Capnogram slope and ventilation dead space parameters: comparison of mainstream and sidestream techniques.

A L Balogh1, F Petak2, G H Fodor3, J Tolnai3, Z Csorba4, B Babik4.   

Abstract

BACKGROUND: Capnography may provide useful non-invasive bedside information concerning heterogeneity in lung ventilation, ventilation-perfusion mismatching and metabolic status. Although the capnogram may be recorded by mainstream and sidestream techniques, the capnogram indices furnished by these approaches have not previously been compared systematically.
METHODS: Simultaneous mainstream and sidestream time and volumetric capnography was performed in anaesthetized, mechanically ventilated patients undergoing elective heart surgery. Time capnography was used to assess the phase II (SII,T) and III slopes (SIII,T). The volumetric method was applied to estimate phase II (SII,V) and III slopes (SIII,V), together with the dead space values according to the Fowler (VDF), Bohr (VDB), and Enghoff (VDE) methods and the volume of CO2 eliminated per breath ([Formula: see text]). The partial pressure of end-tidal CO2 ([Formula: see text]) was registered.
RESULTS: Excellent correlation and good agreement were observed in SIII,T measured by the mainstream and sidestream techniques [ratio=1.05 (sem 0.16), R(2)=0.92, P<0.0001]. Although the sidestream technique significantly underestimated [Formula: see text] and overestimated SIII,V [1.32 (0.28), R(2)=0.93, P<0.0001], VDF, VDB, and VDE, the agreement between the mainstream and sidestream techniques in the difference between VDE and VDB, reflecting the intrapulmonary shunt, was excellent [0.97 (0.004), R(2)=0.92, P<0.0001]. The [Formula: see text] exhibited good correlation and mild differences between the mainstream and sidestream approaches [0.025 (0.005) kPa].
CONCLUSIONS: Sidestream capnography provides adequate quantitative bedside information about uneven alveolar emptying and ventilation-perfusion mismatching, because it allows reliable assessments of the phase III slope, [Formula: see text] and intrapulmonary shunt. Reliable measurement of volumetric parameters (phase II slope, dead spaces, and eliminated CO2 volumes) requires the application of a mainstream device.
© The Author 2016. Published by Oxford University Press on behalf of the British Journal of Anaesthesia. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Keywords:  capnography; carbon dioxide; intraoperative monitoring; mechanical ventilation; ventilation-perfusion ratio

Mesh:

Substances:

Year:  2016        PMID: 27317710      PMCID: PMC4913399          DOI: 10.1093/bja/aew127

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  26 in total

1.  A method of reconstruction of clinical gas-analyzer signals corrupted by positive-pressure ventilation.

Authors:  A D Farmery; C E Hahn
Journal:  J Appl Physiol (1985)       Date:  2001-04

Review 2.  Capnographic waveforms in the mechanically ventilated patient.

Authors:  John E Thompson; Michael B Jaffe
Journal:  Respir Care       Date:  2005-01       Impact factor: 2.258

Review 3.  Monitoring ventilation with capnography.

Authors:  Rafael Ortega; Christopher Connor; Sora Kim; Robin Djang; Kunal Patel
Journal:  N Engl J Med       Date:  2012-11-08       Impact factor: 91.245

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Determinants of distortions in CO2 catheter sampling systems: a mathematical model.

Authors:  R A Epstein; A M Reznik; M A Epstein
Journal:  Respir Physiol       Date:  1980-07

6.  Disparity between mainstream and sidestream end-tidal carbon dioxide values and arterial carbon dioxide levels.

Authors:  Murat Pekdemir; Orhan Cinar; Serkan Yilmaz; Elif Yaka; Melih Yuksel
Journal:  Respir Care       Date:  2013-01-15       Impact factor: 2.258

7.  Capnometer transport delay: measurement and clinical implications.

Authors:  P H Breen; B Mazumdar; S C Skinner
Journal:  Anesth Analg       Date:  1994-03       Impact factor: 5.108

8.  Lung recruitment improves the efficiency of ventilation and gas exchange during one-lung ventilation anesthesia.

Authors:  Gerardo Tusman; Stephan H Böhm; Fernando Suárez Sipmann; Stefan Maisch
Journal:  Anesth Analg       Date:  2004-06       Impact factor: 5.108

9.  Flow-through versus sidestream capnometry for detection of end tidal carbon dioxide in the sedated patient.

Authors:  Derek J Sakata; Isao Matsubara; Nishant A Gopalakrishnan; Dwayne R Westenskow; Julia L White; Shinji Yamamori; Talmage D Egan; Nathan L Pace
Journal:  J Clin Monit Comput       Date:  2009-03-20       Impact factor: 2.502

10.  Effects of respiratory mechanics on the capnogram phases: importance of dynamic compliance of the respiratory system.

Authors:  Barna Babik; Zsófia Csorba; Dorottya Czövek; Patrick N Mayr; Gábor Bogáts; Ferenc Peták
Journal:  Crit Care       Date:  2012-10-02       Impact factor: 9.097

View more
  4 in total

1.  Effect of body position on the redistribution of regional lung aeration during invasive and non-invasive ventilation of COVID-19 patients.

Authors:  André Dos Santos Rocha; John Diaper; Adam L Balogh; Christophe Marti; Olivier Grosgurin; Walid Habre; Ferenc Peták; Roberta Südy
Journal:  Sci Rep       Date:  2022-06-30       Impact factor: 4.996

2.  End-tidal carbon dioxide measurement in preterm infants with low birth weight.

Authors:  Hsin-Ju Lin; Ching-Tzu Huang; Hsiu-Feng Hsiao; Ming-Chou Chiang; Mei-Jy Jeng
Journal:  PLoS One       Date:  2017-10-17       Impact factor: 3.240

Review 3.  Carbon Dioxide Sensing-Biomedical Applications to Human Subjects.

Authors:  Emmanuel Dervieux; Michaël Théron; Wilfried Uhring
Journal:  Sensors (Basel)       Date:  2021-12-28       Impact factor: 3.576

4.  Volumetric capnography pre- and post-surfactant during initial resuscitation of premature infants.

Authors:  Emma E Williams; Theodore Dassios; Katie A Hunt; Anne Greenough
Journal:  Pediatr Res       Date:  2021-05-22       Impact factor: 3.953

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.