Literature DB >> 27428913

Clinical Evaluation of an Unsintered Hydroxyapatite/Poly-L-Lactide Osteoconductive Composite Device for the Internal Fixation of Maxillofacial Fractures.

Shintaro Sukegawa1, Takahiro Kanno, Naoki Katase, Akane Shibata, Yuka Takahashi, Yoshihiko Furuki.   

Abstract

INTRODUCTION: OSTEOTRANS MX (Takiron Co, Ltd, Osaka, Japan) is a resorbable osteosynthetic material composed of an unsintered hydroxyapatite/poly-L-lactide composite, and its osteoconductive capacity has been documented. The authors here report their clinical experience using OSTEOTRANS MX.
METHODS: The authors treated 35 patients (19 men, 16 women; age, 14-88 years; mean ± standard deviation, 38.4 ± 19.9 years) with maxillofacial fractures. The authors used standard surgery to stabilize fractures in all patients, fitting resorbable plates (thickness, 1.0 or 1.4 mm) and screws (diameter, 2 mm) according to Arbeitsgemeinschaft für Osteosynthesefragen/Association (AO) for the Study of Internal Fixation guidelines.
RESULTS: All patients eventually achieved satisfactory healing with favorable restoration of form and function without foreign body reaction. Complications occurred in 3 patients-plate exposure in 2 and discomfort in 1. However, fracture sites healed in all patients. Scanning electron microscopy revealed that the devices bonded directly to the bone without interposition of nonmineralized tissue.
CONCLUSION: OSTEOTRANS MX is a useful material with few complications. Its osteoconductive bioactivity is advantageous for the early functional improvement of maxillofacial fractures.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27428913      PMCID: PMC5023762          DOI: 10.1097/SCS.0000000000002828

Source DB:  PubMed          Journal:  J Craniofac Surg        ISSN: 1049-2275            Impact factor:   1.046


Fixation of maxillofacial fractures with titanium miniplates is an established and widely used treatment. The mechanical properties of titanium, including its strength, ease of handling, lack of dimensional changes,[1,2] minimal scatter on computed tomography (CT) scanning, and compatibility with radiography and magnetic resonance imaging,[3] have prompted its widespread adoption as the general standard. However, fixation with titanium has several disadvantages, such as potential interference with facial growth,[4,5] thermal sensitivity,[3,6] plate migration,[7,8] and interference with diagnostic imaging.[9] The adverse effects of retained metallic devices also include osteopenia of cortical bone induced by protection from stress and corrosion.[10] Thus, resorbable bone fixation devices have been developed. OSTEOTRANS MX (Takiron Co, Ltd, Osaka, Japan), also called Super FIXSORB MX in Japan, is a bioactive, totally resorbable osteosynthetic bone fixation material. It is composed of a composite of fine particles of unsintered hydroxyapatite (u-HA) and poly-l-lactide (PLLA). The properties of this material have some clinical advantages, as follows: its mechanical strength closely approximates that of cortical bone, it has osteoconductive properties, and it makes direct contacts with the bone. Furthermore, because of its bioactive and biodegradable properties, the u-HA/PLLA composite has the potential for total replacement by bone.[11] In this study, we report our clinical experience using OSTEOTRANS MX.

METHODS

Patients

We enrolled 35 patients (19 men, 16 women) aged 14 to 88 (mean ± standard deviation, 38.4 ± 19.9) years with maxillofacial fractures treated between 2009 and 2014 at the Division of Oral and Maxillofacial Surgery, Kagawa Prefectural Central Hospital, Kagawa, Japan. This study was approved by the Ethics Committee of Kagawa Prefectural Central Hospital (Approval No. 396). All patients provided written informed consent.

Methods

We used a standard process to stabilize fractures in all patients, fitting resorbable plates (thickness, 1.0 or 1.4 mm) and screws (diameter, 2 mm) (Fig. 1) according to Arbeitsgemeinschaft für Osteosynthesefragen/Association (AO) for the Study of Internal Fixation principles.
FIGURE 1

OSTEOTRANS MX (Takiron Co, Ltd, Osaka, Japan) composed of a forged unsintered hydroxyapatite/poly-l-lactide composite.

