| Literature DB >> 27364749 |
Audun Eskeland Rimehaug1,2,3, Ingrid Elise Hoff4,5, Lars Øivind Høiseth5,6, Jonny Hisdal7, Petter Aadahl8,9,10, Idar Kirkeby-Garstad8,9,10.
Abstract
BACKGROUND: Changes in cardiac power parameters incorporate changes in both aortic flow and blood pressure. We hypothesized that dynamic and non-dynamic cardiac power parameters would track hypovolemia better than equivalent flow- and pressure parameters, both during spontaneous breathing and non-invasive positive pressure ventilation (NPPV).Entities:
Keywords: Cardiac power; Critical care; Hypovolemia; Lower body negative pressure; Vascular impedance
Mesh:
Year: 2016 PMID: 27364749 PMCID: PMC4929737 DOI: 10.1186/s12871-016-0195-0
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
Fig. 1Illustration of setup and cardiac power calculation. Top image depicts the setup with the GE Vingmed SD-100 Doppler monitor (A), the Finometer blood pressure monitor (B), the Dräger Evita 4 ventilator (C) and the in-house LBNP chamber (D). The traces from the suprasternal Doppler flow and the Finometer blood pressure measurement were recorded by in-house software, and used as basis for calculation of a cardiac power trace (PWR) in Matlab
Fig. 2Illustration of impedance analysis by Fourier transformation. Pressure and flow from the aorta are converted into a series of sine waves using Fourier transformation. For each harmonic the corresponding waves of pressure and flow are related to calculate an amplitude and phase angle of each harmonic of the impedance. Only the first five harmonics are included in the figure for illustative purposes
Dynamic parameters. Changes in respiratory variation in power, flow and pressure parameters, both during spontaneous breathing and NPPV
| Power parameters, dynamic | ||||||||
| Delta PWRmax (*10-2) | Delta PWR-integral (*10-2) | |||||||
| Spontaneous breathing | NPPV | Spontaneous breathing | NPPV | |||||
| Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | |
| Baseline | 11.3 (7.0) | 16.4 (7.6) | 18.3 (8.9) | 20.8 | ||||
| 20 mmHg | *14.3 (11.6) | +27 % | 16.2 (6.9) | 19.7 (17.6) | 20.3 | |||
| 40 mmHg | **16.3 (9.4) | +44 % | 16.4 (7.7) | 21.6 (13.7) | *26 (15.4) | +25 % | ||
| 60 mmHg | *15.7 (8.9) | +39 % | **20.6 (9.0) | +26 % | 22.8 (12.3) | *25.9 (14) | +25 % | |
| 80 mmHg | **21.6 (11.8) | +91 % | **27.1 (9.7) | +65 % | **30.7 (23.2) | +68 % | **39.9 (22.2) | +92 % |
| Flow parameters, dynamic | ||||||||
| Delta peak aortic flow (*10-2) | Delta stroke volume (*10-2) | |||||||
| Spontaneous breathing | NPPV | Spontaneous breathing | NPPV | |||||
| Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | |
| Baseline | 8.9 (6.0) | 14 (7.2) | 15.8 (7.8) | 20.1 (11.7) | ||||
| 20 mmHg | **12.3 (9.0) | +38 % | 14.5 (7.3) | 17.4 (17.1) | 19.2 (11.4) | |||
| 40 mmHg | **14.6 (9.2) | +64 % | 14.6 (7.3) | 20.4 (14.9) | *25.3 (16.6) | +26 % | ||
| 60 mmHg | **13 (7.6) | +46 % | 15.4 (8.3) | *22.5 (14.6) | +42 % | *25.9 (14.0) | +29 % | |
| 80 mmHg | **18.1 (10.7) | +103 % | **20.3 (7.9) | +45 % | **30.4 (23.3) | +92 % | **40.6 (23.7) | +102 % |
| Pressure parameters, dynamic | ||||||||
| Delta systolic pressure (*10-2) | Delta pulse pressure (*10-2) | |||||||
| Spontaneous breathing | NPPV | Spontaneous breathing | NPPV | |||||
| Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | |
| Baseline | 3.8 (2.6) | 4.3 (2.6) | 7.5 (4.5) | 6.8 (6.0) | ||||
| 20 mmHg | 4.5 (2.6) | 4.7 (2.5) | 8.2 (5.6) | 8 (4.9) | ||||
| 40 mmHg | 4.7 (3.8) | *5.5 (2.6) | +28 % | 8.2 (6.9) | *8.9 (5.0) | +31 % | ||
| 60 mmHg | *5.3 (3.5) | +39 % | *7.3 (3.8) | +70 % | 8.7 (6.7) | **13.1 (8.0) | +93 % | |
| 80 mmHg | **6.7 (3.8) | +76 % | **10 (4.3) | +133 % | **10.8 (5.8) | +44 % | **14.9 (6.5) | +119 % |
*Significant to a level of p < 0.05
**Significant to a level of p < 0.001
Fig. 3Dynamic parameters during spontaneous breathing. Error bars represent 95 % confidence interval. LBNP: lower body negative pressure
