Fracture of VerSys fully bead-coated long femoral stems: report on four fractures in 41 hips.

Hsueh-Che Lu1, Cheng-Li Lin, Chih-Wei Chang, Kuo-An Lai.   

Abstract

The fracture of uncemented, fully porous, coated femoral stems is a rare complication that develops after primary total hip arthroplasty, and it is even rarer after revision surgery. To the best of our knowledge, cases of only 10 broken fully coated stems have been reported. This is the first report on fracture of the VerSys fully bead-coated femoral stem, which is designed for diaphyseal fixation and for application in revision surgery or when a proximal fixation is not feasible. Between 2000 and 2008, we implanted 41 VerSys fully coated stems in 40 patients. We present four cases of femoral stem fractures at 23, 31, 40, and 86 months after surgery. The common risk factors for these four patients were inadequate support for the implant in the proximal femur due to nonunion in this area, a stem diameter ≤12 mm, and a relatively young age.
Copyright © 2012. Published by Elsevier B.V.

Entities:  

Mesh:

Year:  2012        PMID: 22632892     DOI: 10.1016/j.kjms.2011.11.018

Source DB:  PubMed          Journal:  Kaohsiung J Med Sci        ISSN: 1607-551X            Impact factor:   2.744


  6 in total

1.  What Is the Risk of Revision Surgery in Hydroxyapatite-coated Femoral Hip Stems? Findings From a Large National Registry.

Authors:  Maria C Inacio; Michelle Lorimer; David C Davidson; Richard N De Steiger; Peter L Lewis; Stephen E Graves
Journal:  Clin Orthop Relat Res       Date:  2018-12       Impact factor: 4.176

2.  Fracture of Fully-coated Femoral Stem after Primary Total Hip Arthroplasty for Nonunion of Intertrochanteric Fracture: A Case Report.

Authors:  Young Soo Chun; Hyung Suk Juh; Yoon Je Cho; Kee Hyung Rhyu
Journal:  Hip Pelvis       Date:  2015-09-30

3.  Fracture of cobalt chrome, fully-coat beaded femoral revision long stem, a clinical outcomes study.

Authors:  Chih-Hsiang Chang; Po-Chun Lin; Cheng-Min Shih; Chun-Chieh Chen; Pang-Hsin Hsieh; Hsin-Nung Shih
Journal:  Biomed J       Date:  2018-03-30       Impact factor: 4.910

4.  Fracture of Uncemented Revision Femoral Stems in three Arthroplasty Patients: A Case Series with three different brands.

Authors:  Rajiv Merchant; Ian Kelly; John Quinlan
Journal:  J Orthop Case Rep       Date:  2014 Oct-Dec

5.  Rupture of Extensively Porous Coated Stems. - A Case Series of 2 Patients.

Authors:  X Gallart; G Claret; S Garcia; J A Fernández-Valencia; J Riba
Journal:  J Orthop Case Rep       Date:  2014 Jul-Sep

6.  Bilateral Fractures of Anatomic Medullary Locking Hip Arthroplasty Stems in a Single Patient: A Case Report.

Authors:  Sean E Slaven; Colin J Harrington; John P Cody
Journal:  J Orthop Case Rep       Date:  2020
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.