OSTEOTRANS MX (Takiron Co, Ltd, Osaka, Japan) composed of a forged unsintered hydroxyapatite/poly-l-lactide composite. Le Fort I type fractures were treated with exposure of the anterior maxillary walls and piriform aperture. After the fractured segments were reduced, the maxillary segment was repositioned and fixed paranasally and at the infrazygomatic crest with 4-hole plates. Zygomaticomaxillary complex (ZMC) fractures were stabilized with 2-point fixation at the maxillary buttress and frontozygomatic suture following the ZMC reduction. A third fixation point at the infraorbital rim was used if the orbit was explored or reconstructed, and if this was performed, it helped to stabilize the fracture (Fig. 2).
FIGURE 2

A 62-year-old male patient with a Le Fort I fracture from a traffic injury. (A) Preoperative computed tomography scan. (B) Transoral internal fixation.

A 62-year-old male patient with a Le Fort I fracture from a traffic injury. (A) Preoperative computed tomography scan. (B) Transoral internal fixation. Mandible fractures were manually reduced using reduction tools such as reduction forceps, and patients were placed into maxillomandibular fixation with the use of intermaxillary fixation screws with wire and heavy elastics. All fractures were treated with intraoral open reductions. Teeth in the line of fracture were extracted only if they demonstrated mobility, tooth root fracture, apical pathology, were nonrestorable, or interfered with the reduction of the fractures or occlusion. The internal fixation design was the 2-miniplate technique for the parasymphyseal, symphysis, and mandibular body fractures. At a minimum, 2 screws were placed on each side of the fracture (Fig. 3).
FIGURE 3

(A) A 52-year-old male patient with left mandibular body fracture from a fall injury. Preoperative computed tomography scan. (B) Transoral internal fixation.

(A) A 52-year-old male patient with left mandibular body fracture from a fall injury. Preoperative computed tomography scan. (B) Transoral internal fixation. After reducing the fracture, a template of the contours of the bone was created. The optimal shape and size of the plate to be used were selected and shaped according to this template. The material was warmed in hot water and adjusted by bending to match the contours of the template. Finally, the plate was affixed with screws. First, we drilled the bone to a depth of approximately 1 mm less than the total screw length. Second, we tapped the screw hole carefully with the screw tap to open it and rinsed out the bone chips. Third, the device was inserted and carefully fixed with screws to the bone. Regarding postoperative management, although we did not use intermaxillary fixation, guiding elastics were used to guide occlusion in mandibular fractures where appropriate. Patients were followed up postoperatively, and the functional status and complications (including infection, swelling, nonunion, malunion, and exposure of the plate) of treatment were assessed through clinical and radiographical examinations. The unexposed hardware side of the removed plates and screws and tissue adhered to the plate were subjected to evaluation as follows.

Scanning Electron Microscopy

A removed plate was fixed in 10% phosphate-buffered formalin to examine the attached tissues by scanning electron microscopy (SEM; S-3400N; Hitachi, Ltd, Tokyo, Japan).

Energy-Dispersive X-Ray Analysis

Elemental analyses were performed using an energy-dispersive X-ray analyzer (EX-350; Horiba, Ltd, Kyoto, Japan). The operating conditions used for analysis comprised an accelerating voltage of 20 kV and a working distance of 10 mm.

Measurement of Molecular Weight

The viscosity average molecular weight (Mv) of the polymer was determined from the intrinsic viscosity in chloroform at 25°C using the following equation[12]:

Measurement of Crystallinity

The crystallinity (%) of the polymer was determined using the following equation:: where ΔHTm is the melting enthalpy (J/g), and 93.7 is the theoretical melting enthalpy of 100% crystalline PLLA polymer (J/g).[13]

Histological Evaluation

We conducted a histological examination of tissue adhered to the plate that required removal from a patient experiencing discomfort. The tissue sample was fixed in 15% phosphate-buffered formalin, dehydrated with a graded 70% to 100% ethanol series, serially soaked in xylol, embedded in paraffin, and sectioned. The sections were stained with hematoxylin and eosin according to standard procedures.

