| Literature DB >> 27430339 |
Valerie Zaphiratos1, Hubert Chiasson2, Pierre Drolet3, Bruno Benzaquen4, Jacques Lapointe4, Louis-Philippe Fortier3.
Abstract
BACKGROUND: Magnetic fields may potentially interfere with the function of cardiovascular implantable electronic devices. Sterile magnetic drapes used to hold surgical instruments are often placed on the patient's thorax, and they are likely to interfere with the function of these devices.Entities:
Keywords: CIED; Drape; Interference; Magnet; Mat; Pacemaker
Mesh:
Year: 2016 PMID: 27430339 PMCID: PMC4949887 DOI: 10.1186/s12938-016-0205-y
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Fig. 1CVD surgical magnetic drape containing 70 ceramic ferrite magnets
Fig. 2Prototype surgical magnetic drape with a silicone shell similar to the CVD drape, but instead contains 70 bottom-isolated ceramic ferrite magnets
Fig. 3Magnetic field of the prototype drape on the patient facing side demonstrating an overall magnetic remanence less than 10 Gauss as measured at a distance of 0.165″ from the drape
Demographic data
| Demographic | Part 1 | Part 2 |
|---|---|---|
| Patients | N = 27 | N = 20 |
| Gender | M 15:F 12 | M 10:F 10 |
| Age (years) (mean ± SD) | 79 ± 8 | 78 ± 8 |
| Height (m) (mean ± SD) | 1.67 ± 0.07 | 1.66 ± 0.09 |
| Weight (kg) (mean ± SD) | 75 ± 17.6 | 73 ± 16 |
| Body mass index (kg/m2) (mean ± SD) | 27 ± 5.2 | 27 ± 4 |
| Type of pacemaker (Boston Scientific/Medtronic) | 17/10 | 9/11 |
SD standard deviation
Results from Part 1 of the study (n = 27)
| Asynchronous rhythm | ||
|---|---|---|
| CVD drape | Prototype drape | |
| Patients (%) | N = 17 (63 %) | N = 0 |
| Age (years) (mean ± SD) | 81 ± 6 | N/A |
| Gender | M 11/F 6 | N/A |
| Weight (kg) (mean ± SD) | 74 ± 18 | N/A |
| Height (m) (mean ± SD) | 1.68 ± 0.07 | N/A |
| Body mass index (kg/m2) (mean ± SD) | 26 ± 5 | N/A |
| Pacemaker (Boston Scientific/Medtronic) | 7/10 | N/A |
Results from part 2 of the study (n = 20)
| Drapes tested | Prototype | CVD | JCM | SDR | LDR |
|---|---|---|---|---|---|
| Patients who demonstrated asynchronous rhythm (%) | N = 0 | N = 16 (80 %) | N = 5 (25 %) | N = 1 (5 %) | N = 0 |
| Age (years) (mean ± SD) | N/A | 80 ± 7 | 82 ± 7 | 87 | N/A |
| Gender | N/A | M 7/F 9 | M 4/F 1 | M 0/F 1 | N/A |
| Weight (kg) (mean ± SD) | N/A | 69 ± 15 | 70 ± 18 | 44 | N/A |
| Height (m) (mean ± SD) | N/A | 1.66 ± 0.09 | 1.74 ± 0.08 | 1.52 | N/A |
| Body mass index (kg/m2) (mean ± SD) | N/A | 26 ± 4 | 26 ± 3 | 19 | N/A |
| Pacemaker (Boston Scientific/Medtronic) | N/A | 6/10 | 3/2 | 0/1 | N/A |
Detailed results for each patient from part 1 of the study
| Subject | Date | Age (years) | Gender | Weight (kg) | Height (m) | BMI | Pacemaker model | Asynchronous rhythm with CVD drape | Displacementa | Asynchronous rhythm with prototype drape (with and without displacement) |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 20/10/2011 | 82 | F | 68 | 1.68 | 24.02 | Boston Scientific | No | – | No |
| 2 | 20/10/2011 | 73 | M | 82 | 1.78 | 25.75 | Boston Scientific | Yes | Yes | No |
| 3 | 20/10/2011 | 81 | F | 60 | 1.60 | 23.44 | Boston Scientific | No | – | No |
| 4 | 20/10/2011 | 65 | M | 84 | 1.73 | 28.20 | Boston Scientific | No | – | No |
| 5 | 27/10/2011 | 82 | M | 66 | 1.75 | 21.49 | Boston Scientific | No | – | No |
| 6 | 27/10/2011 | 52 | M | 102 | 1.78 | 32.19 | Boston Scientific | No | – | No |
| 7 | 27/10/2011 | 78 | F | 54 | 1.63 | 20.47 | Boston Scientific | Yes | Yes | No |
