Comparison of fixation stability of resorbable versus titanium plate and screws in mandibular angle fractures.

Burak Bayram1, Kenan Araz, Sina Uckan, Cenk Balcik.   

Abstract

PURPOSE: The aim of this study was to compare the fixation reliability and stability of titanium and resorbable plates and screws by simulating chewing forces.
MATERIALS AND METHODS: Mandibular angle fractures in 11 sheep hemimandibles were fixed with 4-hole straight titanium plates and 2.0 x 7-mm titanium screws; in addition, 11 hemimandibles were fixed with 4-hole straight resorbable plates and 2.5 x 6-mm resorbable screws according to the Champy technique. The hemimandibles were mounted with a fixation device in a servohydraulic testing unit for compressive testing. Displacement values under 20, 60, 100, 120, 150, and 200 N; maximum displacements; and maximum forces that the model could resist before breakage were recorded and compared.
RESULTS: Significant differences were found between resorbable and titanium plates and screws at all forces (20, 60, 100, 120, 150, and 200 N) (P < .05). We found no statistically significant differences in the breaking force and maximum displacement values (displacement values at the breaking forces) between the groups.
CONCLUSIONS: The stability of mandibular angle fractures with titanium miniplates under simulated chewing forces was significantly higher than with the resorbable system. Metallic and resorbable fixation systems cannot be used interchangeably to treat mandibular angle fractures under similar loading conditions.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19615576     DOI: 10.1016/j.joms.2009.03.011

Source DB:  PubMed          Journal:  J Oral Maxillofac Surg        ISSN: 0278-2391            Impact factor:   1.895


  6 in total

1.  Resorbable Versus Titanium Rigid Fixation for Pediatric Mandibular Fractures: A Systematic Review, Institutional Experience and Comparative Analysis.

Authors:  Matthew E Pontell; Eva B Niklinska; Stephane A Braun; Nolan Jaeger; Kevin J Kelly; Michael S Golinko
Journal:  Craniomaxillofac Trauma Reconstr       Date:  2021-06-21

Review 2.  Fixation of mandibular angle fractures: in vitro biomechanical assessments and computer-based studies.

Authors:  Bruno Ramos Chrcanovic
Journal:  Oral Maxillofac Surg       Date:  2012-10-14

3.  Resorbable implants in maxillofacial surgery: a reality check.

Authors:  Suresh Menon; Col Sk Roy Choudhury
Journal:  J Maxillofac Oral Surg       Date:  2011-04-07

4.  Comparison of resorbable plates and titanium plates for fixation stability of combined mandibular symphysis and angle fractures.

Authors:  Ho-Yong Lim; Chang-Hwa Jung; Seong-Yong Kim; Jin-Yong Cho; Jae-Young Ryu; Hyeon-Min Kim
Journal:  J Korean Assoc Oral Maxillofac Surg       Date:  2014-12-26

Review 5.  A Narrative Review of u-HA/PLLA, a Bioactive Resorbable Reconstruction Material: Applications in Oral and Maxillofacial Surgery.

Authors:  Huy Xuan Ngo; Yunpeng Bai; Jingjing Sha; Shinji Ishizuka; Erina Toda; Rie Osako; Akira Kato; Reon Morioka; Mrunalini Ramanathan; Hiroto Tatsumi; Tatsuo Okui; Takahiro Kanno
Journal:  Materials (Basel)       Date:  2021-12-26       Impact factor: 3.623

6.  Clinical Evaluation of an Unsintered Hydroxyapatite/Poly-L-Lactide Osteoconductive Composite Device for the Internal Fixation of Maxillofacial Fractures.

Authors:  Shintaro Sukegawa; Takahiro Kanno; Naoki Katase; Akane Shibata; Yuka Takahashi; Yoshihiko Furuki
Journal:  J Craniofac Surg       Date:  2016-09       Impact factor: 1.046

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.