RESULTS

Clinical Evaluation

All patients eventually achieved satisfactory healing with favorable restoration of form and function without foreign body reaction. The fracture lines healed completely in all patients (Table 1, Fig. 4). Complications occurred in 3 patients (3/35; 8.6%)—plate exposure (n = 2) and discomfort (n = 1). No nonunion or malunion occurred (Table 1). Plate exposure occurred in 2 patients with mandibular fractures (in the parasymphyseal region [n = 1] and in the body [n = 1]; Fig. 5). In both patients, plate exposure was recognized in the alveolar region and the exposed plate was removed after bone healing.
TABLE 1

Clinical Characteristics of 35 Patients With Maxillofacial Fractures

PatientAgeSexType of FractureEtiologyOutcome of Fracture TreatmentPostoperative ComplicationTiming of Plate Removal (mo)Observation Period (mo)
156MZMCTraffic accidentsHealingDiscomfort12
234MMandibular bodyFallHealingExposed plate8
319MMandibular bodyTraffic accidentsHealingExposed plate4
424FSymphysisFallHealingNon68
518MLe Fort ITraffic accidentsHealingNon63
674FZMCTraffic accidentsHealingNon41
742FLe Fort ITraffic accidentsHealingNon30
847FSymphysisFallHealingNon21
924FParasymphysisTraffic accidentsHealingNon20
1023MLe Fort ITraffic accidentsHealingNon19
1130MLe Fort IInterpersonal conflictsHealingNon19
1275MZMCFallHealingNon19
1314FParasymphysisFallHealingNon17
1428FMandibular bodyTraffic accidentsHealingNon16
1548FLe Fort ITraffic accidentsHealingNon16
1619FLe Fort IFallHealingNon15
1778FMandibular bodyFallHealingNon15
1833FLe Fort I/ II FxTraffic accidentsHealingNon14
1929MLe Fort IFallHealingNon13
2030FZMCInterpersonal conflictsHealingNon12
2143MParasymphysisTraffic accidentsHealingNon12
2241MSymphysisTraffic accidentsHealingNon11
2371MMandibular CondyleFallHealingNon11
2415FSymphysisTraffic accidentsHealingNon10
2535MMandibular bodyFallHealingNon10
2626FSymphysisFallHealingNon9
2719MSymphysisSportHealingNon8
2838FParasymphysisFallHealingNon8
2917MMandibular condyleSportHealingNon8
3030FSymphysisFallHealingNon8
3127MSymphysisTraffic accidentsHealingNon7
3262MLe Fort1 FxTraffic accidentsHealingNon7
3388MParasymphysisFallHealingNon6
3452MParasymphysealFallHealingNon6
3534MLe Fort1 FxFallHealingNon6

ZMC, zygomaticomaxillary complex.

FIGURE 4

A 33-year-old female patient with a right Le Fort I and left Le Fort I/II fractures from traffic accident. (A) Frontal aspect at the emergency transportation, showing the remarkable facial deformation. (B) Intraoral view, showing his occlusal deviation due to midfacial fractures. (C) Three-dimensional CT showing complicated and dislocated right Le Fort I and left Le Fort I/ II fractures. (D, E) Bilateral transoral internal fixation with OSTEOTRANS MX. (F) Frontal aspect at 6-month follow-up, showing improvement of facial deformation. (G) Intraoral view, showing improvement of her occlusal. (H) Three-dimensional CT at postoperative 13-month, showing good bone integration. CT, computed tomography.

FIGURE 5

Plate exposure occurred in 2 patients with mandibular fractures. In both patients, exposure was recognized from the alveolar region and was not complete. (A) Plate exposure was recognized in the parasymphyseal region. (B) Plate exposure was recognized in the mandibular body.