| 8 | 27/10/2011 | 75 | F | 88 | 1.60 | 34.38 | Boston Scientific | No | – | No |
| 9 | 27/10/2011 | 84 | F | 59 | 1.60 | 23.05 | Boston Scientific | No | – | No |
| 10 | 27/10/2011 | 86 | M | 77 | 1.70 | 26.68 | Boston Scientific | No | – | No |
| 11 | 27/10/2011 | 82 | M | 95 | 1.73 | 31.84 | Boston Scientific | Yes | No | No |
| 12 | 27/10/2011 | 81 | F | 61 | 1.60 | 23.91 | Boston Scientific | Yes | No | No |
| 13 | 27/10/2011 | 70 | F | 55 | 1.52 | 23.81 | Boston Scientific | No | – | No |
| 14 | 27/10/2011 | 76 | M | 85 | 1.74 | 28.17 | Boston Scientific | Yes | Yes | No |
| 15 | 27/10/2011 | 76 | M | 56 | 1.64 | 20.82 | Boston Scientific | Yes | No | No |
| 16 | 27/10/2011 | 80 | F | 77 | 1.65 | 28.17 | Boston Scientific | Yes | Yes | No |
| 17 | 27/10/2011 | 73 | F | 102 | 1.65 | 37.50 | Boston Scientific | No | – | No |
| 18 | 3/11/2011 | 81 | M | 64 | 1.68 | 22.50 | Medtronic | Yes | Yes | No |
| 19 | 3/11/2011 | 92 | M | 57 | 1.70 | 19.62 | Medtronic | Yes | Yes | No |
| 20 | 3/11/2011 | 85 | F | 69 | 1.63 | 25.78 | Medtronic | Yes | Yes | No |
| 21 | 3/11/2011 | 75 | M | 94 | 1.75 | 30.79 | Medtronic | Yes | Yes | No |
| 22 | 3/11/2011 | 85 | M | 66 | 1.65 | 24.17 | Medtronic | Yes | Yes | No |
| 23 | 3/11/2011 | 77 | M | 93 | 1.65 | 34.31 | Medtronic | Yes | Yes | No |
| 24 | 3/11/2011 | 76 | M | 108 | 1.83 | 32.25 | Medtronic | Yes | No | No |
| 25 | 3/11/2011 | 90 | M | 61 | 1.68 | 21.68 | Medtronic | Yes | Yes | No |
| 26 | 3/11/2011 | 87 | F | 45 | 1.55 | 18.90 | Medtronic | Yes | Yes | No |
| 27 | 3/11/2011 | 77 | F | 93 | 1.65 | 34.16 | Medtronic | Yes | No | No |
aDisplacement was necessary in order to obtain asynchronous rhythm
Detailed results for each patient for part 2 of the study
| Subject | Age | Gender | Weight | Height | BMI | Pacemaker model | Asynchronous rhythm with drape | ||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Prototype | JCM | LDR | SDR | CVD | |||||||
| 1 | 74 | M | 89 | 1.78 | 28.1 | Medtronic | No | No | No | No | No |
| 2 | 87 | F | 44 | 1.52 | 19 | Medtronic | No | No | No | Yes | Yes |
| 3 | 66 | M | 89 | 1.73 | 29.7 | Medtronic | No | No | No | No | Yes |
| 4 | 78 | M | 70 | 1.70 | 24.2 | Medtronic | No | No | No | No | Yes |
| 5 | 86 | F | 52 | 1.60 | 20.3 | Medtronic | No | No | No | No | Yes |
| 6 | 87 | F | 66 | 1.55 | 27.5 | Medtronic | No | No | No | No | Yes |
| 7 | 88 | F | 61 | 1.65 | 22.4 | Medtronic | No | No | No | No | Yes |
| 8 | 81 | M | 77 | 1.80 | 23.8 | Medtronic | No | Yes | No | No | Yes |
| 9 | 69 | M | 80 | 1.60 | 31.2 | Boston Scientific | No | No | No | No | No |
| 10 | 75 | M | 60 | 1,49 | 28.6 | Boston Scientific | No | Yes | No | No | Yes |
| 11 | 64 | F | 94 | 1.60 | 36.7 | Boston Scientific | No | No | No | No | No |
| 12 | 92 | F | 46 | 1,49 | 21.9 | Boston Scientific | No | Yes | No | No | Yes |
| 13 | 78 | F | 89 | 1.70 | 30.8 | Boston Scientific | No | No | No | No | Yes |
| 14 | 76 | M | 91 | 1.75 | 29.7 | Boston Scientific | No | No | No | No | No |
| 15 | 83 | M | 95 | 1.78 | 30 | Boston Scientific | No | Yes | No | No | Yes |
| 16 | 79 | F | 79 | 1.68 | 28 | Boston Scientific | No | No | No | No | Yes |
| 17 | 72 | M | 75 | 1.75 | 24.5 | Medtronic | No | No | No | No | Yes |
| 18 | 80 | F | 63 | 1.60 | 24.6 | Boston Scientific | No | No | No | No | Yes |
| 19 | 72 | F | 66 | 1.52 | 28.6 | Medtronic | No | No | No | No | Yes |
| 20 | 77 | M | 70 | 1.65 | 25.7 | Medtronic | No | Yes | No | No | Yes |