Exploiting thermochromic materials for the rapid quality assurance of physiotherapy ultrasound treatment heads.

Ian Butterworth1, Jill Barrie, Bajram Zeqiri, Gordana Žauhar, Benjamin Parisot.   

Abstract

Significant nonuniformities in the acoustic intensity distribution generated by physiotherapy ultrasound treatment heads are not uncommon, potentially leading to significant localised temperature rises and tissue damage. An acoustic absorber tile containing a thermochromic pigment has been developed to provide rapid quality assurance of physiotherapy ultrasound treatment heads by virtue of a thermochromic colour change, indicating the time-averaged intensity distributions generated by these devices. As a bench-top device, the use of the tile is designed to mimic the nature of the physiotherapeutic application, requiring minimal training. Two designs where thermochromic pigments are added to the various polymeric layers of the tile are presented. Testing has been conducted with two physiotherapy treatment heads of differing performance, one of them notably exhibiting a strong "hot-spot" in localised acoustic time-averaged intensity. Findings show good qualitative agreement with classical hydrophone scans. Techniques are explored for the correction of nonlinearities in the thermochromic relationship, to enhance the accuracy of quantitative assessment. Crown
Copyright © 2012. Published by Elsevier Inc. All rights reserved.

Mesh:

Year:  2012        PMID: 22425374     DOI: 10.1016/j.ultrasmedbio.2012.01.021

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  3 in total

1.  Evaluation of a tissue-mimicking thermochromic phantom for radiofrequency ablation.

Authors:  Andrew S Mikhail; Ayele H Negussie; Cole Graham; Manoj Mathew; Bradford J Wood; Ari Partanen
Journal:  Med Phys       Date:  2016-07       Impact factor: 4.071

2.  Thermochromic tissue-mimicking phantom for optimisation of thermal tumour ablation.

Authors:  Ayele H Negussie; Ari Partanen; Andrew S Mikhail; Sheng Xu; Nadine Abi-Jaoudeh; Subha Maruvada; Bradford J Wood
Journal:  Int J Hyperthermia       Date:  2016-04-20       Impact factor: 3.914

3.  A thermochromic tissue-mimicking phantom model for verification of ablation plans in thermal ablation.

Authors:  Yan Zhou; Lin Zhao; Xinyu Zhong; Jianmin Ding; Hongyu Zhou; Fengmei Wang; Xiang Jing
Journal:  Ann Transl Med       Date:  2021-02
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.