A 33-year-old female patient with a right Le Fort I and left Le Fort I/II fractures from traffic accident. (A) Frontal aspect at the emergency transportation, showing the remarkable facial deformation. (B) Intraoral view, showing his occlusal deviation due to midfacial fractures. (C) Three-dimensional CT showing complicated and dislocated right Le Fort I and left Le Fort I/ II fractures. (D, E) Bilateral transoral internal fixation with OSTEOTRANS MX. (F) Frontal aspect at 6-month follow-up, showing improvement of facial deformation. (G) Intraoral view, showing improvement of her occlusal. (H) Three-dimensional CT at postoperative 13-month, showing good bone integration. CT, computed tomography. Plate exposure occurred in 2 patients with mandibular fractures. In both patients, exposure was recognized from the alveolar region and was not complete. (A) Plate exposure was recognized in the parasymphyseal region. (B) Plate exposure was recognized in the mandibular body. One patient with a ZMC fracture reported discomfort (Fig. 6) of the infraorbital rim with no pus 10 months after surgery. The patient was monitored for a few months, but the situation did not improve, and at the patient's request, the plate was removed.
FIGURE 6

Discomfort occurred in 1 patient with a zygomaticomaxillary complex fracture. (A) The patient complained of discomfort of the infraorbital rim was evident 10 months after surgery. (B) The remnants of the plate and screws with fibrous tissue.

Discomfort occurred in 1 patient with a zygomaticomaxillary complex fracture. (A) The patient complained of discomfort of the infraorbital rim was evident 10 months after surgery. (B) The remnants of the plate and screws with fibrous tissue.

Scanning Electron Microscopy, Energy-Dispersive X-Ray Analysis, and Molecular Weight and Crystallinity Measurements

In a plate removed 4 months after surgery, the Mv and crystallinity of the material were 64 kDa and 54.1%, respectively. The Mv and crystallinity of the screw were 60 kDa and 54.4%, respectively (Table 2). On SEM, the outline of the plate remained distinct. Bone formation was seen in a circle around the plate, and remodeled trabecular bone had formed toward the center (Fig. 7A). The plate was observed to bond directly to the bone without interposition of nonmineralized tissue (Fig. 7B). On energy-dispersive X-ray analysis, the weight concentration of the deposits was 24.06 wt% calcium and 13.63 wt% phosphorus, similar to that of bone.[14] Thus, we consider these deposits to be bone tissue.
TABLE 2

Molecular Weight and Crystallinity Measurements of the Removed Plates

PatientAgeSexTiming of Plate RemovalMolecular Weight (kDa)Crystallinity (%)
119M4M
plate6454.1
screw6053.4
234M8M
plate4854.5
screw3964.8
356M12M
plate4859
screw26
FIGURE 7

Scanning electron microscopy imaging of tissue attached to the plates. Bone formation in a circle around the plate and remodeled trabecular bone formation toward the center. The plate was observed to bond directly to the bone, with no interposition of nonmineralized tissue.

Scanning electron microscopy imaging of tissue attached to the plates. Bone formation in a circle around the plate and remodeled trabecular bone formation toward the center. The plate was observed to bond directly to the bone, with no interposition of nonmineralized tissue. In a plate removed 8 months after surgery, the Mv and crystallinity of the plate were 48 kDa and 54.5%, respectively. The Mv and crystallinity of the screw were 39 kDa and 64.8%, respectively. Bone-like deposits were observed on SEM. In a plate removed 1 year after surgery, the Mv and crystallinity of the plate were 48 kDa and 59%, respectively. The Mv of the screw was 26 kDa. The crystallinity of the screw could not be measured because the sample volume was too small. Bone-like deposits were observed on SEM. Fibrous tissue was also evident around the plate (Fig. 8).
FIGURE 8

Histopathological examination confirmed the presence of fibrous tissue. The specimen was composed of dense fibrous tissue without infiltration of inflammatory cells or blood vessels, resembling the organizing phase of granulation tissue (magnification: ×200).

Histopathological examination confirmed the presence of fibrous tissue. The specimen was composed of dense fibrous tissue without infiltration of inflammatory cells or blood vessels, resembling the organizing phase of granulation tissue (magnification: ×200).