Fig. 4Dynamic parameters during non-invasive positive pressure ventilation (NPPV). Error bars represent 95 % confidence intervals. LBNP: lower body negative pressure
Non-dynamic parameters. Reduction in non-dynamic power, flow and pressure, both during spontaneous breathing and NPPV
| Power parameters, non-dynamic | ||||||||||||
| PWRmax (Watt) | PWR-integral (Joule) | CPO (Watt) | ||||||||||
| Spontaneous breathing | NPPV | Spontaneous breathing | NPPV | Spontaneous breathing | NPPV | |||||||
| Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | |
| Baseline | 13.6 (2.8) | 13.3 (2.4) | 1.03 (0.14) | 1.01 (0.16) | 0.92 (0.23) | 0.95 (0.21) | ||||||
| 20 mmHg | 13.3 (3.2) | **11.8 (3.1) | −11 % | **0.96 (0.17) | −6.9 % | **0.84 (0.19) | −17 % | 0.89 (0.23) | **0.81 (0.20) | −15 % | ||
| 40 mmHg | **11.6 (3.2) | −15 % | **10.2 (3.5) | −23 % | **0.77 (0.19) | −25 % | **0.66 (0.20) | −34 % | **0.73 (0.22) | −21 % | **0.73 (0.25) | −23 % |
| 60 mmHg | **9.9 (2.8) | −27 % | **8.8 (3.3) | −34 % | **0.69 (0.19) | −41 % | **0.51 (0.19) | −50 % | **0.71 (0.21) | −23 % | **0.64 (0.27) | −33 % |
| 80 mmHg | **7.4 (1.9) | −46 % | **6.7 (3.0) | −50 % | **0.42 (0.13) | −59 % | **0.32 (0.11) | −68 % | **0.57 (0.15) | −38 % | **0.48 (0.18) | −49 % |
| Flow parameters, non-dynamic | ||||||||||||
| Aorta maxflow (m/s) | Stroke volume (ml) | CO (L/min) | ||||||||||
| Spontaneous breathing | NPPV | Spontaneous breathing | NPPV | Spontaneous breathing | NPPV | |||||||
| Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | |
| Baseline | 0.88 (0.12) | 0.85 (0.15) | 83.0 (10.6) | 77.9 (13.4) | 4.91 (1.08) | **4.85 (1.33) | ||||||
| 20 mmHg | 0.85 (0.15) | *0.78 (0.16) | −8.2 % | *77.3 (12.1) | −6.9 % | **67.6 (16.4) | −13 % | 4.74 (1.2) | **4.26 (0..99) | −12 % | ||
| 40 mmHg | **0.76 (0.18) | −14 % | **0.66 (0.21) | −22 % | **63.0 (15.3) | −24 % | **52.3 (17.6) | −33 % | **3.93 (1.11) | −20 % | **3.64 (1.28) | −25 % |
| 60 mmHg | **0.67 (0.17) | −24 % | **0.62 (0.18) | −27 % | **50.7 (14.1) | −39 % | **44.2 (13.7) | −43 % | **3.74 (1.08) | −24 % | **3.48 (1.33) | −28 % |
| 80 mmHg | **0.55 (0.14) | −38 % | **0.53 (0.14) | −38 % | **37.0 (15.5) | −55 % | **30.4 (11.9) | −61 % | **3.13 (1.06) | −36 % | **2.89 (1.01) | −40 % |
| Pressure parameters, non-dynamic | ||||||||||||
| Systolic pressure (mmHg) | Pulse pressure (mmHg) | MAP (mmHg) | ||||||||||
| Spontaneous breathing | NPPV | Spontaneous breathing | NPPV | Spontaneous breathing | NPPV | |||||||
| Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | Mean (sd) | Change | |
| Baseline | 118 (12.2) | 123 (17.3) | 59.2 (11.4) | 61.5 (10.8) | 75.7 (10.1) | 80.2 (16.1) | ||||||
| 20 mmHg | 119 (16.2) | 121 (22.2) | 59.9 (16.5) | 59.7 (15.4) | 76.0 (11.7) | 78.0 (18.2) | ||||||
| 40 mmHg | 120 (15.6) | 124 (21.6) | 58.8 (15.8) | *56.1 (18.3) | −9 % | 76.3 (11.7) | 82.1 (18.7) | |||||
| 60 mmHg | 115 (13.4) | *114 (27.0) | −7.3 % | **50 (11.7) | −16 % | **50.2 (19.3) | −18 % | 77.9 (12.3) | 76.8 (21.4) | |||
| 80 mmHg | **108 (−12.6) | −8.4 % | **98 (27.7) | −20 % | **44.2 (10.7) | −25 % | **40.3 (18.1) | −34 % | 76.5 (13.0) | **69.7 (22.8) | −13.1 % | |
*Significant to a level of p < 0.05
**Significant to a level of p < 0.001
Fig. 5Non-dynamic hemodynamic parameters during spontaneous breathing. Error bars represent 95 % confidence intervals. LBNP: lower body negative pressure
Fig. 6Non-dynamic hemodynamic parameters during non-invasive positive pressure ventilation (NPPV). Error bars represent 95 % confidence intervals. LBNP: lower body negative pressure
Fig. 7The PWR trace, which is the product of the synchronized aortic flow trace and the blood pressure trace. The arrows indicate that respiratory variation in aortic flow is 2–3 heart cycles ahead of respiratory variation in blood pressure