Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies.

Ovandir Bazan1, Jayme Pinto Ortiz1.   

Abstract

OBJECTIVE: To complement the ISO 5840 standards concerning the duration of left ventricular systole and diastole as a function of changes in heart rates according to in vivo studies from the physiologic literature review.
METHODS: The systolic and diastolic durations from three in vivo studies were compared with the durations of systole proposed by the ISO 5840:2010 and ISO 5840-2:2015 for hydrodynamic performance assessment of prosthetic heart valves.
RESULTS: Based on the in vivo studies analyzed, the systolic durations proposed by the ISO 5840 standard seemed consistent for 45 and 120 beats per minute (bpm), and showed diverse results for the 70 bpm condition.
CONCLUSION: Information on the realistic validation of the operation of left ventricular models for different heart rates were obtained.

Entities:  

Mesh:

Year:  2016        PMID: 27556318      PMCID: PMC5062727          DOI: 10.5935/1678-9741.20160036

Source DB:  PubMed          Journal:  Braz J Cardiovasc Surg        ISSN: 0102-7638


  9 in total

1.  Pulsatile flow studies of a porcine bioprosthetic aortic valve in vitro: PIV measurements and shear-induced blood damage.

Authors:  W L Lim; Y T Chew; T C Chew; H T Low
Journal:  J Biomech       Date:  2001-11       Impact factor: 2.712

2.  In vitro evaluation of aortic valve prosthesis in a novel valved conduit with pseudosinuses of Valsalva.

Authors:  Ruggero De Paulis; Christoph Schmitz; Raffaele Scaffa; Paolo Nardi; Luigi Chiariello; Helmut Reul
Journal:  J Thorac Cardiovasc Surg       Date:  2005-10       Impact factor: 5.209

3.  Effects of mechanical valve orifice direction on the flow pattern in a ventricular assist device.

Authors:  Eiki Akagawa; Hwansung Lee; Eisuke Tatsumi; Akihiko Homma; Tomonori Tsukiya; Nobumasa Katagiri; Yukihide Kakuta; Tomohiro Nishinaka; Toshihide Mizuno; Kei Ota; Rei Kansaku; Yoshiyuki Taenaka
Journal:  J Artif Organs       Date:  2007-06-20       Impact factor: 1.731

4.  Linear relationship between electrical systole, mechanical systole, and heart rate.

Authors:  H Boudoulas; P Geleris; R P Lewis; S E Rittgers
Journal:  Chest       Date:  1981-11       Impact factor: 9.410

5.  Heart-rate-adapted image reconstruction in multidetector-row cardiac CT: influence of physiological and technical prerequisite on image quality.

Authors:  Christopher Herzog; Nasreddin Abolmaali; Joern O Balzer; Sabine Baunach; Hanns Ackermann; Selami Dogan; Martina B Britten; Thomas J Vogl
Journal:  Eur Radiol       Date:  2002-08-21       Impact factor: 5.315

6.  Duration of diastole and its phases as a function of heart rate during supine bicycle exercise.

Authors:  Charles S Chung; Mustafa Karamanoglu; Sándor J Kovács
Journal:  Am J Physiol Heart Circ Physiol       Date:  2004-06-24       Impact factor: 4.733

7.  Coronary artery motion and cardiac phases: dependency on heart rate -- implications for CT image reconstruction.

Authors:  Lars Husmann; Sebastian Leschka; Lotus Desbiolles; Tiziano Schepis; Oliver Gaemperli; Burkhardt Seifert; Philippe Cattin; Thomas Frauenfelder; Thomas G Flohr; Borut Marincek; Philipp A Kaufmann; Hatem Alkadhi
Journal:  Radiology       Date:  2007-09-11       Impact factor: 11.105

8.  Energy loss due to paravalvular leak with transcatheter aortic valve implantation.

Authors:  Ali N Azadani; Nicolas Jaussaud; Peter B Matthews; Liang Ge; T Sloane Guy; Timothy A M Chuter; Elaine E Tseng
Journal:  Ann Thorac Surg       Date:  2009-12       Impact factor: 4.330

9.  Diastolic time - frequency relation in the stress echo lab: filling timing and flow at different heart rates.

Authors:  Tonino Bombardini; Vincenzo Gemignani; Elisabetta Bianchini; Lucia Venneri; Christina Petersen; Emilio Pasanisi; Lorenza Pratali; David Alonso-Rodriguez; Mascia Pianelli; Francesco Faita; Massimo Giannoni; Giorgio Arpesella; Eugenio Picano
Journal:  Cardiovasc Ultrasound       Date:  2008-04-21       Impact factor: 2.062

  9 in total
  2 in total

1.  Mathematical evaluation of cardiac beat synchronization control used for a rotary blood pump.

Authors:  Daisuke Ogawa; Shinji Kobayashi; Kenji Yamazaki; Tadashi Motomura; Takashi Nishimura; Junichi Shimamura; Tomonori Tsukiya; Toshihide Mizuno; Yoshiaki Takewa; Eisuke Tatsumi
Journal:  J Artif Organs       Date:  2019-07-20       Impact factor: 1.731

2.  Initial Simulated FFR Investigation Using Flow Measurements in Patient-specific 3D Printed Coronary Phantoms.

Authors:  Lauren Shepard; Kelsey Sommer; Richard Izzo; Alexander Podgorsak; Michael Wilson; Zaid Said; Frank J Rybicki; Dimitrios Mitsouras; Stephen Rudin; Erin Angel; Ciprian N Ionita
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2017-03-13
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.