Literature DB >> 27456083

Fusion Imaging to Support Endovascular Aneurysm Repair Using 3D-3D Registration.

Christof J Schulz1, Matthias Schmitt1, Dittmar Böckler1, Philipp Geisbüsch1.   

Abstract

PURPOSE: To evaluate the feasibility and accuracy of fusion imaging (FI) during endovascular aneurysm repair (EVAR).
METHODS: FI was performed in 101 consecutive EVAR patients (median age 72 years; 93 men) using automatic registration of the preoperative computed tomography angiography (CTA) with an intraoperative noncontrast cone beam CT (nCBCT; 3D-3D registration). Operative landmarks defined on the CTA were then overlaid in 3 dimensions on fluoroscopy images. Accuracy was measured as the deviation of the position of the lowest renal artery between the FI and angiography. Factors potentially influencing accuracy (α angle, β angle, anesthesia, tortuosity index, neck calcification, neck length, CTA slice thickness, and conventional or sac sealing stent-graft) were analyzed in a multivariate linear regression model.
RESULTS: Median procedure time for nCBCT was 3 minutes (range 2-20), with 4 minutes (range 0.4-15) for registration. An automatic registration tool was used successfully in 90 (89%) patients. Median craniocaudal deviation of the FI was 3 mm (range 0-15). Full accuracy (<1-mm deviation) was seen in 23 (23%) patients, 1- to 3-mm deviation in 23 (23%), 4- to 5-mm deviation in 22 (22%), and >5-mm deviation in 33 (33%). Caudal deviation potentially resulting in renal coverage was seen in 9 (9%). Lateral plus craniocaudal deviation was a median 5.8 mm (range 0-22). The position of the lowest renal artery compared to the FI was left and cranial in 62 (61%). Aneurysm morphology (β angle, p=0.04), CTA slice thickness (p=0.02), and the use of 2 stiff guidewires in endovascular aneurysm sealing (p=0.01) influenced the overlay accuracy.
CONCLUSION: Fusion imaging can be integrated into a daily workflow adding little to the procedure time. Craniocaudal accuracy (<5 mm) was achieved in 68% of cases, allowing optimal C-arm and angiographic catheter positioning or cannulation of target vessels in most patients. However, the accuracy of FI does not allow a noncontrast EVAR procedure without confirmation of FI overlay by a minimal contrast injection or vessel cannulation.
© The Author(s) 2016.

Entities:  

Keywords:  3D imaging; 3D reconstructions; abdominal aortic aneurysm; computed tomography angiography; cone beam computed tomography; endovascular aneurysm repair; fluoroscopy; fusion imaging; image overlay; navigation

Mesh:

Year:  2016        PMID: 27456083     DOI: 10.1177/1526602816660327

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  6 in total

1.  Simultaneous reconstruction of multiple stiff wires from a single X-ray projection for endovascular aortic repair.

Authors:  Katharina Breininger; Moritz Hanika; Mareike Weule; Markus Kowarschik; Marcus Pfister; Andreas Maier
Journal:  Int J Comput Assist Radiol Surg       Date:  2019-08-22       Impact factor: 2.924

Review 2.  Intraoperative Imaging and Image Fusion for Venous Interventions.

Authors:  Ponraj Chinnadurai; Jean Bismuth
Journal:  Methodist Debakey Cardiovasc J       Date:  2018 Jul-Sep

3.  Performance of a feature-based algorithm for 3D-3D registration of CT angiography to cone-beam CT for endovascular repair of complex abdominal aortic aneurysms.

Authors:  Giasemi Koutouzi; Behrooz Nasihatkton; Monika Danielak-Nowak; Henrik Leonhardt; Mårten Falkenberg; Fredrik Kahl
Journal:  BMC Med Imaging       Date:  2018-11-08       Impact factor: 1.930

Review 4.  Augmented Reality in Vascular and Endovascular Surgery: Scoping Review.

Authors:  Joshua Eves; Dimitri Amiras; Abhilash Sudarsanam; Joseph Shalhoub
Journal:  JMIR Serious Games       Date:  2022-09-23       Impact factor: 3.364

5.  Intraoperative stent segmentation in X-ray fluoroscopy for endovascular aortic repair.

Authors:  Katharina Breininger; Shadi Albarqouni; Tanja Kurzendorfer; Marcus Pfister; Markus Kowarschik; Andreas Maier
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-05-19       Impact factor: 2.924

6.  Use of three-dimensional-three-dimensional digital subtraction angiography spin fusion protocol for complex aortoiliac endovascular interventions.

Authors:  Edvard Skripochnik; Nicole D'Ambrosio; William E Crawford; Shang A Loh
Journal:  J Vasc Surg Cases Innov Tech       Date:  2021-04-20
  6 in total