Interleaf leakage for 5 and 10 mm dynamic multileaf collimation systems incorporating patient motion.

E E Klein1, D A Low.   

Abstract

Use of dynamic multileaf collimation (DMLC) for intensity modulated radiation therapy (IMRT) is accelerating. Delivery systems have the ailment of interleaf leakage (IL). This is compounded by the inefficiency of IMRT delivery, estimated to be a factor of 5 for DMLC. With IL on the order of 4%, it is possible to deliver as much as 20% of the prescribed dose to nonprescribed regions. However, IL is characterized by narrow Gaussian peaks of approximately 0.5-1.0 mm full-width-half-maximum (FWHM). We performed a leakage study for 5 and 10 mm leaf systems, accounting for intratreatment and intertreatment motions. In solid phantoms, film was placed perpendicular to beams. DMLC patterns delivered step-wedged distributions. The same field was duplicated using a collimating jaw in a segmented fashion to obtain baseline data of primary and scatter contributions. Longitudinal shifts up to 4 mm and angulations up to 4 degrees were introduced during beam delivery by running multiple patterns, to arrive at a composite delivery. The intent of these rigid body motion experiments was to replicate patient motion. Clinical IMRT fields using segmented MLC were also tested. Films were scanned and converted to dose. A microionization chamber confirmed film data at discrete points. In all cases shifts diminished IL peak values. In the step-wedge case, the net 18 MV IL peaks diminished from 3.6% to 3.2% for the 10 mm system. The 5 mm system IL values decreased from 4.0% to 3.2% with a 2 mm shift but increased to 4.0% with 4 mm shifts. The clinical field data followed the same pattern with a washing out of peak values, but the overall transmission to shielded regions slightly increased. Therefore nonprescribed regions are influenced by an effective transmission value rather than discrete peak IL values. The 5 mm leaf system does not introduce increased IL and is an appropriate system for IMRT.

Entities:  

Mesh:

Year:  2001        PMID: 11548940     DOI: 10.1118/1.1388216

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Leakage-Penumbra effect in intensity modulated radiation therapy step-and-shoot dose delivery.

Authors:  Grigor N Grigorov; James Cl Chow
Journal:  World J Radiol       Date:  2016-01-28

2.  Dosimetric Effect of Jaw Tracking in Volumetric-Modulated Arc Therapy.

Authors:  Sangutid Thongsawad; Chirasak Khamfongkhruea; Chirapha Tannanonta
Journal:  J Med Phys       Date:  2018 Jan-Mar

3.  Dosimetric characteristics of LinaTech DMLC H multi leaf collimator: Monte Carlo simulation and experimental study.

Authors:  Mikaeil Molazadeh; Ahad Zeinali; Mostafa Robatjazi; Alireza Shirazi; Ghazale Geraily
Journal:  J Appl Clin Med Phys       Date:  2017-03-06       Impact factor: 2.102

4.  Determination of MLC model parameters for Monaco using commercial diode arrays.

Authors:  Paul Kinsella; Laura Shields; Patrick McCavana; Brendan McClean; Brian Langan
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.