290101502

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419