DISCUSSION

A resorbable osteosynthesis system has been developed for use similar to titanium mini plates in craniofacial, oral, maxillofacial, plastic, and reconstructive surgeries.[2,15-19] Using this system, there is no need for a second surgery to remove the plate. OSTEOTRANS MX is made from a composite of uncalcined and unsintered u-HA and PLLA. Resorbable plates have been previously made from PLLA alone. However, PLLA osteosynthesis devices have several disadvantages, including lower dynamic strength, an inability to fuse with bone, and long resorption and replacement times.[20-23] OSTEOTRANS MX was designed to overcome these problems, and the u-HA/PLLA composite material was developed by adding particulate resorbable uncalcined and unsintered u-HA to PLLA. Surgeons have recognized that bone fixation devices for use in maxillofacial surgery require much higher initial mechanical strength than natural cortical bone (which has a bending strength of 200 MPa). The u-HA/PLLA composite material has demonstrated higher mechanical strength, including bending strength, bending modulus, shears strength, and impact strength, than PLLA devices.[24] Indeed, the material has an initial bending strength of 270 MPa,[24] far higher than the bending strength of cortical bone. Hence, the high mechanical strength of OSTEOTRANS MX is thus advantageous for bone healing. Taken together, because of its bioactive, osteoconductive, as well as biodegradable properties, this u-HA/PLLA composite fixation system has the higher potentials for clinical advantages and may broaden the further indications in maxillofacial fracture repair surgery such as more complex, displaced, or comminuted fractures as a feasible next generative material. The mechanical strength of the described osteosynthesis system is basically inferior to that of metallic, titanium devices. Recent evidence has revealed biomechanical differences between conventional bioresorbable plates and titanium hardware specifically designed for repair of mandible fractures. Stanton et al[25] reported that the use of resorbable plating systems combined with a short postoperative period of intermaxillary fixation is a very effective method for internal fixation of mandibular fractures. Some studies demonstrated that conventional resorbable plates provide insufficient stability to counteract the masticatory forces of the mandible, resulting in higher rates of delayed union, as compared with a traditional titanium-based system.[26,27] The resorbable plates used in these previous studies were constructed from PLLA alone or composites of PLLA and PLDLA/PGA. On the other hand, with respect to postoperative management, we did not use intermaxillary fixation, but rather elastics, to guide occlusion in mandibular fractures where appropriate. The intensity of this bioresorbable plate system is very high and its bending strength is far greater than that for cortical bone, whereas its modulus is almost equivalent to that of cortical bone.[24] This plate system has high mechanical strength compared with that of other resorbable plates, although it is considered to be subject to appropriate occlusal management of maxillofacial fractures. One of the disadvantages of resorbable plates and screws is the lack of self-tapping. Because of these properties, it is necessary to pay much closer attention when using these devices for repair of complex midfacial injuries, severe displacement, and comminuted fractures. The thin and fragile maxillofacial bone regions, such as maxillary sinus and orbital walls, require careful attention when tapping in a technically demanding procedure.[26-28] However, we occasionally selected a surgical procedure to perform only drilling without tapping, which was not a problem when carefully applied for fixation of extremely thin and fragile bone for repair of complex midfacial fractures. We confirmed fracture healing postoperatively using CT scanning. We found that the fracture site had healed in all patients. It was possible to observe the plates because the u-HA particles render them radiopaque. CT scanning clearly showed the plates at various stages of fracture healing in our patients. We observed that the form of the plates was maintained until bone healing. Shikinami et al[29] reported that HA crystals were deposited and grew on the surface of this material 3 to 6 days after immersion in simulated body fluid, suggesting that it can bond directly to bone. Yasunaga et al[30] reported that an u-HA/PLLA plate fixed to rabbit tibial cortices made direct contact with the bone with no intervening fibrous tissue. These reports indicate that u-HA/PLLA plates can bond directly to bone and possess osteoconductivity. We performed scanning capacitance microscopy analysis on an incompletely exposed plate after its removal 4 months postoperatively. Deposits of bone-like tissue were evident on the bone-contacting surface of the removed plate. The fibrous appearance of these deposits was suggestive of collagen fibrils of bone tissue, as evident in magnified images. This proved that the plates directly bonded to bone and proved the osteoconductivity of the u-HA/PLLA plate system. Its early osteoconductive bioactivity is advantageous for the early functional improvement of maxillofacial fractures. Foreign-body reaction was not observed in all patients. The conventional devices made from the highly crystalline PLLA and PGA have shown long-term foreign body reactions.