Evaluation of a tissue-mimicking thermochromic phantom for radiofrequency ablation.

Andrew S Mikhail1, Ayele H Negussie1, Cole Graham1, Manoj Mathew1, Bradford J Wood1, Ari Partanen2.   

Abstract

PURPOSE: This work describes the characterization and evaluation of a tissue-mimicking thermochromic phantom (TMTCP) for direct visualization and quantitative determination of temperatures during radiofrequency ablation (RFA).
METHODS: TMTCP material was prepared using polyacrylamide gel and thermochromic ink that permanently changes color from white to magenta when heated. Color vs temperature calibration was generated in matlab by extracting RGB color values from digital photographs of phantom standards heated in a water bath at 25-75 °C. RGB and temperature values were plotted prior to curve fitting in mathematica using logistic functions of form f(t) = a + b/(1 + e((c(t-d)))), where a, b, c, and d are coefficients and t denotes temperature. To quantify temperatures based on TMTCP color, phantom samples were heated to temperatures blinded to the investigators, and two methods were evaluated: (1) visual comparison of sample color to the calibration series and (2) in silico analysis using the inverse of the logistic functions to convert sample photograph RGB values to absolute temperatures. For evaluation of TMTCP performance with RFA, temperatures in phantom samples and in a bovine liver were measured radially from an RF electrode during heating using fiber-optic temperature probes. Heating and cooling rates as well as the area under the temperature vs time curves were compared. Finally, temperature isotherms were generated computationally based on color change in bisected phantoms following RFA and compared to temperature probe measurements.
RESULTS: TMTCP heating resulted in incremental, permanent color changes between 40 and 64 °C. Visual and computational temperature estimation methods were accurate to within 1.4 and 1.9 °C between 48 and 67 °C, respectively. Temperature estimates were most accurate between 52 and 62 °C, resulting in differences from actual temperatures of 0.6 and 1.6 °C for visual and computational methods, respectively. Temperature measurements during RFA using fiber-optic probes matched closely with maximum temperatures predicted by color changes in the TMTCP. Heating rate and cooling rate, as well as the area under the temperature vs time curve were similar for TMTCP and ex vivo liver.
CONCLUSIONS: The TMTCP formulated for use with RFA can be used to provide quantitative temperature information in mild hyperthermic (40-45 °C), subablative (45-50 °C), and ablative (>50 °C) temperature ranges. Accurate visual or computational estimates of absolute temperatures and ablation zone geometry can be made with high spatial resolution based on TMTCP color. As such, the TMTCP can be used to assess RFA heating characteristics in a controlled, predictable environment.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27370145      PMCID: PMC4912564          DOI: 10.1118/1.4953394

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  24 in total

1.  Exploiting thermochromic materials for the rapid quality assurance of physiotherapy ultrasound treatment heads.

Authors:  Ian Butterworth; Jill Barrie; Bajram Zeqiri; Gordana Žauhar; Benjamin Parisot
Journal:  Ultrasound Med Biol       Date:  2012-03-16       Impact factor: 2.998

2.  Pulsatile flow phantom for ultrasound image-guided HIFU treatment of vascular injuries.

Authors:  Robyn Greaby; Vesna Zderic; Shahram Vaezy
Journal:  Ultrasound Med Biol       Date:  2007-04-27       Impact factor: 2.998

Review 3.  MR thermometry for monitoring tumor ablation.

Authors:  Baudouin Denis de Senneville; Charles Mougenot; Bruno Quesson; Iulius Dragonu; Nicolas Grenier; Chrit T W Moonen
Journal:  Eur Radiol       Date:  2007-05-22       Impact factor: 5.315

Review 4.  Percutaneous tumor ablation with radiofrequency.

Authors:  Bradford J Wood; Jeffrey R Ramkaransingh; Tito Fojo; McClellan M Walther; Stephen K Libutti
Journal:  Cancer       Date:  2002-01-15       Impact factor: 6.860

5.  Tissue-mimicking gel phantoms for thermal therapy studies.

Authors:  Ali Dabbagh; Basri Johan Jeet Abdullah; Chanthiriga Ramasindarum; Noor Hayaty Abu Kasim
Journal:  Ultrason Imaging       Date:  2014-03-13       Impact factor: 1.578

6.  Thermal dosimetry and temperature measurements.

Authors:  D A Christensen
Journal:  Cancer Res       Date:  1979-06       Impact factor: 12.701

7.  Thermal dose determination in cancer therapy.

Authors:  S A Sapareto; W C Dewey
Journal:  Int J Radiat Oncol Biol Phys       Date:  1984-06       Impact factor: 7.038

8.  Current tumor ablation technologies: basic science and device review.

Authors:  David F Saldanha; Vishal L Khiatani; Tami C Carrillo; Felix Y Yap; James T Bui; M Grace Knuttinen; Charles A Owens; Ron C Gaba
Journal:  Semin Intervent Radiol       Date:  2010-09       Impact factor: 1.513

Review 9.  Arrhenius relationships from the molecule and cell to the clinic.

Authors:  W C Dewey
Journal:  Int J Hyperthermia       Date:  1994 Jul-Aug       Impact factor: 3.914

Review 10.  Magnetic Resonance-Guided Drug Delivery.

Authors:  Andrew S Mikhail; Ari Partanen; Pavel Yarmolenko; Aradhana M Venkatesan; Bradford J Wood
Journal:  Magn Reson Imaging Clin N Am       Date:  2015-07-09       Impact factor: 2.266

View more
  7 in total

1.  Gel Phantom Models for Radiofrequency and Microwave Ablation of the Liver.

Authors:  Willa J Chen; Qi Wang; Charles Y Kim
Journal:  Dig Dis Interv       Date:  2020-09-28

2.  RF-ablation pattern shaping employing switching channels of dual bipolar needle electrodes: ex vivo results.

Authors:  Jens Ziegle; Chloé Audigier; Johannes Krug; Ghazanfar Ali; Younsu Kim; Emad M Boctor; Michael Friebe
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-04-20       Impact factor: 2.924

3.  Ultrasound thermal monitoring with an external ultrasound source for customized bipolar RF ablation shapes.

Authors:  Younsu Kim; Chloé Audigier; Jens Ziegle; Michael Friebe; Emad M Boctor
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-04-04       Impact factor: 2.924

Review 4.  Application of Nanoparticles and Nanomaterials in Thermal Ablation Therapy of Cancer.

Authors:  Zhannat Ashikbayeva; Daniele Tosi; Damir Balmassov; Emiliano Schena; Paola Saccomandi; Vassilis Inglezakis
Journal:  Nanomaterials (Basel)       Date:  2019-08-24       Impact factor: 5.076

5.  A feasibility study of microwave therapy for precancerous actinic keratosis.

Authors:  D N Jackson; F J Hogarth; D Sutherland; E M Holmes; P T Donnan; C M Proby
Journal:  Br J Dermatol       Date:  2020-03-23       Impact factor: 9.302

6.  Technical advance in silico and in vitro development of a new bipolar radiofrequency ablation device for renal denervation.

Authors:  Noel Pérez; Karl Muffly; Stephen E Saddow
Journal:  BMC Cardiovasc Disord       Date:  2021-10-16       Impact factor: 2.298

7.  Matrix 3D ultrasound-assisted thyroid nodule volume estimation and radiofrequency ablation: a phantom study.

Authors:  T Boers; S J Braak; M Versluis; S Manohar
Journal:  Eur Radiol Exp       Date:  2021-07-29
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.