Literature DB >> 17297377

A study of the accuracy of cyberknife spinal radiosurgery using skeletal structure tracking.

Anthony K Ho1, Dongshan Fu, Cristian Cotrutz, Steven L Hancock, Steven D Chang, Iris C Gibbs, Calvin R Maurer, John R Adler.   

Abstract

OBJECTIVE: New technology has enabled the increasing use of radiosurgery to ablate spinal lesions. The first generation of the CyberKnife (Accuray, Inc., Sunnyvale, CA) image-guided radiosurgery system required implanted radiopaque markers (fiducials) to localize spinal targets. A recently developed and now commercially available spine tracking technology called Xsight (Accuray, Inc.) tracks skeletal structures and eliminates the need for implanted fiducials. The Xsight system localizes spinal targets by direct reference to the adjacent vertebral elements. This study sought to measure the accuracy of Xsight spine tracking and provide a qualitative assessment of overall system performance.
METHODS: Total system error, which is defined as the distance between the centroids of the planned and delivered dose distributions and represents all possible treatment planning and delivery errors, was measured using a realistic, anthropomorphic head-and-neck phantom. The Xsight tracking system error component of total system error was also computed by retrospectively analyzing image data obtained from eleven patients with a total of 44 implanted fiducials who underwent CyberKnife spinal radiosurgery.
RESULTS: The total system error of the Xsight targeting technology was measured to be 0.61 mm. The tracking system error component was found to be 0.49 mm.
CONCLUSION: The Xsight spine tracking system is practically important because it is accurate and eliminates the use of implanted fiducials. Experience has shown this technology to be robust under a wide range of clinical circumstances.

Entities:  

Mesh:

Year:  2007        PMID: 17297377     DOI: 10.1227/01.NEU.0000249248.55923.EC

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  31 in total

1.  Assessment of image-guided CyberKnife radiosurgery for metastatic spine tumors.

Authors:  Jo-Ting Tsai; Jia-Wei Lin; Wen-Ta Chiu; Woei-Chyn Chu
Journal:  J Neurooncol       Date:  2009-03-03       Impact factor: 4.130

Review 2.  Radiosurgery for spinal malignant tumors.

Authors:  Berndt Wowra; Alexander Muacevic; Stefan Zausinger; Jörg-Christian Tonn
Journal:  Dtsch Arztebl Int       Date:  2009-02-13       Impact factor: 5.594

3.  Surgery for chordomas of the craniocervical junction: lessons learned.

Authors:  David Choi; Michael Gleeson
Journal:  Skull Base       Date:  2010-01

4.  A Systematic Analysis of Errors in Target Localization and Treatment Delivery for Stereotactic Radiosurgery Using 2D/3D Image Registration.

Authors:  Hao Xu; Stephen Brown; Indrin J Chetty; Ning Wen
Journal:  Technol Cancer Res Treat       Date:  2016-08-31

5.  Validation for 2D/3D registration. II: The comparison of intensity- and gradient-based merit functions using a new gold standard data set.

Authors:  Christelle Gendrin; Primoz Markelj; Supriyanto Ardjo Pawiro; Jakob Spoerk; Christoph Bloch; Christoph Weber; Michael Figl; Helmar Bergmann; Wolfgang Birkfellner; Bostjan Likar; Franjo Pernus
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

Review 6.  Spinal radiosurgery: technology and clinical outcomes.

Authors:  M Avanzo; P Romanelli
Journal:  Neurosurg Rev       Date:  2008-09-24       Impact factor: 3.042

7.  Quality of radiosurgery for single brain metastases with respect to treatment technology: a matched-pair analysis.

Authors:  Berndt Wowra; Alexander Muacevic; Jörg-Christian Tonn
Journal:  J Neurooncol       Date:  2009-02-01       Impact factor: 4.130

Review 8.  Extracranial radiosurgery--applications in the management of benign intradural spinal neoplasms.

Authors:  Christine Saraceni; Jonathan B Ashman; James S Harrop
Journal:  Neurosurg Rev       Date:  2009-01-28       Impact factor: 3.042

Review 9.  18F-FDG PET/CT for image-guided and intensity-modulated radiotherapy.

Authors:  Eric C Ford; Joseph Herman; Ellen Yorke; Richard L Wahl
Journal:  J Nucl Med       Date:  2009-09-16       Impact factor: 10.057

10.  Dosimetric impact of angular deviations in positioning for spinal radiosurgery.

Authors:  Jinkoo Kim; Jian-Yue Jin; Ning Wen; Samir H Patel; Benjamin Movsas; Jack Rock; Indrin J Chetty; Samuel Ryu
Journal:  J Radiosurg SBRT       Date:  2012
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.