[31,32] Furthermore, it has been confirmed that PGA degrades too rapidly after implantation and gives rise to considerable side effects that created inflammatory foreign-body reaction around the collapsing devices due to lowering of the pH and irritation from the fibrinous debris.[33] Although the reinforced fabricated PLLA is more hydrophobic than PGA and therefore degrades slowly, PLLA-only devices have the potential to release uneven PLLA fragments at irregular time intervals during the degradation process. This can induce physical inflammatory responses. On the other hand, uniform hydrolysis occurred throughout the PLLA matrix in the u-HA/PLLA material, and a steady release of small amounts of PLLA debris during PLLA degradation did not provoke adverse tissue responses.[33] Additionally, a thin PLLA film provided interface with invading water molecules, which allowed homogeneous hydrolysis and steady degradation of PLLA. Thus, u-HA/PLLA material has the advantage of preventing a foreign-body reaction by its moderate and stable hydrolysis.[34] Complications occurred in 3 patients, comprising exposed plates (n = 2) and discomfort (n = 1). The plate was removed in all these patients. Wound dehiscence and infection following plate exposure are the most frequent reasons for plate removal.[35,36] Plate exposure occurred in 2 patients with mandibular fractures in the parasymphyseal region and the body. In both patients, plate exposure was recognized in the alveolar region and was incomplete. For increased strength, resorbable plates are of increased thickness; a resorbable plate must be thick to match the strength of a titanium plate.[37] Although thicker plates are advantageous for mechanical strength, their increased mass makes them undesirable, as they are at a greater risk of exposure. The gingival alveolar portion is thin and easily suffers wound dehiscence, leading to exposure. Thus, it is necessary to cover it sufficiently using an oral vestibular approach. Discomfort occurred in only 1 patient with a ZMC fracture. The patient complained of discomfort in the infraorbital rim approximately 1 year postoperatively. In this patient, we removed the plate after bone healing at the patient's request. The plate was palpable through the patient's facial skin, and the patient was experiencing a foreign-body sensation. In the immediate postoperative period, there was no discomfort, probably because of the general swelling of the face. However, the patient complained of mild discomfort after the facial swelling had subsided. Using thin plates is an option to reduce the discomfort caused by plates with midfacial fracture treatment, although there is controversy regarding their fixation strength.[38,39] Yoshitaka et al[40] reported that thicker plates were related to more discomfort than were thinner plates in their study. Polymerized PLLA plates that are used to fix unstable zygomatic fractures are voluminous; they are approximately 3 times the size of a titanium plate.[41] In fact, these bioresorbable plates have a thickness of 1.0 or 1.4 mm. All of them are thicker than the thinnest titanium plate of 0.4 mm. Histopathological examination showed that the area of discomfort contained much fibrous tissue. Rarely, when the gross geometry of the implant is rapidly lost at approximately 1-year postsurgery, there is an increased risk of discomfort due to adverse tissue reactions.[42] Covered with fibrous connective tissue around device, since there is no acute inflammation, it was not a foreign-body reaction due to the absorption process. In this patient, the removed plate had median Mv and low crystallinity compared with those of the new plate and screw. It was suggested that the stage of resorption was in progress, but it was not at the stage of phagocytic degradation by cells. Shikinami et al[34] reported that the PLLA matrix in a composite was completely degraded and totally absorbed after 4 to 5 years, by which point natural new bone had replaced the majority of the u-HA particles. Additionally, it was previously reported that a decrease in Mv indicated the degradation of PLLA and the composite,[22] and that the crystallinity of PLLA increased during hydrolysis.[43] During the degradation of the screw and plate, the screw (which is in the bone) tends to decompose earlier than the plate (which is outside the bone). This indicates that degradation occurs more rapidly in the bone than in plates covered with fibrous tissue. Therefore, when using a resorbable plate in an area of thin facial skin, such as the periorbital region or frontozygomatic suture, it is necessary to consider the following: It is easier to palpate a thicker resorbable plate than a titanium plate. It is possible that discomfort may be exacerbated over time because of a further increase in volume due to fibrous tissue covering the plate. This material, composed of a reinforced composite of u-HA and PLLA, is suitable for use as an internal bone fixation device for facial fractures[19] not only because of its mechanical properties but also because of its bioactivities, including osteoconduction. We believe that this material may supersede titanium plates in the treatment of maxillofacial fractures if the appropriate technique is used to optimize the characteristics of this device

CONCLUSION

The present prospective clinical study of maxillofacial fractures showed successful treatment with resorbable implants composed of composite an unsintered hydroxyapatite/PLLA. All patients had successful fracture stabilization and reossification, with low complications. Its osteoconductive bioactivity is advantageous for the early functional improvement of maxillofacial fractures.
  36 in total

1.  Bonding behavior of ultrahigh strength unsintered hydroxyapatite particles/poly(L-lactide) composites to surface of tibial cortex in rabbits.

Authors:  T Yasunaga; Y Matsusue; T Furukawa; Y Shikinami; M Okuno; T Nakamura
Journal:  J Biomed Mater Res       Date:  1999-12-05

2.  Comparison of stability of titanium and absorbable plate and screw fixation for mandibular angle fractures.

Authors:  Alparslan Esen; Hanife Ataoğlu; Lokman Gemi
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2008-08-20

Review 3.  Removal of asymptomatic bone plates used for orthognathic surgery and facial fractures.

Authors:  B Alpert; D Seligson
Journal:  J Oral Maxillofac Surg       Date:  1996-05       Impact factor: 1.895

4.  Foreign-body reactions to fracture fixation implants of biodegradable synthetic polymers.

Authors:  O Böstman; E Hirvensalo; J Mäkinen; P Rokkanen
Journal:  J Bone Joint Surg Br       Date:  1990-07

5.  A 5-year in vitro and in vivo study of the biodegradation of polylactide plates.

Authors:  R Suuronen; T Pohjonen; J Hietanen; C Lindqvist
Journal:  J Oral Maxillofac Surg       Date:  1998-05       Impact factor: 1.895

6.  Association between plate location and plate removal following facial fracture repair.

Authors:  Yoshitaka Kubota; Tomoaki Kuroki; Shinsuke Akita; Tomoe Koizumi; Masakazu Hasegawa; Naoaki Rikihisa; Nobuyuki Mitsukawa; Kaneshige Satoh
Journal:  J Plast Reconstr Aesthet Surg       Date:  2011-10-24       Impact factor: 2.740

7.  Structural pillars of the facial skeleton: an approach to the management of Le Fort fractures.

Authors:  P N Manson; J E Hoopes; C T Su
Journal:  Plast Reconstr Surg       Date:  1980-07       Impact factor: 4.730

8.  Ultra-high-strength absorbable self-reinforced polyglycolide (SR-PGA) composite rods for internal fixation of bone fractures: in vitro and in vivo study.

Authors:  P Törmälä; J Vasenius; S Vainionpää; J Laiho; T Pohjonen; P Rokkanen
Journal:  J Biomed Mater Res       Date:  1991-01

9.  The use of resorbable plates in association with dental arch stabilization in the treatment of mandibular fractures in children.

Authors:  Zhi Li; Ongodia David; Zu-Bing Li
Journal:  J Craniomaxillofac Surg       Date:  2013-09-12       Impact factor: 2.078

10.  Segmental stability in orthognathic surgery: hydroxyapatite/Poly-l-lactide osteoconductive composite versus titanium miniplate osteosyntheses.

Authors:  Constantin A Landes; Alexander Ballon; Andreas Tran; Shahram Ghanaati; Robert Sader
Journal:  J Craniomaxillofac Surg       Date:  2014-01-21       Impact factor: 2.078

View more
  15 in total

1.  Use of the bioactive resorbable plate system for zygoma and zygomatic arch replacement and fixation with modified Crockett's method for maxillectomy: A technical note.

Authors:  Shintaro Sukegawa; Takahiro Kanno; Akane Shibata; Kenichi Matsumoto; Yuka Sukegawa-Takahashi; Kyousuke Sakaida; Yoshihiko Furuki
Journal:  Mol Clin Oncol       Date:  2017-05-23

2.  Resorbable Versus Titanium Rigid Fixation for Pediatric Mandibular Fractures: A Systematic Review, Institutional Experience and Comparative Analysis.

Authors:  Matthew E Pontell; Eva B Niklinska; Stephane A Braun; Nolan Jaeger; Kevin J Kelly; Michael S Golinko
Journal:  Craniomaxillofac Trauma Reconstr       Date:  2021-06-21

3.  Efficacy of bioresorbable plates in the osteosynthesis of linear mandibular fractures.

Authors:  Satyavrat Arya; Krushna Bhatt; Ongkila Bhutia; Ajoy Roychoudhury
Journal:  Natl J Maxillofac Surg       Date:  2020-06-18

4.  Complications of a poly-L-lactic acid and polyglycolic acid osteosynthesis device for internal fixation in maxillofacial surgery.

Authors:  Shintaro Sukegawa; Takahiro Kanno; Kenichi Matsumoto; Yuka Sukegawa-Takahashi; Masanori Masui; Yoshihiko Furuki
Journal:  Odontology       Date:  2018-02-07       Impact factor: 2.634

5.  The Clinical Feasibility of Newly Developed Thin Flat-Type Bioresorbable Osteosynthesis Devices for the Internal Fixation of Zygomatic Fractures: Is There a Difference in Healing Between Bioresorbable Materials and Titanium Osteosynthesis?

Authors:  Shintaro Sukegawa; Takahiro Kanno; Daiki Nagano; Akane Shibata; Yuka Sukegawa-Takahashi; Yoshihiko Furuki
Journal:  J Craniofac Surg       Date:  2016-11       Impact factor: 1.046

6.  Biomechanical Loading Evaluation of Unsintered Hydroxyapatite/poly-l-lactide Plate System in Bilateral Sagittal Split Ramus Osteotomy.

Authors:  Shintaro Sukegawa; Takahiro Kanno; Yoshiki Manabe; Kenichi Matsumoto; Yuka Sukegawa-Takahashi; Masanori Masui; Yoshihiko Furuki
Journal:  Materials (Basel)       Date:  2017-07-07       Impact factor: 3.623

Review 7.  Overview of innovative advances in bioresorbable plate systems for oral and maxillofacial surgery.

Authors:  Takahiro Kanno; Shintaro Sukegawa; Yoshihiko Furuki; Yoshiki Nariai; Joji Sekine
Journal:  Jpn Dent Sci Rev       Date:  2018-04-05

8.  Bioactive Regeneration Potential of the Newly Developed Uncalcined/Unsintered Hydroxyapatite and Poly-l-Lactide-Co-Glycolide Biomaterial in Maxillofacial Reconstructive Surgery: An In Vivo Preliminary Study.

Authors:  Shinji Ishizuka; Quang Ngoc Dong; Huy Xuan Ngo; Yunpeng Bai; Jingjing Sha; Erina Toda; Tatsuo Okui; Takahiro Kanno
Journal:  Materials (Basel)       Date:  2021-05-10       Impact factor: 3.623

9.  Feasibility of a Three-Dimensional Porous Uncalcined and Unsintered Hydroxyapatite/poly-d/l-lactide Composite as a Regenerative Biomaterial in Maxillofacial Surgery.

Authors:  Yunpeng Bai; Takahiro Kanno; Hiroto Tatsumi; Kenichi Miyamoto; Jingjing Sha; Katsumi Hideshima; Yumi Matsuzaki
Journal:  Materials (Basel)       Date:  2018-10-20       Impact factor: 3.623

10.  Advantage of Alveolar Ridge Augmentation with Bioactive/Bioresorbable Screws Made of Composites of Unsintered Hydroxyapatite and Poly-L-lactide.

Authors:  Shintaro Sukegawa; Hotaka Kawai; Keisuke Nakano; Kiyofumi Takabatake; Takahiro Kanno; Hitoshi Nagatsuka; Yoshihiko Furuki
Journal:  Materials (Basel)       Date:  2019-11-08       Impact factor: 3.623

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.